Combined cardiopulmonary assessments with implantable telemetry device in conscious freely moving cynomolgus monkeys

被引:14
|
作者
Authier, S. [1 ,2 ]
Haefner, Paul [3 ]
Fournier, S. [1 ]
Troncy, E. [2 ]
Moon, L. B. [3 ]
机构
[1] LAB Res Inc, Laval, PQ H7V 4B3, Canada
[2] Univ Montreal, Fac Med Vet, GREPAQ, Dept Vet Biomed, St Hyacinthe, PQ J2S 7C6, Canada
[3] Data Sci Int, St Paul, MN 55112 USA
关键词
Respiratory safety pharmacology; Cynomolgus monkeys; Impedance; Tidal volume; Respiratory rate; Minute volume; Telemetry; ICH S7A guideline; CARDIOVASCULAR MONITORING-SYSTEM; REGULATORY SAFETY PHARMACOLOGY; IMPEDANCE MINUTE VENTILATION; PACEMAKER; VALIDATION; SENSOR;
D O I
10.1016/j.vascn.2010.05.014
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Female cynomolgus monkeys were surgically implanted with telemetry transmitters recording ECG (DII), arterial pressure, physical activity, body temperature, and tidal volume. Respiratory rate (RR) and tidal volume (TV) were monitored simultaneously with the telemetry transmitter using impedance. Impedance-based monitoring of RR and TV by telemetry correlated with controlled TV and with pneumotachometer (>98%) in restrained animals. Control drugs with cardiovascular and respiratory effects, including saline, medetomidine (0.01, 0.02 and 0.04 mg/kg) and cocaine (0.5, 1.0 and 1.5 mg/kg) were administered intravenously. An averaging epoch of 5 min was used for analysis of respiratory data. Medetomidine induced significant respiratory depression with decrease in RR and TV in freely moving animals while cocaine increased TV, RR and minute ventilation (MV) with concomitant increase in heart rate when compared with time matched values from saline-treated animals. The onset, duration and magnitude of cardiovascular and respiratory changes were correlated. This highlights the dependency of the cardiovascular and respiratory systems. The use of cardiopulmonary monitoring can allow continuous monitoring including during night time when variability of respiratory parameters is lower. Monitoring of cardiovascular and respiratory parameters in the same animals could also help to decrease the number of animals used in research. (C) 2010 Elsevier Inc. All rights reserved.
引用
收藏
页码:6 / 11
页数:6
相关论文
共 39 条
  • [1] Combined Cardiopulmonary Assessments Using Impedance and Digital Implants in Conscious Freely Moving Cynomolgus Monkeys, Beagle Dogs, and Gottingen Minipigs: Pharmacological Characterization and Social Housing Effects
    Boulay, Emmanuel
    Troncy, Eric
    Pugsley, Michael
    St-Pierre, Joey
    Downey, Anne-Marie
    Smutova, Viktoria
    Guerrier, Mireille
    Maghezzi, Said
    Authier, Simon
    INTERNATIONAL JOURNAL OF TOXICOLOGY, 2021, 40 (06) : 530 - 541
  • [2] Advantages of Glucose Monitoring with Implantable Telemetry in Freely Moving Conscious Nonhuman Primates
    Wang, Bingdi
    Qiao, Wei
    Sun, Guofeng
    Liu, Yongqiang
    Qiao, Jiqiu
    Yi, Weiwei
    Wang, Hui
    Wang, Xiaoli
    Lindquist, Ryan
    Wang, Yixin
    Xiao, Yongfu
    DIABETES, 2016, 65 : A234 - A234
  • [3] Dose-response effects of sotalol on cardiovascular function in conscious, freely moving cynomolgus monkeys
    Lynch, J. J., III
    Wilson, A. W.
    Hernandez, L. E.
    Nelson, R. A.
    Marsh, K. C.
    Cox, B. F.
    Mittelstadt, S. W.
    BRITISH JOURNAL OF PHARMACOLOGY, 2008, 154 (07) : 1439 - 1445
  • [4] Accurate evaluation of QT interval in conscious cynomolgus monkeys by use of telemetry ECG
    Hijioka, M
    Kuwano, K
    Jikuzono, T
    Otsuka, S
    Miyake, T
    Uchino, T
    Shoki, M
    Koga, T
    Fukuzaki, K
    Nagata, R
    Kito, G
    JOURNAL OF PHARMACOLOGICAL SCIENCES, 2005, 97 : 240P - 240P
  • [5] RECORDING MYOMETRIAL ACTIVITY IN FREELY-MOVING RHESUS MONKEYS BY TELEMETRY
    MICHAEL, RP
    WELLER, C
    JOURNAL OF ENDOCRINOLOGY, 1968, 42 (02) : 357 - &
  • [6] Long-term blood glucose monitoring with implanted telemetry device in conscious and stress-free cynomolgus monkeys
    Wang, B.
    Sun, G.
    Qiao, W.
    Liu, Y.
    Qiao, J.
    Ye, W.
    Wang, H.
    Wang, X.
    Lindquist, R.
    Wang, Y.
    Xiao, Y. -F.
    JOURNAL OF ENDOCRINOLOGICAL INVESTIGATION, 2017, 40 (09): : 967 - 977
  • [7] Long-term blood glucose monitoring with implanted telemetry device in conscious and stress-free cynomolgus monkeys
    B. Wang
    G. Sun
    W. Qiao
    Y. Liu
    J. Qiao
    W. Ye
    H. Wang
    X. Wang
    R. Lindquist
    Y. Wang
    Y.-F. Xiao
    Journal of Endocrinological Investigation, 2017, 40 : 967 - 977
  • [8] Assessment of Subcutaneous Radio-Telemetry Device Implantation in Cynomolgus Monkeys
    Queudot, J-C.
    Forget, J.
    Nechev, S.
    Celori, A.
    Megrath, B.
    Hernandez, T.
    INTERNATIONAL JOURNAL OF TOXICOLOGY, 2021, 40 (01) : 71 - 71
  • [9] Feasibility of Continuous and Prolonged Intravenous Infusion in Freely Moving Pregnant Cynomolgus Monkeys
    Korte, S. H.
    Nowak, P.
    Bee, W.
    Fuchs, A.
    Weinbauer, G. F.
    BIRTH DEFECTS RESEARCH PART A-CLINICAL AND MOLECULAR TERATOLOGY, 2010, 88 (05) : 390 - 390
  • [10] Simultaneous measurement of arterial and left ventricular pressure in conscious freely moving rats by telemetry
    Segreti, Jason A.
    Polakowski, James S.
    Blomme, Eric A.
    King, Andrew J.
    JOURNAL OF PHARMACOLOGICAL AND TOXICOLOGICAL METHODS, 2016, 79 : 23 - 33