Incubator-based Sound Attenuation: Active Noise Control In A Simulated Clinical Environment

被引:14
作者
Hutchinson, George [1 ]
Du, Lilin [1 ]
Ahmad, Kaashif [2 ,3 ]
机构
[1] Invictus Med Inc, San Antonio, TX 78232 USA
[2] Pediatrix Med Grp, San Antonio, TX USA
[3] Baylor Coll Med, San Antonio, TX USA
来源
PLOS ONE | 2020年 / 15卷 / 07期
基金
美国国家卫生研究院;
关键词
INTENSIVE-CARE-UNIT; SINGLE-FAMILY ROOM; RECOMMENDED STANDARDS; PRETERM INFANTS; OUTCOMES; RESPONSES; EARMUFFS; DESIGN; TALK;
D O I
10.1371/journal.pone.0235287
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Objective Noise in the neonatal intensive care unit can be detrimental to the health of the hospitalized infant. Means of reducing that noise include staff training, warning lights, and ear coverings, all of which have had limited success. Single family rooms, while an improvement, also expose the hospitalized infant to the same device alarms and mechanical noises found in open bay units. Methods We evaluated a non-contact incubator-based active noise control device (Neoasis (TM), Invictus Medical, San Antonio, Texas) in a simulated neonatal intensive care unit (NICU) setting to determine whether it could effectively reduce the noise exposure of infants within an incubator. In the NICU simulation center, we generated a series of clinically appropriate sound sequences with bedside medical devices such as a patient monitor and fluid infusion devices, hospital air handling systems, and device mechanical sounds. A microphoneequipped infant mannequin was oriented within an incubator. Measurements were made with the microphones with the Neoasis (TM) deactivated and activated. Results The active noise control device decreased sound pressure levels for certain alarm sounds by as much as 14.4 dB (a 5.2-fold reduction in sound pressure) at the alarm tone's primary frequency. Frequencies below the 2 kHz octave band were more effectively attenuated than frequencies at or above the 2 kHz octave band. Background noise levels below 40 dBA were essentially not impacted by the active noise control device. Conclusions The active noise control device further reduces noise inside infant incubators. Device safety and potential health benefits of the quieter environment should be verified in a clinical setting.
引用
收藏
页数:12
相关论文
共 42 条
  • [1] Abdeen Ziad, 2014, ScientificWorldJournal, V2014, P878704, DOI [10.1155/2014/864780, 10.1155/2014/878704]
  • [2] A randomized clinical trial evaluating silicone earplugs for very low birth weight newborns in intensive care
    Abou Turk, C.
    Williams, A. L.
    Lasky, R. E.
    [J]. JOURNAL OF PERINATOLOGY, 2009, 29 (05) : 358 - 363
  • [3] Sound reduction management in the neonatal intensive care unit for preterm or very low birth weight infants
    Almadhoob, Abdulraoof
    Ohlsson, Arne
    [J]. COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2015, (01):
  • [4] [Anonymous], 2012, S1268 ANSIASA, V2007
  • [5] Blackstock DT, 2000, FUNDAMENTALS PHYS AC, P163
  • [6] OUTCOME OF EXTREMELY LOW-BIRTH-WEIGHT INFANTS - 1980-1990
    BLAYMOREBIER, J
    PEZZULLO, J
    KIM, E
    OH, W
    GARCIACOLL, C
    VOHR, BR
    [J]. ACTA PAEDIATRICA, 1994, 83 (12) : 1244 - 1248
  • [7] Noise and the premature infant: Physiological effects and practice implications
    Bremmer, P
    Byers, JF
    Kiehl, E
    [J]. JOGNN-JOURNAL OF OBSTETRIC GYNECOLOGIC AND NEONATAL NURSING, 2003, 32 (04): : 447 - 454
  • [8] The characterization of noise levels in a neonatal intensive care unit and the implications for noise management
    Carlos Fortes-Garrido, Juan
    Mauricio Velez-Pereira, Andres
    Gazquez, Manuel
    Hidalgo-Hidalgo, Montserrat
    Pedro Bolivar, Juan
    [J]. JOURNAL OF ENVIRONMENTAL HEALTH SCIENCE AND ENGINEERING, 2014, 12
  • [9] Adult Talk in the NICU With Preterm Infants and Developmental Outcomes
    Caskey, Melinda
    Stephens, Bonnie
    Tucker, Richard
    Vohr, Betty
    [J]. PEDIATRICS, 2014, 133 (03) : E578 - E584
  • [10] Importance of Parent Talk on the Development of Preterm Infant Vocalizations
    Caskey, Melinda
    Stephens, Bonnie
    Tucker, Richard
    Vohr, Betty
    [J]. PEDIATRICS, 2011, 128 (05) : 910 - 916