Emissivity Measurement Under Vacuum from to and from to at PTB

被引:36
作者
Adibekyan, A. [1 ]
Monte, C. [1 ]
Kehrt, M. [1 ]
Gutschwager, B. [1 ]
Hollandt, J. [1 ]
机构
[1] Phys Tech Bundesanstalt, Berlin, Germany
关键词
Blackbody; Emissivity; Spectral radiance; Uncertainty; Vacuum;
D O I
10.1007/s10765-014-1745-7
中图分类号
O414.1 [热力学];
学科分类号
摘要
A new facility of the Physikalisch-Technische Bundesanstalt for emissivity measurements under vacuum was brought into operation, capable of measuring directional spectral emissivity in the wavelength range from to and temperature range from to . The wide spectral and thermal ranges in combination with a target uncertainty of less than 0.01 meet the needs of the solar thermal energy industry and climate research. Emissivities determined in these ranges and at that uncertainty levels would allow significantly improving solar thermal absorber coatings and designing improved reference blackbodies for remote sensing. Here the measurement scheme which is based on the measurement of the spectral radiance of a sample with respect to the spectral radiance of two blackbodies at different temperatures is presented, and the subsequent evaluation of the emissivity is discussed. The first results of emissivity measurements under vacuum in the wavelength range from to and temperature range from to are presented and compared with results obtained in air, and it is shown that samples with very low emissivity can be measured with the required low target uncertainty of less than 0.01.
引用
收藏
页码:283 / 289
页数:7
相关论文
共 9 条
[1]   The development of emissivity measurements under vacuum at the PTB [J].
Adibekyan, A. R. ;
Monte, C. ;
Kehrt, M. ;
Morozova, S. P. ;
Gutschwager, B. ;
Hollandt, J. .
MEASUREMENT TECHNIQUES, 2013, 55 (10) :1163-1171
[2]  
Kennedy A.C.E., 2002, REV MID TO HIGH TEMP
[3]   A high-accuracy blackbody for CLARREO [J].
Latvakoski, Harri ;
Watson, Mike ;
Topham, Shane ;
Scott, Deron ;
Wojcik, Mike ;
Bingham, Gail .
INFRARED REMOTE SENSING AND INSTRUMENTATION XVIII, 2010, 7808
[4]   Radiation Thermometry and Emissivity Measurements Under Vacuum at the PTB [J].
Monte, C. ;
Gutschwager, B. ;
Morozova, S. P. ;
Hollandt, J. .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2009, 30 (01) :203-219
[5]  
Monte C., 2010, High Temperatures-High Pressures, V39, P151
[6]   The determination of the uncertainties of spectral emissivity measurements in air at the PTB [J].
Monte, Christian ;
Hollandt, Joerg .
METROLOGIA, 2010, 47 (02) :S172-S181
[7]   Vacuum variable-temperature blackbody VTBB100 [J].
Morozova, S. P. ;
Parfentiev, N. A. ;
Lisiansky, B. E. ;
Sapritsky, V. I. ;
Dovgilov, N. L. ;
Melenevsky, U. A. ;
Gutschwager, B. ;
Monte, C. ;
Hollandt, J. .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2008, 29 (01) :341-351
[8]   Vacuum Variable Medium Temperature Blackbody [J].
Morozova, S. P. ;
Parfentiev, N. A. ;
Lisiansky, B. E. ;
Melenevsky, U. A. ;
Gutschwager, B. ;
Monte, C. ;
Hollandt, J. .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2010, 31 (8-9) :1809-1820
[9]   Analysis of the accuracy of methods for the direct measurement of emissivity [J].
Perez-Saez, Raul B. ;
del Campo, Leire ;
Tello, Manuel J. .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2008, 29 (03) :1141-1155