Optimizing the Minimum Detectable Difference of Gamma Camera SPECT Images via the Taguchi Analysis: A Feasibility Study with a V-Shaped Slit Gauge

被引:4
作者
Ke, Ching-Hsiu [1 ,2 ]
Liu, Wan-Ju [1 ,3 ]
Peng, Bing-Ru [1 ,4 ]
Pan, Lung-Fa [1 ,5 ]
Pan, Lung-Kwang [1 ]
机构
[1] Cent Taiwan Univ Sci & Technol, Dept Med Imaging & Radiol Sci, Taichung 406, Taiwan
[2] Cent Taiwan Univ Sci & Technol, Dept Optometry, Taichung 406, Taiwan
[3] Taichung Tzu Chi Hosp, Buddhist Tzu Chi Med Fdn, Dept Nucl Med, Taichung 427, Taiwan
[4] Taichung Armed Forces Gen Hosp, Dept Radiol, Taichung 411, Taiwan
[5] Taichung Armed Forces Gen Hosp, Dept Cardiol, Taichung 411, Taiwan
来源
APPLIED SCIENCES-BASEL | 2022年 / 12卷 / 05期
关键词
SPECT; gamma camera; Tc-99m MDP; spatial resolution; minimum detectable difference; OPTIMIZATION; QUALITY;
D O I
10.3390/app12052708
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
This study tried to propose an innovated idea of solidifying the resolution of gamma camera in routine quality control and recommended a quantified index as minimum detectable difference (MDD) of gamma camera SPECT images using the Taguchi analysis and an indigenous V-shaped slit gauge. The gauge was customized to fulfill the quantitative requirement of the Taguchi analysis. The MDD among slit gauge of derived SPECT image was calculated from two overlapped peak profiles collected from a tangent slice of the V-shaped slit gauge with two nearby peaks. In particular, MDD was evaluated as minimum distance between two peak centers through the Student's t-test with a constant, 1.96, which indicates that two peak centers separated distant enough to create a 95% confidence level of separation. Eighteen combinations of six gamma camera scanned factors were organized according to Taguchi analysis. Accordingly, (A) collimator, (B) detector to targe distance, (C) total counts, (D) acquired energy width, (E) Matrix size, and (F) zoom of collected ROI with each of two or three levels were organized into 18 groups to collect the slit gauge images according to Taguchi L-18 orthogonal array. Then, three well-trained radiologists were ranked the scanned gauge images to derive the fish-bone-plot of signal-to-noise ratio (S/N, dB) and correlated ANOVA. Furthermore, the quantified MDD was proposed to verify the optimal suggestion of gamma camera scanned protocol, and obtained the MDD as 8.4, 7.9, and 7.1 mm for the second group of original L-18 preset, conventional, and the optimal preset, respectively. Thus, the optimal preset of gamma camera was achieved in this study. The MDD proved to be a successful index in quantifying the imaging resolution of a gamma camera.
引用
收藏
页数:16
相关论文
共 25 条
[1]   The role of 18F-NaF PET/CT in metastatic bone disease [J].
Araz, Mine ;
Aras, Gulseren ;
Kucuk, Ozlem N. .
JOURNAL OF BONE ONCOLOGY, 2015, 4 (03) :92-97
[2]   Optimizing the TLD-100 readout system for various radiotherapy beam doses using the Taguchi methodology [J].
Chen, Chien Yi ;
Liu, Kuo Chi ;
Chen, Hsin Hsiung ;
Pan, Lung Kwang .
APPLIED RADIATION AND ISOTOPES, 2010, 68 (03) :481-488
[3]   MINIMUM DETECTABLE DIFFERENCE OF CT ANGIOGRAPHY SCANS AT VARIOUS CARDIAC BEATS: EVALUATION VIA A CUSTOMIZED OBLIQUE V-SHAPED LINE GAUGE AND PMMA PHANTOM [J].
Chiang, Chao-Yu ;
Chen, Yi-Hua ;
Pan, Lung-Fa ;
Cho, Chien-Chou ;
Peng, Bing-Ru ;
Pan, Lung-Kwang .
JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2021, 21 (10)
[4]  
Das D., 2014, IOSR J PHARM, V4, P56
[5]  
Dickerscheid D, 2013, EJNMMI RES, V3, DOI 10.1186/2191-219X-3-21
[6]   Proof of concept for low-dose molecular breast imaging with a dual-head CZT gamma camera. Part I. Evaluation in phantoms [J].
Hruska, Carrie B. ;
Weinmann, Amanda L. ;
O'Connor, Michael K. .
MEDICAL PHYSICS, 2012, 39 (06) :3466-3475
[7]   Prospective Evaluation of 99mTc MDP Scintigraphy, 18F NaF PET/CT, and 18F FDG PET/CT for Detection of Skeletal Metastases [J].
Iagaru, Andrei ;
Mittra, Erik ;
Dick, David W. ;
Gambhir, Sanjiv Sam .
MOLECULAR IMAGING AND BIOLOGY, 2012, 14 (02) :252-259
[8]   Rare splenic metastasis of renal cell carcinoma detected on Tc-99m-MDP bone scan [J].
Jatale, Archi Agrawal Prafulla ;
Purandare, Nilendu ;
Shah, Sneha ;
Rangarajan, Venkatesh .
INDIAN JOURNAL OF NUCLEAR MEDICINE, 2014, 29 (01) :60-61
[9]   The statistical analysis of data from small groups [J].
Kenny, DA ;
Mannetti, L ;
Pierro, A ;
Livi, S ;
Kashy, DA .
JOURNAL OF PERSONALITY AND SOCIAL PSYCHOLOGY, 2002, 83 (01) :126-137
[10]   The Optimization of the Single Photon Emission Computed Tomography Image Quality via Taguchi Analysis: A Feasibility Study of a V -Shaped Phantom [J].
Kittipayak, Samrit ;
Pan, Lung-Fa ;
Chiang, Fu-Tsai ;
Pan, Lung-Kwang ;
Lin, Cheng-Hsun .
JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2017, 7 (01) :143-148