MS-QI: A Modulation Spectrum-Based ECG Quality Index for Telehealth Applications

被引:56
作者
Tobon, Diana P., V [1 ]
Falk, Tiago H. [1 ]
Maier, Martin [1 ]
机构
[1] Inst Natl Rech Sci, Energie Mat Telecommun, Montreal, PQ G1K 9A9, Canada
关键词
Electrocardiogram (ECG); kurtosis; modulation spectrum; quality index; telehealth; ELECTROCARDIOGRAM SIGNAL; IMPACT;
D O I
10.1109/TBME.2014.2355135
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
As telehealth applications emerge, the need for accurate and reliable biosignal quality indices has increased. One typical modality used in remote patient monitoring is the electrocardiogram (ECG), which is inherently susceptible to several different noise sources, including environmental (e.g., powerline interference), experimental (e.g., movement artifacts), and physiological (e.g., muscle and breathing artifacts). Accurate measurement of ECG quality can allow for automated decision support systems to make intelligent decisions about patient conditions. This is particularly true for in-home monitoring applications, where the patient is mobile and the ECG signal can be severely corrupted by movement artifacts. In this paper, we propose an innovative ECG quality index based on the so-called modulation spectral signal representation. The representation quantifies the rate of change of ECG spectral components, which are shown to be different from the rate of change of typical ECG noise sources. The proposed modulation spectral-based quality index, MS-QI, was tested on 1) synthetic ECG signals corrupted by varying levels of noise, 2) single-lead recorded data using the Hexoskin garment during three activity levels (sitting, walking, running), 3) 12-lead recorded data using conventional ECG machines (Computing in Cardiology 2011 dataset), and 4) two-lead ambulatory ECG recorded from arrhythmia patients (MIT-BIH Arrhythmia Database). Experimental results showed the proposed index outperforming two conventional benchmark quality measures, particularly in the scenarios involving recorded data in real-world environments.
引用
收藏
页码:1613 / 1622
页数:10
相关论文
共 33 条
  • [1] del Río BAS, 2011, COMPUT CARDIOL CONF, V38, P609
  • [2] [Anonymous], 2007, P INTERSPEECH
  • [3] Atlas L., 2010, TECH REP
  • [4] Behar J, 2012, COMPUT CARDIOL CONF, V39, P381
  • [5] ECG Signal Quality During Arrhythmia and Its Application to False Alarm Reduction
    Behar, Joachim
    Oster, Julien
    Li, Qiao
    Clifford, Gari D.
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2013, 60 (06) : 1660 - 1666
  • [6] Signal quality indices and data fusion for determining clinical acceptability of electrocardiograms
    Clifford, G. D.
    Behar, J.
    Li, Q.
    Rezek, I.
    [J]. PHYSIOLOGICAL MEASUREMENT, 2012, 33 (09) : 1419 - 1433
  • [7] Signal quality in cardiorespiratory monitoring
    Clifford, Gari D.
    Moody, George B.
    [J]. PHYSIOLOGICAL MEASUREMENT, 2012, 33 (09)
  • [8] Evaluation of an algorithm based on single-condition decision rules for binary classification of 12-lead ambulatory ECG recording quality
    Di Marco, Luigi Yuri
    Duan, Wenfeng
    Bojarnejad, Marjan
    Zheng, Dingchang
    King, Susan
    Murray, Alan
    Langley, Philip
    [J]. PHYSIOLOGICAL MEASUREMENT, 2012, 33 (09) : 1435 - 1448
  • [9] Falk T. H., 2010, New Developments in Biomedical Engineering, P93
  • [10] Clinical evaluation of a wireless ECG sensor system for arrhythmia diagnostic purposes
    Fensli, Rune
    Gundersen, Torstein
    Snaprud, Tormod
    Hejlesen, Ole
    [J]. MEDICAL ENGINEERING & PHYSICS, 2013, 35 (06) : 697 - 703