A robust algorithm of intensity modulated proton therapy for critical tissue sparing and target coverage

被引:33
作者
Inaniwa, Taku [1 ]
Kanematsu, Nobuyuki [1 ]
Furukawa, Takuji [1 ]
Hasegawa, Azusa
机构
[1] Natl Inst Radiol Sci, Res Ctr Charged Particle Therapy, Med Phys Res Program, Inage Ku, Chiba 2638555, Japan
关键词
TREATMENT UNCERTAINTIES; SCANNING IRRADIATION; RANGE UNCERTAINTIES; OPTIMIZATION; SENSITIVITY; PLANS; BEAM;
D O I
10.1088/0031-9155/56/15/008
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Intensity modulated proton therapy (IMPT) offers the possibility of generating excellent target coverage while sparing the neighbouring organs at risk. However, treatment plans optimized for IMPT may be very sensitive to range and setup uncertainties. We developed a method to deal with these uncertainties in the dose optimization. This method aims at two objectives: one for maintaining the dose coverage within the target, and the other for preventing undesired exposure to organs at risk. The former objective was achieved by the algorithm described in our previous paper to suppress the in-field dose gradient within the target. In this study, the latter objective was achieved by a novel algorithm in which we suppressed pencil beams with high risk to deliver undesired doses to organs at risk under conditions where range and setup uncertainties occur. We defined the risk index that quantifies the likelihood of each pencil beam delivering high doses to organs at risk, and introduced it into the objective function of dose optimizations. In order to test the algorithm's performance, this method was applied to an RTOG benchmark phantom geometry and to a cervical chordoma case. These simulations demonstrated that our method provides IMPT plans that are more robust against range and setup errors compared to conventional IMPT plans. Compared to the conventional IMPT plan, the optimization time for the robust plan increased by a factor of only 3, from 4 to 11 min.
引用
收藏
页码:4749 / 4770
页数:22
相关论文
共 20 条
[1]   The influence of the optimization starting conditions on the robustness of intensity-modulated proton therapy plans [J].
Albertini, F. ;
Hug, E. B. ;
Lomax, A. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (10) :2863-2878
[2]   Sensitivity of intensity modulated proton therapy plans to changes in patient weight [J].
Albertini, Francesca ;
Bolsi, Alessandra ;
Lomax, Antony J. ;
Rutz, Hans Peter ;
Timmerman, Beate ;
Goitein, Gudrun .
RADIOTHERAPY AND ONCOLOGY, 2008, 86 (02) :187-194
[3]   Development of treatment planning for scanning irradiation at HIMAC [J].
Inaniwa, Taku ;
Furukawa, Takuji ;
Sato, Shinji ;
Tomitani, Takehiro ;
Kobayashi, Masanao ;
Minohara, Shnichi ;
Noda, Koji ;
Kanai, Tatsuaki .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2008, 266 (10) :2194-2198
[4]   Optimization for fast-scanning irradiation in particle therapy [J].
Inaniwa, Taku ;
Furukawa, Takuji ;
Tomitani, Takehiro ;
Sato, Shinji ;
Noda, Koji ;
Kanai, Tatsuaki .
MEDICAL PHYSICS, 2007, 34 (08) :3302-3311
[5]   Optimization algorithm for overlapping-field plans of scanned ion beam therapy with reduced sensitivity to range and setup uncertainties [J].
Inaniwa, Taku ;
Kanematsu, Nobuyuki ;
Furukawa, Takuji ;
Noda, Koji .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (06) :1653-1669
[6]   Treatment planning for a scanned carbon beam with a modified microdosimetric kinetic model [J].
Inaniwa, Taku ;
Furukawa, Takuji ;
Kase, Yuki ;
Matsufuji, Naruhiro ;
Toshito, Toshiyuki ;
Matsumoto, Yoshitaka ;
Furusawa, Yoshiya ;
Noda, Koji .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (22) :6721-6737
[7]   Field-size effect of physical doses in carbon-ion scanning using range shifter plates [J].
Inaniwa, Taku ;
Furukawa, Takuji ;
Nagano, Ai ;
Sato, Shinji ;
Saotome, Naoya ;
Noda, Koji ;
Kanai, Tatsuaki .
MEDICAL PHYSICS, 2009, 36 (07) :2889-2897
[8]   A CT calibration method based on the polybinary tissue model for radiotherapy treatment planning [J].
Kanematsu, N ;
Matsufuji, N ;
Kohno, R ;
Minohara, S ;
Kanai, T .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (08) :1053-1064
[9]   A novel patch-field design using an optimized grid filter for passively scattered proton beams [J].
Li, Yupeng ;
Zhang, Xiaodong ;
Dong, Lei ;
Mohan, Radhe .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (12) :N265-N275
[10]   Intensity modulation methods for proton radiotherapy [J].
Lomax, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (01) :185-205