Imaging Facilities' Adherence to PI-RADS v2 Minimum Technical Standards for the Performance of Prostate MRI

被引:66
作者
Esses, Steven J. [1 ]
Taneja, Samir S. [2 ]
Rosenkrantz, Andrew B. [1 ]
机构
[1] NYU Langone Med Ctr, NYU Sch Med, Ctr Biomed Imaging, Dept Radiol, 660 First Ave,3rd Floor, New York, NY 10016 USA
[2] NYU Langone Med Ctr, Dept Urol Oncol, New York, NY USA
关键词
Prostate cancer; MRI; PI-RADS; diffusion-weighted imaging; dynamic contrast-enhanced imaging; DIFFUSION-WEIGHTED MRI; CONTRAST-ENHANCED MRI; TEMPORAL RESOLUTION; VERSION; CANCER; SYSTEM; EPI;
D O I
10.1016/j.acra.2017.08.013
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: This study aimed to assess variability in imaging facilities' adherence to the minimum technical standards for prostate magnetic resonance imaging acquisition established by Prostate Imaging -Reporting and Data System (PI-RADS) version 2 (v2). Methods: A total of 107 prostate magnetic resonance imaging examinations performed at 107 unique imaging facilities after the release of PI-RADS v2 and that were referred to a tertiary care center for secondary interpretation were included. Image sets, DICOM headers, and outside reports were reviewed to assess adherence to 21 selected PI-RADS v2 minimum technical standards. Results: Hardware arrangements were 23.1%, 1.5T without endorectal coil; 7.7%, 1.5T with endorectal coil; 63.5%, 3T without endorectal coil; and 5.8%, 3T with endorectal coil. Adherence to minimum standards was lowest on T2 weighted imaging (T2WI) for frequency resolution 50.4 mm (16.8%) and phase resolution <= 0.7 mm (48.6%), lowest on diffusion-weighted imaging (DWI) for field of view (FOV) 120-220 mm (30.0%), and lowest on dynamic contrast-enhanced (DCE) imaging for slice thickness 3 mm (33.3%) and temporal resolution <10 s (31.5%). High b-value (>= 1400 s/mm(2)) images were included in 58.0% (calculated in 25.9%). Adherence to T2WI phase resolution and DWI inter-slice gap were greater (P <.05) at 3T than at 1.5T. Adherence did not differ (P >.05) for any parameter between examinations performed with and without an endorectal coil. Adherence was greater for examinations performed at teaching facilities for T2WI slice thickness and DCE temporal resolution (P <.05). Adherence was not better for examinations performed in 2016 than in 2015 for any parameter (P >.05). Conclusion: Facilities' adherence to PI-RADS v2 minimum technical standards was variable, being particularly poor for T2WI frequency resolution and DCE temporal resolution. The standards warrant greater community education. Certain technical standards may be too stringent, and revisions should be considered.
引用
收藏
页码:188 / 195
页数:8
相关论文
共 20 条
[1]  
Accreditation Council for Graduate Medical Education, RAD DIAGN PROGR AC Y
[2]   Small Field-of-view single-shot EPI-DWI of the prostate: Evaluation of spatially-tailored two-dimensional radiofrequency excitation pulses [J].
Attenberger, Ulrike I. ;
Rathmann, Nils ;
Sertdemir, Metin ;
Riffel, Philipp ;
Weidner, Anja ;
Kannengiesser, Stefan ;
Morelli, John N. ;
Schoenberg, Stefan O. ;
Hausmann, Daniel .
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2016, 26 (02) :168-176
[3]   CARCINOMA OF THE PROSTATE-GLAND - MR-IMAGING WITH PELVIC PHASED-ARRAY COILS VERSUS INTEGRATED ENDORECTAL PELVIC PHASED-ARRAY COILS [J].
HRICAK, H ;
WHITE, S ;
VIGNERON, D ;
KURHANEWICZ, J ;
KOSCO, A ;
LEVIN, D ;
WEISS, J ;
NARAYAN, P ;
CARROLL, PR .
RADIOLOGY, 1994, 193 (03) :703-709
[4]  
Kim SH, 2017, AM J ROENTGENOL, P1
[5]   Clinical utility of apparent diffusion coefficient values obtained using high b-value when diagnosing prostate cancer using 3 tesla MRI: Comparison between ultra-high b-value (2000 s/mm2) and standard high b-value (1000 s/mm2) [J].
Kitajima, Kazuhiro ;
Takahashi, Satoru ;
Ueno, Yoshiko ;
Yoshikawa, Takeshi ;
Ohno, Yoshiharu ;
Obara, Makoto ;
Miyake, Hideaki ;
Fujisawa, Masato ;
Sugimura, Kazuro .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2012, 36 (01) :198-205
[6]   Reduced-FOV excitation decreases susceptibility artifact in diffusion-weighted MRI with endorectal coil for prostate cancer detection [J].
Korn, Natalie ;
Kurhanewicz, John ;
Banerjee, Suchandrima ;
Starobinets, Olga ;
Saritas, Emine ;
Noworolski, Susan .
MAGNETIC RESONANCE IMAGING, 2015, 33 (01) :56-62
[7]   High-Resolution Diffusion-Weighted Imaging of the Prostate [J].
Medved, Milica ;
Soylu-Boy, Fatma N. ;
Karademir, Ibrahim ;
Sethi, Ila ;
Yousuf, Ambereen ;
Karczmar, Gregory S. ;
Oto, Aytekin .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 203 (01) :85-90
[8]  
Missouri Census Data Center, MET DAT PUB DAT GEOR
[9]   Effect of Temporal Resolution on Diagnostic Performance of Dynamic Contrast-Enhanced Magnetic Resonance Imaging of the Prostate [J].
Othman, Ahmed E. ;
Falkner, Florian ;
Weiss, Jakob ;
Kruck, Stephan ;
Grimm, Robert ;
Martirosian, Petros ;
Nikolaou, Konstantin ;
Notohamiprodjo, Mike .
INVESTIGATIVE RADIOLOGY, 2016, 51 (05) :290-296
[10]   Accuracy and Interobserver Agreement for Prostate Imaging Reporting and Data System, Version 2, for the Characterization of Lesions Identified on Multiparametric MRI of the Prostate [J].
Purysko, Andrei S. ;
Bittencourt, Leonardo K. ;
Bullen, Jennifer A. ;
Mostardeiro, Thomaz R. ;
Herts, Brian R. ;
Klein, Eric A. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2017, 209 (02) :339-345