Radiation dose optimization in pediatric temporal bone computed tomography: influence of tube tension on image contrast and image quality

被引:6
作者
Nauer, Claude Bertrand [1 ,2 ]
Zubler, Christoph [2 ]
Weisstanner, Christian [2 ]
Stieger, Christof [3 ]
Senn, Pascal [4 ]
Arnold, Andreas [3 ,4 ]
机构
[1] Kantonsspital Graubunden, Zent Roentgeninst, CH-7000 Chur, Switzerland
[2] Univ Hosp Bern, Inst Diagnost & Intervent Neuroradiol, CH-3010 Bern, Switzerland
[3] Univ Bern, ARTORG Ctr, Grp Artificial Hearing Res, Bern, Switzerland
[4] Univ Hosp Bern, Dept ENT Head & Neck Surg, CH-3010 Bern, Switzerland
关键词
Image quality; Temporal bone CT; Tube tension; Radiation exposure to patients; Pediatric CT; RAY TECHNIQUE FACTORS; PATIENT SIZE; CT; HEAD; EXPOSURE; CHILDREN;
D O I
10.1007/s00234-011-0961-0
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The purpose of this experimental study was to investigate the effect of tube tension reduction on image contrast and image quality in pediatric temporal bone computed tomography (CT). Seven lamb heads with infant-equivalent sizes were scanned repeatedly, using four tube tensions from 140 to 80 kV while the CT-Dose Index (CTDI) was held constant. Scanning was repeated with four CTDI values from 30 to 3 mGy. Image contrast was calculated for the middle ear as the Hounsfield unit (HU) difference between bone and air and for the inner ear as the HU difference between bone and fluid. The influence of tube tension on high-contrast detail delineation was evaluated using a phantom. The subjective image quality of eight middle and inner ear structures was assessed using a 4-point scale (scores 1-2 = insufficient; scores 3-4 = sufficient). Middle and inner ear contrast showed a near linear increase with tube tension reduction (r = -0.94/-0.88) and was highest at 80 kV. Tube tension had no influence on spatial resolution. Subjective image quality analysis showed significantly better scoring at lower tube tensions, with highest image quality at 80 kV. However, image quality improvement was most relevant for low-dose scans. Image contrast in the temporal bone is significantly higher at low tube tensions, leading to a better subjective image quality. Highest contrast and best quality were found at 80 kV. This image quality improvement might be utilized to further reduce the radiation dose in pediatric low-dose CT protocols.
引用
收藏
页码:247 / 254
页数:8
相关论文
共 20 条
[1]  
[Anonymous], 2009, FUNDAMENTALS MED IMA, DOI DOI 10.1017/CBO9780511596803
[2]   Morphometric study of the external and middle ear anatomy in sheep: A possible model for ear experiments [J].
Ayres Seibel, Valter Alberto ;
Lavinsky, Luiz ;
Preto De Oliveira, Jost Alfredo .
CLINICAL ANATOMY, 2006, 19 (06) :503-509
[3]   MRI versus CT in assessment of cochlear patency in cochlear implant candidates [J].
Bettman, R ;
Beek, E ;
Van Olphen, A ;
Zonneveld, F ;
Huizing, E .
ACTA OTO-LARYNGOLOGICA, 2004, 124 (05) :577-581
[4]   Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[5]   Estimated risks of radiation-induced fatal cancer from pediatric CT [J].
Brenner, DJ ;
Elliston, CD ;
Hall, EJ ;
Berdon, WE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) :289-296
[6]   Exposure to Low-Dose Ionizing Radiation from Medical Imaging Procedures. [J].
Fazel, Reza ;
Krumholz, Harlan M. ;
Wang, Yongfei ;
Ross, Joseph S. ;
Chen, Jersey ;
Ting, Henry H. ;
Shah, Nilay D. ;
Nasir, Khurram ;
Einstein, Andrew J. ;
Nallamothu, Brahmajee K. .
NEW ENGLAND JOURNAL OF MEDICINE, 2009, 361 (09) :849-857
[7]   The Image Gently campaign:: Working together to change practice [J].
Goske, Marilyn J. ;
Applegate, Kimberly E. ;
Boylan, Jennifer ;
Butler, Priscilla F. ;
Callahan, Michael J. ;
Coley, Brian D. ;
Farley, Shawn ;
Frush, Donald P. ;
Hernanz-Schulman, Marta ;
Jaramillo, Diego ;
Johnson, Neil D. ;
Kaste, Sue C. ;
Morrison, Greg ;
Strauss, Keith J. ;
Tuggle, Nora .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (02) :273-274
[8]   Patient size and x-ray technique factors in head computed tomography examinations. II. Image quality [J].
Huda, W ;
Lieberman, KA ;
Chang, J ;
Roskopf, ML .
MEDICAL PHYSICS, 2004, 31 (03) :595-601
[9]   Patient size and x-ray technique factors in head computed tomography examinations. I. Radiation doses [J].
Huda, W ;
Lieberman, KA ;
Chang, J ;
Roskopf, ML .
MEDICAL PHYSICS, 2004, 31 (03) :588-594
[10]   Influence of patient age on normalized effective doses calculated for CT examinations [J].
Khursheed, A ;
Hillier, MC ;
Shrimpton, PC ;
Wall, BF .
BRITISH JOURNAL OF RADIOLOGY, 2002, 75 (898) :819-830