Musculoskeletal imaging with a prototype photon-counting detector

被引:1
作者
Gruber, M. [1 ]
Homolka, P. [2 ]
Chmeissani, M. [3 ]
Uffmann, M. [1 ]
Pretterklieber, M. [4 ]
Kainberger, F. [1 ]
机构
[1] Med Univ Vienna, Dept Radiol, AT-1090 Vienna, Austria
[2] Med Univ Vienna, Ctr Med Phys & Biomed Engn, AT-1090 Vienna, Austria
[3] Inst High Energy Phys, Barcelona, Spain
[4] Med Univ Vienna, Ctr Anat & Cell Biol, AT-1090 Vienna, Austria
关键词
Photon-counting detector; Digital imaging; Low dose; Direct radiography; Prototype X-ray detector; COMPUTED-TOMOGRAPHY; MEDIPIX2; SYSTEM;
D O I
10.1007/s00330-011-2246-8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To test a digital imaging X-ray device based on the direct capture of X-ray photons with pixel detectors, which are coupled with photon-counting readout electronics. The chip consists of a matrix of 256 x 256 pixels with a pixel pitch of 55 mu m. A monolithic image of 11.2 cm x 7 cm was obtained by the consecutive displacement approach. Images of embalmed anatomical specimens of eight human hands were obtained at four different dose levels (skin dose 2.4, 6, 12, 25 mu Gy) with the new detector, as well as with a flat-panel detector. The overall rating scores for the evaluated anatomical regions ranged from 5.23 at the lowest dose level, 6.32 at approximately 6 mu Gy, 6.70 at 12 mu Gy, to 6.99 at the highest dose level with the photon-counting system. The corresponding rating scores for the flat-panel detector were 3.84, 5.39, 6.64, and 7.34. When images obtained at the same dose were compared, the new system outperformed the conventional DR system at the two lowest dose levels. At the higher dose levels, there were no significant differences between the two systems. The photon-counting detector has great potential to obtain musculoskeletal images of excellent quality at very low dose levels. Novel detectors offer further opportunities to limit radiation dose in radiology. A prototype photon-counting detector has recently been introduced for mammography. This detector has now been shown to have potential for musculoskeletal imaging. It can provide high quality musculoskeletal images at very low dose levels.
引用
收藏
页码:205 / 210
页数:6
相关论文
共 16 条
[1]   MediSPECT: Single photon emission computed tomography system for small field of view small animal imaging based on a CdTe hybrid pixel detector [J].
Accorsi, R. ;
Autiero, M. ;
Celentano, L. ;
Chmeissani, M. ;
Cozzolino, R. ;
Curion, A. S. ;
Frallicciardi, P. ;
Laccetti, P. ;
Lanza, R. C. ;
Lauria, A. ;
Maiorino, M. ;
Marotta, M. ;
Mettivier, G. ;
Montesi, M. C. ;
Riccio, P. ;
Roberti, G. ;
Russo, P. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2) :44-47
[2]  
[Anonymous], 2008, WORLD MED ASS DECL H
[3]   A pixel detector-based single photon-counting system as fast spectrometer for diagnostic X-ray beams [J].
Carpentieri, C. ;
Bisogni, M. G. ;
Del Guerra, A. ;
Delogu, P. ;
Fantacci, M. E. ;
Fogli, J. ;
Marchi, A. ;
Marzulli, V. ;
Rosso, V. ;
Stefanini, A. ;
Tofani, A. .
RADIATION PROTECTION DOSIMETRY, 2008, 129 (1-3) :119-122
[4]   First experimental tests with a CdTe photon counting pixel detector hybridized with a Medipix2 readout chip [J].
Chmeissani, M ;
Frojdh, C ;
Gal, O ;
Llopart, X ;
Ludwig, J ;
Maiorino, M ;
Manach, E ;
Mettivier, G ;
Montesi, MC ;
Ponchut, C ;
Russo, P ;
Tlustos, L ;
Zwerger, A .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (05) :2379-2385
[5]  
Chmeissani M., 2004, Proceedings of the 21st IEEE Instrumentation and Measurement Technology Conference (IEEE Cat. No.04CH37510), P787, DOI 10.1109/IMTC.2004.1351164
[6]   Characterization of a pixellated CdTe detector with single-photon processing readout [J].
Fröjdh, C ;
Graafsma, H ;
Nilsson, HE ;
Ponchut, C .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 563 (01) :128-132
[7]   Average glandular dose in routine mammography screening using a Sectra MicroDose Mammography unit [J].
Hemdal, B ;
Herrnsdorf, L ;
Andersson, I ;
Bengtsson, G ;
Heddson, B ;
Olsson, M .
RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) :436-443
[8]  
*INT COMM RAD PROT, 2004, ANN IRCP, V34
[9]   Photon Counting Energy Dispersive Detector Arrays for X-ray Imaging [J].
Iwanczyk, Jan S. ;
Nygard, Einar ;
Meirav, Oded ;
Arenson, Jerry ;
Barber, William C. ;
Hartsough, Neal E. ;
Malakhov, Nail ;
Wessel, Jan C. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (03) :535-542
[10]  
JOHNS PC, 2002, P SPIE TD, V1, P367