Design considerations and initial performance of a 1.2 cm(2) beta imaging intra-operative probe

被引:41
作者
Tornai, MP [1 ]
MacDonald, LR [1 ]
Levin, CS [1 ]
Siegel, S [1 ]
Hoffman, EJ [1 ]
机构
[1] UNIV CALIF LOS ANGELES, DEPT PHYS, LOS ANGELES, CA 90095 USA
关键词
D O I
10.1109/23.531898
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A navel small area beta (beta(+/-)) detector is under development for nuclear emission imaging of surgically exposed, radiolabeled tumor beds. The imaging device front-end consists of a 0.5 mm thick by 1.25 cm diameter CaF2(Eu) scintillator disk coupled to a rigid bundle of 2 mm diameter double clad optical fibers through a polystyrene light diffuser. The detector area (1.2 cm(2)) was determined by the requirement of introducing the probe into small cavities, e.g. during neuro-surgical lesion resection, but large enough to produce images of clinical significance. Flexible back-end optical fibers (1.9 m long) were coupled to the front-end components allowing similar to 75 photoelectrons to be detected for mean beta energies of 250 keV, indicating that sufficient signal can be obtained with clinical beta emitters (e.g. F-18, I-131). The long flexible fibers guide the scintillation light to a Philips XP1700 series, fiber optic faceplate, Multi-Channel PMT. The parallel MC-PMT outputs are fed into a variable gain, charge divider network and an i-V pre-amplifier/line driver network, whose resulting four outputs are digitized and histogrammed with standard Anger positioning logic. The various components in the imaging chain were evaluated and optimized by both simulations and measurements. Line spread functions measured in the 10.8 mm FOV were 0.50 mm +/- 0.038 mm and 0.55 mm +/- 0.065 mm FWHM in X and Y, respectively. A 20% variation in pulse height and minimal variation in spatial resolution was observed. The differential image uniformity was measured to be +/-15.6% with similar to 150 cts/pixel. Preliminary images show excellent reproduction of phantom activity distributions.
引用
收藏
页码:2326 / 2335
页数:10
相关论文
共 41 条
  • [1] SCINTILLATION CAMERA
    ANGER, HO
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1958, 29 (01) : 27 - 33
  • [2] AN INVESTIGATION OF PARAMETERS IN SCINTILLATION CAMERA DESIGN
    BAKER, RG
    SCRIMGER, JW
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1967, 12 (01) : 51 - +
  • [3] SMALL RADIATION DETECTORS FOR BRONCHOSCOPIC TUMOR-LOCALIZATION
    BARBER, HB
    WOOLFENDEN, JM
    DONAHUE, DJ
    NEVIN, WS
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1980, 27 (01) : 496 - 502
  • [4] COMPARISON OF INVIVO SCINTILLATION PROBES AND GAMMA CAMERAS FOR DETECTION OF SMALL, DEEP TUMORS
    BARBER, HB
    BARRETT, HH
    WOOLFENDEN, JM
    MYERS, KJ
    HICKERNELL, TS
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (06) : 727 - 739
  • [5] BARRETT HH, 1981, RAIDOLOGICAL IMAGING, V1, pCH5
  • [6] PERFORMANCE OF MULTICLAD SCINTILLATING AND CLEAR WAVE-GUIDE FIBERS READ-OUT WITH VISIBLE-LIGHT PHOTON COUNTERS
    BAUMBAUGH, B
    ERDMAN, J
    GASKELL, D
    LU, Q
    MARCHANT, J
    RUCHTI, R
    WAYNE, M
    COOPER, C
    HINSON, J
    KOLTICK, DS
    SCHMITZ, CJ
    SHIBATA, EI
    CHUNG, M
    MARGULIES, S
    SOLOMON, J
    ATAC, M
    BAUMBAUGH, A
    BROSS, AD
    ELIAS, J
    MISHINA, M
    CHANEY, R
    HAMMACK, H
    ADAMS, D
    BIRD, F
    LEWIS, RA
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 345 (02) : 271 - 278
  • [7] APPLICATION OF MONTE-CARLO METHODS TO THE DESIGN OF SPECT DETECTOR SYSTEMS
    BRADSHAW, J
    BURNHAM, C
    CORREIA, J
    ROGERS, WL
    CLINTHORNE, NH
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1985, 32 (01) : 753 - 757
  • [8] TEST OF A NEW 64-CHANNEL PMT FOR IMAGING
    COMBY, G
    MEUNIER, R
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 269 (01) : 246 - 260
  • [9] A POSITION-SENSITIVE DETECTOR FOR THE G-2 EXPERIMENT USING PENCIL FIBERS
    CUSHMAN, P
    HOU, S
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 339 (03) : 456 - 467
  • [10] INTRAOPERATIVE BETA-PROBE - A DEVICE FOR DETECTING TISSUE LABELED WITH POSITRON OR ELECTRON EMITTING ISOTOPES DURING SURGERY
    DAGHIGHIAN, F
    MAZZIOTTA, JC
    HOFFMAN, EJ
    SHENDEROV, P
    ESHAGHIAN, B
    SIEGEL, S
    PHELPS, ME
    [J]. MEDICAL PHYSICS, 1994, 21 (01) : 153 - 157