Ex ova chick chorioallantoic membrane as a novel in vivo model for testing biosensors

被引:33
作者
Valdes, TI
Klueh, U
Kreutzer, D
Moussy, F
机构
[1] Dept Mech Engn, Tampa, FL 33620 USA
[2] Ctr Diabet, Tampa, FL 33620 USA
[3] Univ Connecticut, Ctr Hlth, Ctr Biomat, Farmington, CT 06030 USA
[4] Univ Connecticut, Ctr Hlth, Dept Pathol, Farmington, CT 06030 USA
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A | 2003年 / 67A卷 / 01期
关键词
chicken chorioallantoic membrane; biomaterials; biosensors; animal models; in vivo studies;
D O I
10.1002/jbm.a.10055
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A major problem with implantable sensors is their short in vivo lifetime, due to strong tissue reactions (i.e., inflammation and fibrosis) caused by the implant and the failure of sensor components. The tissue reactions to the sensor, the biocompatibility of components, and the function of the sensor must be evaluated by using in vivo models. Current methods of in vivo biosensor testing are time- and labor- intensive and expensive. In addition, the results often vary on the basis of the surgical skills of the investigator. The in ova chorioallantoic membrane (CAM) of the developing chicken embryo was previously developed in our laboratory as a novel in vivo system to test biomaterials. In this new article, we describe a novel approach for testing biosensors in vivo using the ex ova CAM model as an alternative to the traditional mammalian models. Fertilized chicken eggs were incubated for 3 days in ova and then transferred into a petri dish (ex ova) for further incubation at 37degreesC and 80% humidity. After 1 week of incubation, acetaminophen biosensors, used as model sensors, were placed on top of the CAM and allowed to incorporate for 1 week. Biosensors were then tested for their sensitivity to acetaminophen. CAM venules were injected with 0.2 mL of a 3.6 mM acetaminophen solution. The current produced by the sensor reflected the change in blood acetaminophen levels. Sensors were also assessed by using gross and histological evaluations. We previously reported on the similarity of the tissue response of the CAM with the mammalian models. The low cost, simplicity, and possibility to continuously visualize the sensor test site through a cell culture dish make this animal model particularly attractive for the rapid in vivo screening of biosensors. (C) 2003 Wiley Periodicals, Inc.
引用
收藏
页码:215 / 223
页数:9
相关论文
共 20 条
[1]   ION-TRANSPORT ACROSS THE EARLY CHICK-EMBRYO .2. CHARACTERIZATION AND PH SENSITIVITY OF THE TRANSEMBRYONIC SHORT-CIRCUIT CURRENT [J].
ABRIEL, H ;
KATZ, U ;
KUCERA, P .
JOURNAL OF MEMBRANE BIOLOGY, 1994, 141 (02) :159-166
[2]  
BURLEY RW, 1989, AVAIN EGG CHEM BIOL
[3]   OVEREXPRESSION OF VASCULAR ENDOTHELIAL GROWTH-FACTOR IN THE AVIAN EMBRYO INDUCES HYPERVASCULARIZATION AND INCREASED VASCULAR-PERMEABILITY WITHOUT ALTERATIONS OF EMBRYONIC PATTERN-FORMATION [J].
FLAMME, I ;
VONREUTERN, M ;
DREXLER, HCA ;
SYEDALI, S ;
RISAU, W .
DEVELOPMENTAL BIOLOGY, 1995, 171 (02) :399-414
[4]   Performance of subcutaneously implanted glucose sensors for continuous monitoring [J].
Gerritsen, M ;
Jansen, JA ;
Lutterman, JA .
NETHERLANDS JOURNAL OF MEDICINE, 1999, 54 (04) :167-179
[5]  
Insel PA., 1996, Goodman and Gilman's The Pharmacological Basis of Therapeutics, P617
[6]   Effects of anti-inflammatory drugs on increased vascular permeability in acute inflammatory response in the chicken [J].
Jain, NK ;
Vegad, JL ;
Katiyar, AK ;
Awadhiya, RP .
AVIAN PATHOLOGY, 1995, 24 (04) :723-729
[7]   CHICK-EMBRYO CHORIOALLANTOIC MEMBRANE AS A BIOASSAY FOR ANGIOGENESIS FACTORS - REACTIONS INDUCED BY CARRIER MATERIALS [J].
JAKOB, W ;
JENTZSCH, KD ;
MAUERSBERGER, B ;
HEDER, G .
EXPERIMENTELLE PATHOLOGIE, 1978, 15 (05) :241-249
[8]   THE FUNCTION OF A HYDROGEN PEROXIDE-DETECTING ELECTROENZYMATIC GLUCOSE ELECTRODE IS MARKEDLY IMPAIRED IN HUMAN SUB-CUTANEOUS TISSUE AND PLASMA [J].
KERNER, W ;
KIWIT, M ;
LINKE, B ;
KECK, FS ;
ZIER, H ;
PFEIFFER, EF .
BIOSENSORS & BIOELECTRONICS, 1993, 8 (9-10) :473-482
[9]  
LEIGTON JL, 1967, SPREAD CANC
[10]  
Lucarelli E, 1999, INT J MOL MED, V4, P593