Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the MKEH, Hungary and the BIPM in medium-energy x-rays
被引:0
|作者:
Burns, D. T.
[1
]
Csete, I.
[2
]
Roger, P.
[1
]
机构:
[1] BIPM, F-92310 Sevres, France
[2] Hungarian Trade Licensing Off MKEH, Budapest, Hungary
来源:
关键词:
D O I:
10.1088/0026-1394/48/1A/06017
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
A key comparison has been made between the air-kerma standards of the MKEH, Hungary and the BIPM in the medium-energy x-ray range. The results show the standards to be in agreement within the standard uncertainty for the comparison of 3.4 parts in 10(3). No significant trend is observed in the results for the different radiation qualities. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database.