The influence of abutment angulation on micromotion level for immediately loaded dental implants: A 3-D finite element analysis

被引:2
作者
Kao, Hung-Chan [2 ,3 ]
Gung, Yih-Wen [1 ]
Chung, Tai-Foong [1 ]
Hsu, Ming-Lun [1 ]
机构
[1] Natl Yang Ming Univ, Sch Dent, Taipei 112, Taiwan
[2] Natl Yang Ming Univ, Joint Prosthesis Technol Res Ctr, Taipei 112, Taiwan
[3] Natl Yang Ming Univ, Inst Biomech Engn, Taipei 112, Taiwan
关键词
angled abutment; finite element analysis; immediate loading; micromotion;
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: To investigate the micromotion between the implant and surrounding bone caused by the implementation of an angled abutment for an immediately loaded single dental implant located in the anterior maxilla. Materials and Methods: A simplified half premaxillary bone model was fabricated. The dimension of the alveolar ridge was adopted from a dry human skull. Based on Branemark protocol for Mk IV implants in type-3 bone, an immediate loading model was developed by press-fitting a 4-mm-diameter cylinder implant into a 3.15-mm osteotomy site in a numeric model. Material properties were assigned to the simulated model, and the model was meshed. A bite force of 89 N was applied to the tops of the 0-degree, 15-degree, and 25-degree angled abutments at a 120-degree angle to the abutment long axis. The micromotion between the bone-implant interfaces was calculated using ANSYS 9.0 software featuring a nonlinear contact algorithm. Results: The micromotion values for 15-degree and 25-degree angled abutments were 119% and 134%, respectively, compared to the corresponding values for straight abutments. Compared to straight abutments, the 25-degree abutments resulted in increased maximum von Mises stresses to a level of 18%. Most of the stresses were concentrated within the cortical bone around the neck of the implants. Conclusion: Within the limits of the present finite element analysis study, abutment angulation up to 25 degrees can increase the stress in the peri-implant bone by 18% and the micromotion level by 30%.
引用
收藏
页码:623 / 630
页数:8
相关论文
共 45 条
[1]  
ADELL R, 1990, International Journal of Oral and Maxillofacial Implants, V5, P347
[2]   Implant design and intraosseous stability of immediately placed implants: a human cadaver study [J].
Akkocaoglu, M ;
Uysal, S ;
Tekdemir, I ;
Akca, K ;
Cehreli, MC .
CLINICAL ORAL IMPLANTS RESEARCH, 2005, 16 (02) :202-209
[3]   Immediate and early implant loading protocols: A literature review of clinical studies [J].
Attard, NJ ;
Zarb, GA .
JOURNAL OF PROSTHETIC DENTISTRY, 2005, 94 (03) :242-258
[4]   Implant prosthodontic management of partially edentulous patients missing posterior teeth: The Toronto experience [J].
Attard, NJ ;
Zarb, GA .
JOURNAL OF PROSTHETIC DENTISTRY, 2003, 89 (04) :352-359
[5]   THE HYDRODYNAMIC MODEL OF SLIDING INCLINED PLANES AND ITS 2 LIMITS - COULOMB AMONTON FRICTION AND FLUID SLUG RIGID BODY FLOW [J].
AVITZUR, B .
TRIBOLOGY TRANSACTIONS, 1993, 36 (02) :249-257
[6]   TITANIUM PLASMA-SPRAYED (TPS) SCREW IMPLANTS FOR THE RECONSTRUCTION OF THE EDENTULOUS MANDIBLE [J].
BABBUSH, CA ;
KENT, JN ;
MISIEK, DJ .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1986, 44 (04) :274-282
[7]  
Balshi TJ, 1997, INT J ORAL MAX IMPL, V12, P52
[8]   Implant stability measurement of delayed and immediately loaded implants during healing. A clinical resonance-frequency analysis study with sandblasted-and-etched ITI implants [J].
Bischof, M ;
Nedir, R ;
Szmukler-Moncler, S ;
Bernard, JP ;
Samson, J .
CLINICAL ORAL IMPLANTS RESEARCH, 2004, 15 (05) :529-539
[9]  
BRANEMARK P I, 1969, Scandinavian Journal of Plastic and Reconstructive Surgery, V3, P81
[10]   The influence of abutment angulation on strains and stresses along the implant/bone interface: Comparison between two experimental techniques [J].
Brosh, T ;
Pilo, R ;
Sudai, D .
JOURNAL OF PROSTHETIC DENTISTRY, 1998, 79 (03) :328-334