New filtered-backprojection-based algorithms for image reconstruction in fan-beam scans

被引:0
作者
Xia, D [1 ]
Zou, Y [1 ]
Yu, LF [1 ]
Pan, XC [1 ]
机构
[1] Univ Chicago, Dept Radiol, Chicago, IL 60637 USA
来源
2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7 | 2004年
关键词
computed tomography (CT); image reconstruction; fan-beam;
D O I
暂无
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Fan-beam scan is a configuration widely used in clinical computed tomography (CT) because it is easy to implement and control. Filtered-backprojection(FBP)-based algorithms have been developed previously for exact reconstruction of the entire image or an ROI within the image from the data acuqired in fan-beam scans. Recently, we have developed a 3D FBP-based algorithm for image reconstruction on PI-line segments in a helical cone-beam scan. In this work, we demonstrated that the 3D FBP-based algorithm indeed provided a rather general FBP-based formulation for image reconstruction. Based upon this formulation we derived new FBP-based algorithms for image reconstruction in the fan-beam scans, which can be interpreted as the special case of the helical scan. Existing algorithms corresponding to the new derived algorithms were identified. We also performed a preliminary numerical study to verify our theoretical results. The results in the work can readily be generalized to other non-circular trajectories.
引用
收藏
页码:2530 / 2533
页数:4
相关论文
共 8 条
[1]  
Kak A.C. Slaney M., 1999, PRINCIPLES COMPUTERI
[2]   Image reconstruction from fan-beam projections on less than a short scan [J].
Noo, F ;
Defrise, M ;
Clackdoyle, R ;
Kudo, H .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (14) :2525-2546
[3]   Optimal noise control in and fast reconstruction of fan-beam computed tomography image [J].
Pan, XC .
MEDICAL PHYSICS, 1999, 26 (05) :689-697
[4]   Image reconstruction with shift-variant filtration and its implication for noise and resolution properties in fan-beam computed tomography [J].
Pan, XC ;
Yu, LF .
MEDICAL PHYSICS, 2003, 30 (04) :590-600
[5]   OPTIMAL SHORT SCAN CONVOLUTION RECONSTRUCTION FOR FANBEAM CT [J].
PARKER, DL .
MEDICAL PHYSICS, 1982, 9 (02) :254-257
[6]   Exact image reconstruction on PI-lines from minimum data in helical cone-beam CT [J].
Zou, Y ;
Pan, XC .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (06) :941-959
[7]   Image reconstruction on PI-lines by use of filtered backprojection in helical cone-beam CT [J].
Zou, Y ;
Pan, XC .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (12) :2717-2731
[8]  
ZOU Z, 2004, UNPUB PHSY MED BIOL