Imaging Near Metal with a MAVRIC-SEMAC Hybrid

被引:184
作者
Koch, K. M. [1 ]
Brau, A. C. [2 ]
Chen, W. [2 ]
Gold, G. E. [3 ]
Hargreaves, B. A. [3 ]
Koff, M. [4 ]
McKinnon, G. C. [1 ]
Potter, H. G. [4 ]
King, K. F. [1 ]
机构
[1] GE Healthcare, Appl Sci Lab, Waukesha, WI 53188 USA
[2] GE Healthcare, Appl Sci Lab, Menlo Pk, CA USA
[3] Stanford Univ, Dept Radiol, Palo Alto, CA 94304 USA
[4] Hosp Special Surg, Dept Radiol, New York, NY 10021 USA
关键词
MRI of arthroplasty; metal implants; image artifacts; 3D imaging; TOTAL HIP-ARTHROPLASTY; ARTIFACT CORRECTION; MRI; IMPLANTS;
D O I
10.1002/mrm.22523
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The recently developed multi-acquisition with variable resonance image combination (MAVRIC) and slice-encoding metal artifact correction (SEMAC) techniques can significantly reduce image artifacts commonly encountered near embedded metal hardware. These artifact reductions are enabled by applying alternative spectral and spatial-encoding schemes to conventional spin-echo imaging techniques. Here, the MAVRIC and SEMAC concepts are connected and discussed. The development of a hybrid technique that utilizes strengths of both methods is then introduced. The presented technique is shown capable of producing minimal artifact, high-resolution images near total joint replacements in a clinical setting. Magn Reson Med 65:71-82, 2011. (c) 2010 Wiley-Liss, Inc.
引用
收藏
页码:71 / 82
页数:12
相关论文
共 26 条
[1]  
Beatty PJ., 2007, Proceedings of the 15th Annual Meeting of ISMRM, P1749
[2]   Comparison of reconstruction accuracy and efficiency among autocalibrating data-driven parallel imaging methods [J].
Brau, Anja C. S. ;
Beatty, Philip J. ;
Skare, Stefan ;
Bammer, Roland .
MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (02) :382-395
[3]   Effects of refocusing flip angle modulation and view ordering in 3D fast spin echo [J].
Busse, Reed F. ;
Brau, Anja C. S. ;
Vu, Anthony ;
Michelich, Charles R. ;
Bayram, Ersin ;
Kijowski, Richard ;
Reeder, Scott B. ;
Rowley, Howard A. .
MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (03) :640-649
[4]  
Butts K, 1999, JMRI-J MAGN RESON IM, V9, P586, DOI 10.1002/(SICI)1522-2586(199904)9:4<586::AID-JMRI13>3.0.CO
[5]  
2-X
[6]  
Chen C., 2010, P ISMRM STOCKH SWED, P130
[7]  
Chen W., 2009, Proc Int Soc Magnetic Resonance in Medicine, page, P2783
[8]   TOTAL INHOMOGENEITY CORRECTION INCLUDING CHEMICAL-SHIFTS AND SUSCEPTIBILITY BY VIEW ANGLE TILTING [J].
CHO, ZH ;
KIM, DJ ;
KIM, YK .
MEDICAL PHYSICS, 1988, 15 (01) :7-11
[9]  
GOLD GE, 2010, P ISMRM STOCKH SWED, P893
[10]   Accelerated Slice Encoding for Metal Artifact Correction [J].
Hargreaves, Brian A. ;
Chen, Weitian ;
Lu, Wenmiao ;
Alley, Marcus T. ;
Gold, Garry E. ;
Brau, Anja C. S. ;
Pauly, John M. ;
Pauly, Kim Butts .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2010, 31 (04) :987-996