Ion funnel trap interface for orthogonal time-of-flight mass spectrometry

被引:64
作者
Ibrahim, Yehia [1 ]
Belov, Mikhail E. [1 ]
Tolmachev, Aleksey V. [1 ]
Prior, David C. [1 ]
Smith, Richard D. [1 ]
机构
[1] Pacific NW Natl Lab, Div Biol Sci, Richland, WA 99352 USA
关键词
D O I
10.1021/ac071091m
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A combined electrodynamic ion funnel and ion trap coupled to an orthogonal acceleration (oa)-time-of-flight mass spectrometer was developed and characterized. The ion trap was incorporated through the use of added terminal electrodynamic ion funnel electrodes enabling control over the axial dc gradient in the trap section. The ion trap operates efficiently at a pressure of similar to 1 Torr, and measurements indicate a maximum charge capacity of similar to 3 x 10(7) charges. An order of magnitude increase in sensitivity was observed in the analysis of low concentration peptides mixtures with orthogonal acceleration (oa)time-of-flight mass spectrometry (oa-TOF MS) in the trapping mode as compared to the continuous regime. A signal increase in the trapping mode was accompanied by reduction in the chemical background, due to more efficient desolvation of, for example, solvent related clusters. Controlling the ion trap ejection time was found to result in efficient removal of singly charged species and improving signal-to-noise ratio (S/N) for the multiply charged analytes.
引用
收藏
页码:7845 / 7852
页数:8
相关论文
共 41 条
[1]   Zeptomole-sensitivity electrospray ionization - Fourier transform ion cyclotron resonance mass spectrometry of proteins [J].
Belov, ME ;
Gorshkov, MV ;
Udseth, HR ;
Anderson, GA ;
Smith, RD .
ANALYTICAL CHEMISTRY, 2000, 72 (10) :2271-2279
[2]   Ion discrimination during ion accumulation in a quadrupole interface external to a Fourier transform ion cyclotron resonance mass spectrometer [J].
Belov, ME ;
Nikolaev, EN ;
Harkewicz, R ;
Masselon, CD ;
Alving, K ;
Smith, RD .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2001, 208 (1-3) :205-225
[3]  
Campbell JM, 1998, RAPID COMMUN MASS SP, V12, P1463, DOI 10.1002/(SICI)1097-0231(19981030)12:20<1463::AID-RCM357>3.3.CO
[4]  
2-8
[5]   Practical implications of some recent studies in electrospray ionization fundamentals [J].
Cech, NB ;
Enke, CG .
MASS SPECTROMETRY REVIEWS, 2001, 20 (06) :362-387
[6]   Charge state separation for protein applications using a quadrupole time-of-flight mass spectrometer [J].
Chernushevich, IV ;
Fell, LM ;
Bloomfield, N ;
Metalnikov, PS ;
Loboda, AV .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2003, 17 (13) :1416-1424
[7]   Orthogonal-injection TOFMS for analyzing biomolecules [J].
Chernushevich, IV ;
Ens, W ;
Standing, KG .
ANALYTICAL CHEMISTRY, 1999, 71 (13) :452A-461A
[8]  
CHERNUSHEVICH IV, 2003, Patent No. 6507019
[9]   A combined linear ion trap time-of-flight system with improved performance and MSn capabilities [J].
Collings, BA ;
Campbell, JM ;
Mao, DM ;
Douglas, DJ .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2001, 15 (19) :1777-1795
[10]  
Dodonov A, 1997, RAPID COMMUN MASS SP, V11, P1649, DOI 10.1002/(SICI)1097-0231(19971015)11:15<1649::AID-RCM67>3.0.CO