Feasibility study of portable multi-energy computed tomography with photon-counting detector for preclinical and clinical applications

被引:8
作者
Lee, Chang-Lae [1 ]
Hong, Key Jo [1 ]
Kim, Namwoo [1 ]
Han, Kwanhee [1 ]
Kim, Dongkyu [2 ]
Jung, Hoe-Su [2 ]
Lee, Sangmin [1 ]
Park, Junyoung [1 ]
Lee, Kyoung-Yong [1 ]
Lee, Jee Eun [3 ]
Choi, Yuna [1 ]
Cho, Minkook [1 ]
机构
[1] Samsung Elect, Hlth & Med Equipment Business Unit, PCD R&D Grp, Suwon, Gyeonggido, South Korea
[2] Daegu Gyeongbuk Med Innovat Fdn, Lab Anim Ctr, Daegu, South Korea
[3] Bupyeong Serim Hosp, Dept Family Med, Incheon, South Korea
关键词
CEREBROSPINAL-FLUID; CT; STROKE; PERFUSION; MATTER; WHITE; MRI;
D O I
10.1038/s41598-021-02210-5
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
In this study, preclinical experiments were performed with an in-house developed prototypal photon-counting detector computed tomography (PCD CT) system. The performance of the system was compared with the conventional energy-integrating detector (EID)-based CT, concerning the basic image quality biomarkers and the respective capacities for material separation. The pre- and the post-contrast axial images of a canine brain captured by the PCD CT and EID CT systems were found to be visually similar. Multi-energy images were acquired using the PCD CT system, and machine learning-based material decomposition was performed to segment the white and gray matters for the first time in soft tissue segmentation. Furthermore, to accommodate clinical applications that require high resolution acquisitions, a small, native, high-resolution (HR) detector was implemented on the PCD CT system, and its performance was evaluated based on animal experiments. The HR acquisition mode improved the spatial resolution and delineation of the fine structures in the canine's nasal turbinates compared to the standard mode. Clinical applications that rely on high-spatial resolution expectedly will also benefit from this resolution-enhancing function. The results demonstrate the potential impact on the brain tissue segmentation, improved detection of the liver tumors, and capacity to reconstruct high-resolution images both preclinically and clinically.
引用
收藏
页数:11
相关论文
共 33 条
[1]   Magnetic resonance imaging of acute stroke [J].
Baird, AE ;
Warach, S .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1998, 18 (06) :583-609
[2]   Prediction of stroke outcome with echoplanar perfusion- and diffusion-weighted MRI [J].
Barber, PA ;
Darby, DG ;
Desmond, PM ;
Yang, Q ;
Gerraty, RP ;
Jolley, D ;
Donnan, GA ;
Tress, BM ;
Davis, SM .
NEUROLOGY, 1998, 51 (02) :418-426
[3]   Photon-counting normalized metal artifact reduction (NMAR) in diagnostic CT [J].
Byl, Achim ;
Klein, Laura ;
Sawall, Stefan ;
Heinze, Sarah ;
Schlemmer, Heinz-Peter ;
Kachelriess, Marc .
MEDICAL PHYSICS, 2021, 48 (07) :3572-3582
[4]   Assessing Response to Stroke Thrombolysis Validation of 24-Hour Multimodal Magnetic Resonance Imaging [J].
Campbell, Bruce C. V. ;
Tu, Hans T. H. ;
Christensen, Soren ;
Desmond, Patricia M. ;
Levi, Christopher R. ;
Bladin, Christopher F. ;
Hjort, Niels ;
Ashkanian, Mahmoud ;
Solling, Christine ;
Donnan, Geoffrey A. ;
Davis, Stephen M. ;
Ostergaard, Leif ;
Parsons, Mark W. .
ARCHIVES OF NEUROLOGY, 2012, 69 (01) :46-50
[5]   Thresholds for infarction vary between gray matter and white matter in acute ischemic stroke: A CT perfusion study [J].
Chen, Chushuang ;
Bivard, Andrew ;
Lin, Longting ;
Levi, Christopher R. ;
Spratt, Neil J. ;
Parsons, Mark W. .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2019, 39 (03) :536-546
[6]   Diagnostic accuracy of contrast-enhanced dynamic CT for small hypervascular hepatocellular carcinoma and assessment of dynamic enhancement patterns: Results of two-year follow-up using cone-beam CT hepatic arteriography [J].
Choi, Ye Ra ;
Chung, Jin Wook ;
Yu, Mi Hye ;
Lee, Myungsu ;
Kim, Jung Hoon .
PLOS ONE, 2018, 13 (09)
[7]   Resolution characterization of a silicon-based, photon-counting computed tomography prototype capable of patient scanning [J].
da Silva, Joakim ;
Gronberg, Fredrik ;
Cederstrom, Bjorn ;
Persson, Mats ;
Sjolin, Martin ;
Alagic, Zlatan ;
Bujila, Robert ;
Danielsson, Mats .
JOURNAL OF MEDICAL IMAGING, 2019, 6 (04)
[8]   COMPUTER-ASSISTED CATEGORIZATION OF BRAIN COMPUTERIZED-TOMOGRAPHY PIXELS INTO CEREBROSPINAL-FLUID, WHITE MATTER, AND GRAY-MATTER [J].
DELEO, JM ;
SCHWARTZ, M ;
CREASEY, H ;
CUTLER, N ;
RAPOPORT, SI .
COMPUTERS AND BIOMEDICAL RESEARCH, 1985, 18 (01) :79-88
[9]   Quantitative assessment of colorectal cancer tumor vascular parameters by using perfusion CT: Influence of tumor region of interest [J].
Goh, Vicky ;
Halligan, Steve ;
Gharpuray, Anita ;
Wellsted, David ;
Sundin, Josefin ;
Bartram, Clive I. .
RADIOLOGY, 2008, 247 (03) :726-732
[10]   Automatic Segmentation of Cerebrospinal Fluid, White and Gray Matter in Unenhanced Computed Tomography Images [J].
Gupta, Varsha ;
Ambrosius, Wojciech ;
Qian, Guoyu ;
Blazejewska, Anna ;
Kazmierski, Radoslaw ;
Urbanik, Andrzej ;
Nowinski, Wieslaw L. .