Reliability of intraocular pressure measurements after myopic excimer photorefractive keratectomy

被引:56
作者
Abbasoglu, KE [1 ]
Bowman, RW [1 ]
Cavanagh, HD [1 ]
McCulley, JP [1 ]
机构
[1] Univ Texas, SW Med Ctr, Dept Ophthalmol, Dallas, TX 75235 USA
关键词
D O I
10.1016/S0161-6420(98)91215-5
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objective: To determine the reliability of intraocular pressure (IOP) measurements by Goldmann applanation tonometry and pneumotonometry in eyes treated with excimer myopic photorefractive keratectomy (PRK), Design: A prospective case series. Participants: Forty consecutive eyes treated with PRK were evaluated. Intervention and Main Outcome Measures: Central and peripheral corneal Goldmann tonometry and pneumotonometry measurements were done before surgery, at 1 week, and at I and 3 months after surgery. Results: The IOP by Goldmann tonometry from the central cornea was significantly lower than the peripheral IOP; however, there was no difference between IOP measured from central and peripheral corneas by pneumotonometry, which, in turn, correlated with peripheral Goldmann measurements. There was a trend, but not a statistically significant correlation, between the spherical equivalent of the treatment and the amount of decrease in central Goldmann IOP. Conclusions: Pneumotonometry measures the IOP reliably after PRK from all parts of the cornea, whereas central Goldmann tonometry underestimates the IOP by 2.40 +/- 1.23 mmHg.