Effects of different insertion methods on reducing needle deflection

被引:24
作者
Abolhassani, Niki [1 ,2 ]
Patel, Rajni [1 ,2 ]
Ayazi, Farzam [3 ]
机构
[1] Univ Western Ontario, Dept Elect & Comp Engn, London, ON, Canada
[2] Univ Western Ontario, CSTAR, London, ON, Canada
[3] Lawson Hlth Res Inst, London, ON, Canada
来源
2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16 | 2007年
基金
加拿大自然科学与工程研究理事会; 加拿大创新基金会;
关键词
D O I
10.1109/IEMBS.2007.4352330
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Needle steering in medical procedures has attracted considerable attention in recent years. For example, in prostate brachytherapy, it is desired to insert a flexible beveled-tip needle with minimum deflection. To date, different methods of insertion which incorporate needle rotation about its insertion axis have been proposed in order to reduce needle deflection and target displacement. In this paper, needle deflection resulting from different methods of insertion are compared with our "model-based" method which estimates the amount of needle deflection using Euler-Bernoulli beam equations. Experiments are performed in gelatin phantoms and animal tissue. The results show that the proposed "model-based" method reduces the amount of needle deflection more than other methods. In this paper, some factors for choosing the appropriate method of insertion are also discussed.
引用
收藏
页码:491 / +
页数:2
相关论文
共 14 条
[1]  
ABOLHASSANI N, 2006, P INT C IEEE ENG MED, P3858
[2]  
ABOLHASSANI N, 2006, P COMP ASS RAD SURG, P269
[3]   Needle insertion into soft tissue: A survey [J].
Abolhassani, Niki ;
Patel, Rajni ;
Moallem, Mehrdad .
MEDICAL ENGINEERING & PHYSICS, 2007, 29 (04) :413-431
[4]   The effect of introducer gauge, design and bevel direction on the deflection of spinal needles [J].
Ahn, WS ;
Bahk, JH ;
Lim, YJ ;
Kim, YC .
ANAESTHESIA, 2002, 57 (10) :1007-1011
[5]  
ALTEROVITZ R, 2006, WORKSH ALG FDN ROB W
[6]   Needle steering and motion planning in soft tissues [J].
DiMaio, SP ;
Salcudean, SE .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2005, 52 (06) :965-974
[7]  
GERE JM, 2001, MECH MAT
[8]  
KATAOKA K, 2001, P MED IM COMP COMP A, P966
[9]   Force modeling for needle insertion into soft tissue [J].
Okamura, AM ;
Simone, C ;
O'Leary, MD .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2004, 51 (10) :1707-1716
[10]  
Podder TK, 2005, IEEE C ENG MED BIOL, P5766