A Clinical Pilot Study of a Modular Video-CT Augmentation System for Image-Guided Skull Base Surgery

被引:2
作者
Liu, Wen P. [1 ]
Mirota, Daniel J. [1 ]
Uneri, Ali [1 ]
Otake, Yoshito
Hager, Gregory [1 ]
Reh, Douglas D.
Ishii, Masaru
Gallia, Gary L.
Siewerdsen, Jeffrey H. [1 ]
机构
[1] Johns Hopkins Univ, Dept Comp Sci, Baltimore, MD 21218 USA
来源
MEDICAL IMAGING 2012: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING | 2012年 / 8316卷
基金
美国国家卫生研究院;
关键词
video-CT augmentation; image-guided surgery; skull-base surgery; surgical navigation; video endoscopy; ENDOSCOPIC SINUS SURGERY; CALIBRATION; NAVIGATION;
D O I
10.1117/12.911724
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Augmentation of endoscopic video with preoperative or intraoperative image data [ e. g., planning data and/or anatomical segmentations defined in computed tomography (CT) and magnetic resonance (MR)], can improve navigation, spatial orientation, confidence, and tissue resection in skull base surgery, especially with respect to critical neurovascular structures that may be difficult to visualize in the video scene. This paper presents the engineering and evaluation of a video augmentation system for endoscopic skull base surgery translated to use in a clinical study. Extension of previous research yielded a practical system with a modular design that can be applied to other endoscopic surgeries, including orthopedic, abdominal, and thoracic procedures. A clinical pilot study is underway to assess feasibility and benefit to surgical performance by overlaying CT or MR planning data in real-time, high-definition endoscopic video. Preoperative planning included segmentation of the carotid arteries, optic nerves, and surgical target volume (e. g., tumor). An automated camera calibration process was developed that demonstrates mean re-projection accuracy (0.7 +/- 0.3) pixels and mean target registration error of (2.3 +/- 1.5) mm. An IRB-approved clinical study involving fifteen patients undergoing skull base tumor surgery is underway in which each surgery includes the experimental video-CT system deployed in parallel to the standard-of-care (un-augmented) video display. Questionnaires distributed to one neurosurgeon and two otolaryngologists are used to assess primary outcome measures regarding the benefit to surgical confidence in localizing critical structures and targets by means of video overlay during surgical approach, resection, and reconstruction.
引用
收藏
页数:9
相关论文
共 17 条
[1]  
[Anonymous], 2000, Opencv. Dr. Dobb's journal of software tools
[2]   Intraoperative cone-beam CT for guidance of head and neck surgery: Assessment of dose and image quality using a C-arm prototype [J].
Daly, M. J. ;
Siewerdsen, J. H. ;
Moseley, D. J. ;
Jaffray, D. A. ;
Irish, J. C. .
MEDICAL PHYSICS, 2006, 33 (10) :3767-3780
[3]   Hand-eye calibration using dual quaternions [J].
Daniilidis, K .
INTERNATIONAL JOURNAL OF ROBOTICS RESEARCH, 1999, 18 (03) :286-298
[4]  
Deguet A., 2008, MIDAS J SYST ARCH CO
[5]  
Fraser J. F., 2008, MINIM INVAS NEUROSUR, V5, P1
[6]   Endoscopic endonasal skull base surgery: analysis of complications in the authors' initial 800 patients A review [J].
Kassam, Amin B. ;
Preveoello, Daniel M. ;
Carrau, Ricardo L. ;
Snyderman, Carl H. ;
Thomas, Ajith ;
Gardner, Paul ;
Zanation, Adam ;
Duz, Bulent ;
Stefko, S. Tonya ;
Byers, Karin ;
Horowitz, Michael B. .
JOURNAL OF NEUROSURGERY, 2011, 114 (06) :1544-1568
[7]   Endoscopic augmented reality navigation system for endonasal transsphenoidal surgery to treat pituitary tumors: Technical note [J].
Kawamata, T ;
Iseki, H ;
Shibasaki, T ;
Hori, T .
NEUROSURGERY, 2002, 50 (06) :1393-1397
[8]   Image-enhanced surgical navigation for endoscopic sinus surgery: evaluating calibration, registration and tracking [J].
Lapeer, R. ;
Chen, M. S. ;
Gonzalez, G. ;
Linney, A. ;
Alusi, G. .
INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2008, 4 (01) :32-45
[9]  
Mirota D. J., 2011, SPIE MED IM ORL
[10]  
Pieper S, 2006, I S BIOMED IMAGING, P698