Gamma spectrometry calibrations with natural radioactive materials

被引:3
作者
Pantelic, G
机构
关键词
D O I
10.1016/S0168-9002(96)80053-2
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Natural radioactive materials were used for detector calibration. We found that KCl is a very suitable material for this purpose. The efficiency curve shape was derived by using gamma ray lines of Bi-214 normalised using a known quantity of KCl in the same geometry. The best fit was found by the least squares method. The summing correction coefficients for Bi-214 are determined.
引用
收藏
页码:572 / 573
页数:2
相关论文
共 2 条
[1]  
DEBERTIN K, 1990, PTB RA24
[2]  
KUKOC AH, 1993, RADIOCHIM ACTA, V60, P25