A CD probe with a tailored cantilever for 3D-AFM measurement

被引:11
作者
Zhang, Rui [1 ]
Wu, Sen [1 ]
Liu, Lu [1 ]
Lu, Nian-Hang [1 ]
Fu, Xing [1 ]
Gao, Si-Tian [2 ]
Hu, Xiao-Dong [1 ]
机构
[1] Tianjin Univ, Sch Precis Instruments & Optoelect Engn, State Key Lab Precis Measurement Technol & Instru, Tianjin 300072, Peoples R China
[2] Natl Inst Metrol, Beijing 100029, Peoples R China
基金
中国国家自然科学基金;
关键词
3D-AFM; CD probe; tailored cantilever; deformation; sidewall; SCANNING METHOD; AFM; NANOSTRUCTURES;
D O I
10.1088/1361-6501/aae8f4
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A critical dimensional (CD) probe, such as a boot shaped probe, is widely used for measuring the sidewall profiles of nanostructures in atomic force microscopy (AFM). However, conventional CD probes usually have high spring constants that restrict their ability to operate in critical dimension scanning modes. Furthermore, the flexure spring constant of the tip is much smaller than the torsion spring constant of the cantilever, which probably makes the sidewall measurement unstable. Here, we proposed a method of tailoring the probe where the cantilever was trimmed to reduce the spring constant. The simulation results, using a finite element method (FEM), illustrated that both the flexure and torsion spring constants decreased nearly ten times. Meanwhile, the torsional resonant frequency decreased to below 1 MHz, which could benefit the torsional oscillation scanning. By using a focused ion beam (FIB), a CD probe with a rectangular cantilever was reprocessed. It was mounted to a homebuilt three-dimensional AFM system (3D-AFM), and had the flexure and torsion sensitivities precisely calibrated. The ratio of the flexure sensitivity to torsion sensitivity showed good accordance with the simulation results. Additionally, a grating standard sample was measured using the tailored probe under a modified step-in scanning mode, and the error due to torsional deformation of the tip was compensated. The measurement results demonstrated the validity of the tailored CD probe in 3D metrology.
引用
收藏
页数:9
相关论文
共 25 条
[1]  
Bodermann B., 2009, P SOC PHOTO-OPT INS, V7470
[2]   Three-dimensional imaging of undercut and sidewall structures by atomic force microscopy [J].
Cho, Sang-Joon ;
Ahn, Byung-Woon ;
Kim, Joonhui ;
Lee, Jung-Min ;
Hua, Yueming ;
Yoo, Young K. ;
Park, Sang-il .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2011, 82 (02)
[3]   Comparison of line width calibration using critical dimension atomic force microscopes between PTB and NIST [J].
Dai, Gaoliang ;
Hahm, Kai ;
Bosse, Harald ;
Dixson, Ronald G. .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2017, 28 (06)
[4]   Measurements of CD and sidewall profile of EUV photomask structures using CD-AFM and tilting-AFM [J].
Dai, Gaoliang ;
Hahm, Kai ;
Scholze, Frank ;
Henn, Mark-Alexander ;
Gross, Hermann ;
Fluegge, Jens ;
Bosse, Harald .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2014, 25 (04)
[5]   Development of a 3D-AFM for true 3D measurements of nanostructures [J].
Dai, Gaoliang ;
Haessler-Grohne, Wolfgang ;
Hueser, Dorothee ;
Wolff, Helmut ;
Danzebrink, Hans-Ulrich ;
Koenders, Ludger ;
Bosse, Harald .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2011, 22 (09)
[6]   Effects of lateral tip control in CD-AFM width metrology [J].
Dixson, Ronald ;
Ng, Boon Ping ;
Orji, Ndubuisi .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2014, 25 (09)
[7]  
Faurie P., 2009, P SPIE, V7520, p75200F, DOI [10.1117/12.837311, DOI 10.1117/12.837311]
[8]   Introduction of Next-Generation 3D AFM for Advanced Process Control [J].
Foucher, J. ;
Therese, R. ;
Lee, Y. ;
Park, S. -, I ;
Cho, S. -J. .
METROLOGY, INSPECTION, AND PROCESS CONTROL FOR MICROLITHOGRAPHY XXVII, 2013, 8681
[9]  
Foucher J, 2005, P SOC PHOTO-OPT INS, V5752, P5752
[10]  
Foucher J., 2008, SPIE NEWSROOM