Comparison of Automated Breast Volume Scanning and Hand-Held Ultrasound in the Detection of Breast Cancer: an Analysis of 5,566 Patient Evaluations

被引:40
作者
Choi, Woo Jung [1 ]
Cha, Joo Hee [1 ]
Kim, Hak Hee [1 ]
Shin, Hee Jung [1 ]
Kim, Hyunji [1 ]
Chae, Eun Young [1 ]
Hong, Min Ji [1 ]
机构
[1] Univ Ulsan, Coll Med, Dept Radiol, Res Inst Radiol,Asan Med Ctr, Seoul, South Korea
关键词
Breast cancer; automated breast volume scanning; ultrasound; screening; SCREENING MAMMOGRAPHY; ULTRASONOGRAPHY; WOMEN; KOREA; DIAGNOSIS; RISK; US; PERFORMANCE; ACCURACY; PROGRAM;
D O I
10.7314/APJCP.2014.15.21.9101
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: The purpose of this study was to compare the accuracy and effectiveness of automated breast volume scanning (ABVS) and hand-held ultrasound (HHUS) in the detection of breast cancer in a large population group with a long-term follow-up, and to investigate whether different ultrasound systems may influence the estimation of cancer detection. Materials and Methods: Institutional review board approval was obtained for this retrospective study, and informed consent was waived. From September 2010 to August 2011, a total of 1,866 ABVS and 3,700 HHUS participants, who underwent these procedures at our institute, were included in this study. Cancers occurring during the study and subsequent follow-up were evaluated. The reference standard was a combination of histology and follow-up imaging (>= 12 months). The recall rate, cancer detection yield, diagnostic accuracy, sensitivity, specificity, and positive (PPV) and negative (NPV) predictive values were calculated with exact 95% confidence intervals. Results: The recall rate was 2.57 per 1,000 (48/1,866) for ABVS and 3.57 per 1,000 (132/3,700) for HHUS, with. significant difference (p=0.048). The cancer detection yield was 3.8 per 1,000 for ABVS and 2.7 per 1,000 for HHUS. The diagnostic accuracy was 97.7% for ABVS and 96.5% for HHUS with statistical significance (p=0.018). The specificity of ABVS and HHUS were 97.8%, 96.7%, respectively (p=0.022). Conclusions: ABVS shows a comparable diagnostic performance to HHUS. ABVS is an effective supplemental tool for mammography in breast cancer detection in a large population.
引用
收藏
页码:9101 / 9105
页数:5
相关论文
共 40 条
  • [1] Comparison of Automated Breast Ultrasound and Hand-Held Breast Ultrasound in the Screening of Dense Breasts
    Philadelpho, Fernanda
    Gregorio Calas, Maria Julia
    Coutinho Carneiro, Gracy de Almeida
    Silveira, Isabela Cunha
    Ribeiro Vaz, Andreia Brandao
    Coelho Nogueira, Adriana Maria
    Bergmann, Anke
    Proenca Lobo Lopes, Flavia Paiva
    REVISTA BRASILEIRA DE GINECOLOGIA E OBSTETRICIA, 2021, 43 (03): : 190 - 199
  • [2] Comparison of automated breast volume scanning to hand-held ultrasound and mammography
    Wang, Zhi Li
    Xw, Jian Hong
    Li, Jun Lai
    Huang, Yan
    Tang, Jie
    RADIOLOGIA MEDICA, 2012, 117 (08): : 1287 - 1293
  • [3] Whole Breast Ultrasound: Comparison of the Visibility of Suspicious Lesions with Automated Breast Volumetric Scanning Versus Hand-Held Breast Ultrasound
    Kuzmiak, Cherie M.
    Ko, Eun Y.
    Tuttle, Laura A.
    Steed, Doreen
    Zeng, Donglin
    Yoon, Sora C.
    ACADEMIC RADIOLOGY, 2015, 22 (07) : 870 - 879
  • [4] Automated Breast Volume Scanner Is More Valuable Than Hand-Held Ultrasound in Diagnosis of Small Breast cancer
    Yan, Lixia
    Jing, Luxia
    Lu, Qing
    Wang, Xi
    Mao, Wujian
    Wang, Peilei
    Zhan, Mengna
    Huang, Beijian
    ULTRASOUND QUARTERLY, 2024, 40 (01) : 66 - 73
  • [5] Automated Breast Volume Scanning Versus Conventional Ultrasound in Breast Cancer Screening
    Xiao, Yuanming
    Zhou, Qichang
    Chen, Zhiheng
    ACADEMIC RADIOLOGY, 2015, 22 (03) : 387 - 399
  • [6] Diagnostic performance of combined use of automated breast volume scanning & hand-held ultrasound for breast lesions
    Liu, Jialin
    Zhou, Yang
    Wu, Jialing
    Li, Peng
    Liang, Xinyu
    Duan, Haonan
    Wu, Xueqing
    Hou, Xiukun
    Li, Xiaofeng
    INDIAN JOURNAL OF MEDICAL RESEARCH, 2021, 154 (02) : 347 - 354
  • [7] Diagnostic value of an automated breast volume scanner compared with a hand-held ultrasound: a meta-analysis
    Zhang, Xiaohui
    Chen, Juan
    Zhou, Yidong
    Mao, Feng
    Lin, Yan
    Shen, Songjie
    Sun, Qiang
    Ouyang, Zhaolian
    GLAND SURGERY, 2019, 8 (06) : 698 - +
  • [8] Automated breast ultrasound compared to hand-held ultrasound in surveillance after breast-conserving surgery
    Depretto, Catherine
    Liguori, Alessandro
    Primolevo, Alessandra
    Di Cosimo, Serena
    Cartia, Francesco
    Ferranti, Claudio
    Scaperrotta, Gianfranco P.
    TUMORI JOURNAL, 2021, 107 (02): : 132 - 138
  • [9] A Comparison of the Acceptance of Female Subjects Between Mammography, Automated Breast Ultrasound and Hand-Held Ultrasound
    Gao, Xican
    Liu, Liwen
    Yan, Jingru
    Yang, Shanling
    Shu, Rui
    Ju, Yan
    Zhang, Ge
    Xiao, Yao
    Song, Hongping
    JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2019, 9 (04) : 768 - 775
  • [10] Comparative analysis of automated breast volume scanner (ABVS) combined with conventional hand-held ultrasound and mammography in female breast cancer detection
    Sherchan, Amir
    Liang, Jin Tian
    Sherchan, Bhakti
    Suwal, Sundar
    Katwal, Shailendra
    ANNALS OF MEDICINE AND SURGERY, 2024, 86 (01): : 159 - 165