Long-term Field Evaluation of Low-cost Particulate Matter Sensors in Nanjing

被引:39
|
作者
Bai, Lu [1 ]
Huang, Lin [1 ]
Wang, Zhenglu [2 ]
Ying, Qi [1 ,2 ]
Zheng, Jun [1 ]
Shi, Xiaowen [1 ]
Hu, Jianlin [1 ]
机构
[1] Nanjing Univ Informat Sci & Technol, Jiangsu Key Lab Atmospher Environm Monitoring & P, Jiangsu Collaborat Innovat Ctr Atmospher Environm, Nanjing 210044, Peoples R China
[2] Texas A&M Univ, Zachry Dept Civil Engn, College Stn, TX 77843 USA
基金
中国国家自然科学基金;
关键词
PPD42NS; Field calibration; Long-term; Artificial neural network; NEURAL-NETWORK MODEL; YANGTZE-RIVER DELTA; AIR-POLLUTION; PREMATURE MORTALITY; PM2.5; CHINA; PM10; EXPOSURE; CITIES; RISK;
D O I
10.4209/aaqr.2018.11.0424
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Low-cost particulate matter (PM) sensors can be widely deployed to measure aerosol concentrations at higher spatial and temporal resolutions than traditional instruments, but they need to be carefully calibrated under ambient conditions. In this study, a long-term field experiment was conducted from December 2015 to May 2017 at a site in Nanjing to evaluate the capabilities of in-house built low-cost PM monitors using the Shinyei PPD42NS sensor for ambient PM2.5 monitoring. A BAM-1020 particulate monitor was co-located with the low-cost sensors to provide reference readings. Least-square regressions with linear and power-law functions, and an artificial neural network (ANN) technique were used to convert electrical instrument readings to ambient aerosol concentrations. Applying the ANN technique resulted in the best estimation of the hourly PM2.5 (R-2 = 0.84; mean normalized bias = 12.7% and mean normalized error (MNE) = 29.7%). The low-cost sensors displayed relatively good performance with high aerosol concentrations but larger errors with concentrations below 35 mu g m(-3). High humidity (RH > 75%) can cause a larger MNE for these sensors, but the impact of temperature was negligible in this study. A clear sensor deterioration trend was observed during the 18-month field calibration. High correlations were found between the data from a single low-cost sensor and the data from the BAM-1020 when the low-cost sensor was individually calibrated, but the correlations between measurements taken by different low-cost sensor units were only moderate, possibly due to internal sensor variations. The results suggest that these low-cost sensors can measure ambient PM2.5 concentrations with an acceptable level of accuracy, which can and should be improved by calibrating each sensor individually. Special attention should be paid to the accuracy of these sensors after long-term application and in highly humid environments.
引用
收藏
页码:242 / 253
页数:12
相关论文
共 50 条
  • [1] Long-term field evaluation of the Plantower PMS low-cost particulate matter sensors
    Sayahi, T.
    Butterfield, A.
    Kelly, K. E.
    ENVIRONMENTAL POLLUTION, 2019, 245 : 932 - 940
  • [2] Feasibility of low-cost particulate matter sensors for long-term environmental monitoring: Field evaluation and calibration
    Feng Z.
    Zheng L.
    Ren B.
    Liu D.
    Huang J.
    Xue N.
    Science of the Total Environment, 2024, 945
  • [3] Field Evaluation of Low-Cost Particulate Matter Sensors in Beijing
    Mei, Han
    Han, Pengfei
    Wang, Yinan
    Zeng, Ning
    Liu, Di
    Cai, Qixiang
    Deng, Zhaoze
    Wang, Yinghong
    Pan, Yuepeng
    Tang, Xiao
    SENSORS, 2020, 20 (16) : 1 - 17
  • [4] Long-term field comparison of multiple low-cost particulate matter sensors in an outdoor urban environment
    Bulot, Florentin M. J.
    Johnston, Steven J.
    Basford, Philip J.
    Easton, Natasha H. C.
    Apetroaie-Cristea, Mihaela
    Foster, Gavin L.
    Morris, Andrew K. R.
    Cox, Simon J.
    Loxham, Matthew
    SCIENTIFIC REPORTS, 2019, 9 (1)
  • [5] Long-term field comparison of multiple low-cost particulate matter sensors in an outdoor urban environment
    Florentin M. J. Bulot
    Steven J. Johnston
    Philip J. Basford
    Natasha H. C. Easton
    Mihaela Apetroaie-Cristea
    Gavin L. Foster
    Andrew K. R. Morris
    Simon J. Cox
    Matthew Loxham
    Scientific Reports, 9
  • [6] Long-Term Evaluation and Calibration of Low-Cost Particulate Matter (PM) Sensor
    Lee, Hoochang
    Kang, Jiseock
    Kim, Sungjung
    Im, Yunseok
    Yoo, Seungsung
    Lee, Dongjun
    SENSORS, 2020, 20 (13) : 1 - 25
  • [7] Laboratory and field evaluation of three low-cost particulate matter sensors
    Ghamari, Mohammad
    Soltanpur, Cinna
    Rangel, Pablo
    Groves, William A.
    Kecojevic, Vladislav
    IET WIRELESS SENSOR SYSTEMS, 2022, 12 (01) : 21 - 32
  • [8] Evaluation methods for low-cost particulate matter sensors
    Bean, Jeffrey K.
    ATMOSPHERIC MEASUREMENT TECHNIQUES, 2021, 14 (11) : 7369 - 7379
  • [9] Performance Evaluation of Low-cost Particulate Matter Sensors
    Marinov, Marin B.
    Hensel, Stefan
    Ganev, Borislav
    Nikolov, Georgi
    2017 XXVI INTERNATIONAL SCIENTIFIC CONFERENCE ELECTRONICS (ET), 2017,
  • [10] Author Correction: Long-term field comparison of multiple low-cost particulate matter sensors in an outdoor urban environment
    Florentin M. J. Bulot
    Steven J. Johnston
    Philip J. Basford
    Natasha H. C. Easton
    Mihaela Apetroaie-Cristea
    Gavin L. Foster
    Andrew K. R. Morris
    Simon J. Cox
    Matthew Loxham
    Scientific Reports, 10