Comparison of reversed-phase liquid chromatography and hydrophilic interaction/cation-exchange chromatography for the separation of amphipathic α-helical peptides with L- and D-amino acid substitutions in the hydrophilic face

被引:31
作者
Hartmann, E
Chen, YX
Mant, CT
Jungbauer, A
Hodges, RS [1 ]
机构
[1] Univ Colorado, Hlth Sci Ctr, Dept Biochem & Mol Genet, Denver, CO 80262 USA
[2] Univ Agr Sci, Inst Appl Microbiol, A-1190 Vienna, Austria
[3] Univ Alberta, Dept Biochem, Edmonton, AB T6G 2H7, Canada
关键词
mixed-mode chromatography; hydrophilic interaction chromatography; peptides;
D O I
10.1016/S0021-9673(03)00620-4
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Mixed-mode hydrophilic interaction/cation-exchange chromatography (HILIC/CEX) is a novel high-performance technique which has excellent potential for peptide separations. Separations by HILIX/CEX are carried out by subjecting peptides to linear increasing salt gradients in the presence of high levels of acetonitrile, which promotes hydrophilic interactions overlaid on ionic interactions with the cation-exchange matrix. In the present study, HILIC/CEX has been compared to reversed-phase liquid chromatography (RP-HPLC) for separation of mixtures of diastereomeric amphipathic alpha-helical peptide analogues, where L- and D-amino acid substitutions were made in the centre of the hydrophilic face of the amphipathic alpha-helix. Unlike RP-HPLC, temperature had a substantial effect on HILIC/CEX of the peptides, with a rise in temperature from 25 to 65 degreesC increasing the retention times of the peptides as well as improving resolution. Our results again highlight the potential of HILIC/CEX as a peptide separation mode in its own right as well as an excellent complement to RP-HPLC. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:61 / 71
页数:11
相关论文
共 29 条
[1]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF AMINO-ACIDS, PEPTIDES AND PROTEINS .128. EFFECT OF D-AMINO-ACID SUBSTITUTIONS ON THE REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY RETENTION BEHAVIOR OF NEUROPEPTIDE Y[18-36] ANALOGS [J].
AGUILAR, MI ;
MOUGOS, S ;
BOUBLIK, J ;
RIVIER, J ;
HEARN, MTW .
JOURNAL OF CHROMATOGRAPHY, 1993, 646 (01) :53-65
[2]   HYDROPHILIC-INTERACTION CHROMATOGRAPHY FOR THE SEPARATION OF PEPTIDES, NUCLEIC-ACIDS AND OTHER POLAR COMPOUNDS [J].
ALPERT, AJ .
JOURNAL OF CHROMATOGRAPHY, 1990, 499 :177-196
[3]   INDUCED CONFORMATIONAL STATES OF AMPHIPATHIC PEPTIDES IN AQUEOUS LIPID ENVIRONMENTS [J].
BLONDELLE, SE ;
OSTRESH, JM ;
HOUGHTEN, RA ;
PEREZPAYA, E .
BIOPHYSICAL JOURNAL, 1995, 68 (01) :351-359
[4]   Determination of stereochemistry stability coefficients of amino acid side-chains in an amphipathic α-helix [J].
Chen, Y ;
Mant, CT ;
Hodges, RS .
JOURNAL OF PEPTIDE RESEARCH, 2002, 59 (01) :18-33
[5]  
Chen Y., UNPUB
[6]   PREDICTION OF PEPTIDE RETENTION TIMES IN REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY .1. DETERMINATION OF RETENTION COEFFICIENTS OF AMINO-ACID-RESIDUES OF MODEL SYNTHETIC PEPTIDES [J].
GUO, DC ;
MANT, CT ;
TANEJA, AK ;
PARKER, JMR ;
HODGES, RS .
JOURNAL OF CHROMATOGRAPHY, 1986, 359 :499-517
[7]   PREDICTION OF PEPTIDE RETENTION TIMES IN REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY .2. CORRELATION OF OBSERVED AND PREDICTED PEPTIDE RETENTION TIMES AND FACTORS INFLUENCING THE RETENTION TIMES OF PEPTIDES [J].
GUO, DC ;
MANT, CT ;
TANEJA, AK ;
HODGES, RS .
JOURNAL OF CHROMATOGRAPHY, 1986, 359 :519-532
[8]   The helix-destabilizing propensity scale of D-amino acids: The influence of side chain steric effects [J].
Krause, E ;
Bienert, M ;
Schmieder, P ;
Wenschuh, H .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2000, 122 (20) :4865-4870
[9]   EFFECTS OF HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC SOLVENTS AND HYDROPHOBIC MATRICES ON THE SECONDARY AND QUATERNARY STRUCTURE OF A MODEL PROTEIN - REVERSED-PHASE AND SIZE-EXCLUSION HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
LAU, SYM ;
TANEJA, AK ;
HODGES, RS .
JOURNAL OF CHROMATOGRAPHY, 1984, 317 (DEC) :129-140
[10]   Separation of acetylated core histones by hydrophilic-interaction liquid chromatography [J].
Lindner, H ;
Sarg, B ;
Meraner, C ;
Helliger, W .
JOURNAL OF CHROMATOGRAPHY A, 1996, 743 (01) :137-144