Evaluation of matrix effects in analysis of estrogen using liquid chromatography-tandem mass spectrometry

被引:27
作者
Nguyen, Hien P. [1 ]
Li, Li [1 ]
Nethrapalli, Imam S. [2 ]
Guo, Ningning [2 ]
Toran-Allerand, C. Dominique [2 ]
Harrison, David E. [3 ]
Astle, C. Michael [3 ]
Schug, Kevin A. [1 ]
机构
[1] Univ Texas Arlington, Dept Chem & Biochem, Arlington, TX 76019 USA
[2] Columbia Univ Coll Phys & Surg, Dept Pathol & Cell Biol, New York, NY 10032 USA
[3] Jackson Lab, Bar Harbor, ME 04609 USA
基金
美国国家科学基金会;
关键词
Biological samples; Estradiol; Matrix effects; Electrospray; LC-MS/MS; LC-MS/MS; DERIVATIZATION; IONIZATION; MORTALITY; ESTRADIOL; ENHANCE; PLASMA; SERUM; ASSAY;
D O I
10.1002/jssc.201100219
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Matrix effects of different biological samples, including phosphate-buffered saline bovine serum albumin (PBS-BSA), gelded horse serum, mouse serum, and mouse brain, were investigated for the determination of 17 alpha- and beta-estradiol using derivatization with dansyl chloride prior to LC-MS/MS. Matrix effects were evaluated based on the slopes of regression lines plotted from results obtained in biological matrices versus pure standard solutions. Such plots indicate the enhancement or suppression of signal based on the presence of a particular biological fluid for a particular method. The matrix effects from PBS-BSA were similar to those of mouse serum. In contrast, analyses performed from horse serum and mouse brain yielded significant ion suppression, especially for 17 beta-estradiol. Precipitation during derivatization was observed when pre-concentrated samples were processed with ethyl acetate as an extraction solvent. This was overcome with the use of methyl tert-butyl ether; however, matrix effects from this preparation were still present, evidenced by signal suppression and poor linearity in the standard curve. This work affirms that caution should be taken in the transfer of methods for use with different biological matrices, especially in the case where surrogate matrices are necessary for calibration purposes.
引用
收藏
页码:1781 / 1787
页数:7
相关论文
共 38 条
  • [1] Derivatization of ethinylestradiol with dansyl chloride to enhance electrospray ionization: Application in trace analysis of ethinylestradiol in rhesus monkey plasma
    Anari, MR
    Bakhtiar, R
    Zhu, B
    Huskey, S
    Franklin, RB
    Evans, DC
    [J]. ANALYTICAL CHEMISTRY, 2002, 74 (16) : 4136 - 4144
  • [2] [Anonymous], 1993, Statistics for Analytical Chemistry
  • [3] ASHWOOD ER, 2005, TIETZ TXB CLIN CHEM, P2251
  • [4] Bakhtiar Ray, 2007, Journal of Pharmacological and Toxicological Methods, V55, P262, DOI 10.1016/j.vascn.2006.10.002
  • [5] BENNETTE P, 2004, P AM SOC MASS SPECTR
  • [6] Estradiol protects against ischemic injury
    Dubal, DB
    Kashon, ML
    Pettigrew, LC
    Ren, JM
    Finklestein, SP
    Rau, SW
    Wise, PM
    [J]. JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1998, 18 (11) : 1253 - 1258
  • [7] Fishman R.A., 1980, Cerebrospinal fluid in diseases of the nervous system
  • [8] Measurement of endogenous estrogens: analytical challenges and recent advances
    Giese, RW
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2003, 1000 (1-2) : 401 - 412
  • [9] HORMONE-THERAPY TO PREVENT DISEASE AND PROLONG LIFE IN POSTMENOPAUSAL WOMEN
    GRADY, D
    RUBIN, SM
    PETITTI, DB
    FOX, CS
    BLACK, D
    ETTINGER, B
    ERNSTER, VL
    CUMMINGS, SR
    [J]. ANNALS OF INTERNAL MEDICINE, 1992, 117 (12) : 1016 - 1037
  • [10] NUTRITIONAL COMPOSITION OF MILK PRODUCED BY MOTHERS DELIVERING PRETERM
    GROSS, SJ
    DAVID, RJ
    BAUMAN, L
    TOMARELLI, RM
    [J]. JOURNAL OF PEDIATRICS, 1980, 96 (04) : 641 - 644