Ion funnel trap interface for orthogonal time-of-flight mass spectrometry

被引:64
作者
Ibrahim, Yehia [1 ]
Belov, Mikhail E. [1 ]
Tolmachev, Aleksey V. [1 ]
Prior, David C. [1 ]
Smith, Richard D. [1 ]
机构
[1] Pacific NW Natl Lab, Div Biol Sci, Richland, WA 99352 USA
关键词
D O I
10.1021/ac071091m
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A combined electrodynamic ion funnel and ion trap coupled to an orthogonal acceleration (oa)-time-of-flight mass spectrometer was developed and characterized. The ion trap was incorporated through the use of added terminal electrodynamic ion funnel electrodes enabling control over the axial dc gradient in the trap section. The ion trap operates efficiently at a pressure of similar to 1 Torr, and measurements indicate a maximum charge capacity of similar to 3 x 10(7) charges. An order of magnitude increase in sensitivity was observed in the analysis of low concentration peptides mixtures with orthogonal acceleration (oa)time-of-flight mass spectrometry (oa-TOF MS) in the trapping mode as compared to the continuous regime. A signal increase in the trapping mode was accompanied by reduction in the chemical background, due to more efficient desolvation of, for example, solvent related clusters. Controlling the ion trap ejection time was found to result in efficient removal of singly charged species and improving signal-to-noise ratio (S/N) for the multiply charged analytes.
引用
收藏
页码:7845 / 7852
页数:8
相关论文
共 41 条
[31]   Development and characterization of an electrospray ionization ion trap linear time-of-flight mass spectrometer [J].
Purves, RW ;
Li, L .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1997, 8 (10) :1085-1093
[32]  
Schwartz J. C., 1996, Patent No. [5572 022, 5572022]
[33]  
Shaffer SA, 1997, RAPID COMMUN MASS SP, V11, P1813
[34]   SIMION analysis of a high performance linear accumulation octopole with enhanced ejection capabilities [J].
Taban, IM ;
McDonnell, LA ;
Römpp, A ;
Cerjak, I ;
Heeren, RMA .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2005, 244 (2-3) :135-143
[35]  
Thomson BA, 1998, RAPID COMMUN MASS SP, V12, P1323, DOI 10.1002/(SICI)1097-0231(19981015)12:19<1323::AID-RCM344>3.0.CO
[36]  
2-T
[37]   A collisional focusing ion guide for coupling aa atmospheric pressure ion source to a mass spectrometer [J].
Tolmachev, AV ;
Chernushevich, IV ;
Dodonov, AF ;
Standing, KG .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1997, 124 (01) :112-119
[38]   Simulation-based optimization of the electrodynamic ion funnel for high sensitivity electrospray ionization mass spectrometry [J].
Tolmachev, AV ;
Kim, T ;
Udseth, HR ;
Smith, RD ;
Bailey, TH ;
Futrell, JH .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2000, 203 (1-3) :31-47
[39]   ELECTROSPRAY IONIZATION COMBINED WITH ION TRAP MASS-SPECTROMETRY [J].
VANBERKEL, GJ ;
GLISH, GL ;
MCLUCKEY, SA .
ANALYTICAL CHEMISTRY, 1990, 62 (13) :1284-1295
[40]   REFLECTING TIME-OF-FLIGHT MASS-SPECTROMETER WITH AN ELECTROSPRAY ION-SOURCE AND ORTHOGONAL EXTRACTION [J].
VERENTCHIKOV, AN ;
ENS, W ;
STANDING, KG .
ANALYTICAL CHEMISTRY, 1994, 66 (01) :126-133