摘要
Single molecule localization microscopy (SMLM) techniques allow for sub-diffraction imaging with spatial resolutions better than 10 nm reported. Much has been discussed relating to different variations of SMLM and all-inclusive microscopes can now be purchased, removing the need for in-house software or hardware development. However, little discussion has occurred examining the reliability and quality of the images being produced, as well as the potential for overlooked preparative artifacts. As a result of the up to an order-of-magnitude improvement in spatial resolution, substantially more detail is observed, including changes in distribution and ultrastructure caused by the many steps required to fix, permeabilize, and stain a sample. Here we systematically investigate many of these steps including different fixatives, fixative concentration, permeabilization concentration and timing, antibody concentration, and buffering. We present three well-optimized fixation protocols for staining microtubules, mitochondria and actin in a mammalian cell line and then discuss various artifacts in relation to images obtained from samples prepared using the protocols. The potential for such errors to go undetected in SMLM images and the complications in defining a 'good' image using previous parameters applied to confocal microscopy are also discussed.
引用
收藏
页数:10
相关论文
共 47 条
[1]   Multicolor super-resolution imaging with photo-switchable fluorescent probes [J].
Bates, Mark ;
Huang, Bo ;
Dempsey, Graham T. ;
Zhuang, Xiaowei .
SCIENCE, 2007, 317 (5845) :1749-1753
[2]   Immunofluorescence imaging of DNA damage response proteins: Optimizing protocols for super-resolution microscopy [J].
Bennett, Brian T. ;
Bewersdorf, Joerg ;
Knight, Kendall L. .
METHODS, 2009, 48 (01) :63-71
[3]   Imaging intracellular fluorescent proteins at nanometer resolution [J].
Betzig, Eric ;
Patterson, George H. ;
Sougrat, Rachid ;
Lindwasser, O. Wolf ;
Olenych, Scott ;
Bonifacino, Juan S. ;
Davidson, Michael W. ;
Lippincott-Schwartz, Jennifer ;
Hess, Harald F. .
SCIENCE, 2006, 313 (5793) :1642-1645
[4]   Formalin Fixation at Low Temperature Better Preserves Nucleic Acid Integrity [J].
Bussolati, Gianni ;
Annaratone, Laura ;
Medico, Enzo ;
D'Armento, Giuseppe ;
Sapino, Anna .
PLOS ONE, 2011, 6 (06)
[5]   FILAMENTOUS CYTOSKELETON IN VERTEBRATE SMOOTH-MUSCLE FIBERS [J].
COOKE, P .
JOURNAL OF CELL BIOLOGY, 1976, 68 (03) :539-556
[6]   Localizer: fast, accurate, open-source, and modular software package for superresolution microscopy [J].
Dedecker, Peter ;
Duwe, Sam ;
Neely, Robert K. ;
Zhang, Jin .
JOURNAL OF BIOMEDICAL OPTICS, 2012, 17 (12)
[7]  
Dempsey GT, 2011, NAT METHODS, V8, P1027, DOI [10.1038/NMETH.1768, 10.1038/nmeth.1768]
[8]   Shrinkage and distortion of the rabbit corneal endothelial cell mosaic caused by a high osmolality glutaraldehyde-formaldehyde fixative compared to glutaraldehyde [J].
Doughty, MJ ;
Bergmanson, JPG ;
Blocker, Y .
TISSUE & CELL, 1997, 29 (05) :533-547
[9]   FORMALDEHYDE FIXATION [J].
FOX, CH ;
JOHNSON, FB ;
WHITING, J ;
ROLLER, PP .
JOURNAL OF HISTOCHEMISTRY & CYTOCHEMISTRY, 1985, 33 (08) :845-853
[10]   Imaging biological structures with fluorescence photoactivation localization microscopy [J].
Gould, Travis J. ;
Verkhusha, Vladislav V. ;
Hess, Samuel T. .
NATURE PROTOCOLS, 2009, 4 (03) :291-308