Analytical solution to Cone-beam SPECT reconstruction with non-uniform attenuation and distance-dependent resolution variation

被引:1
|
作者
Zhang, Hao [1 ]
Wen, Junhai [1 ]
Yang, Jing [1 ]
Hu, Peng [1 ]
Liang, Zhengrong [2 ]
机构
[1] Beijing Inst Technol, Sch Life Sci & Technol, Dept Biomed Engn, Beijing 100081, Peoples R China
[2] SUNY Stony Brook, Dept Radiol, Stony Brook, NY 11794 USA
基金
中国国家自然科学基金;
关键词
SPECT; cone-beam; reconstruction; Poisson noise; DDRV; non-uniform attenuation;
D O I
10.1117/12.843790
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Single photon emission computed tomography (SPECT) is a nuclear medicine imaging technique and widely used in the clinical applications. SPECT image reflects not only organizational structure but also functional activities of human body, such as blood-flow and metabolism condition, therefore diseases can be found much earlier. For many clinical applications, cone-beam geometry is preferred, which can improve count density and spatial resolution, and quantitative reconstruction of radiotracer distribution inside the body is desired. In this paper, we developed an efficient, analytical solution to cone-beam SPECT reconstruction with simultaneous compensation for attenuation and distance-dependent resolution variation (DDRV), as well as accurate treatment of Poisson noise. The simulation results show our reconstruction framework is feasible.
引用
收藏
页数:9
相关论文
共 47 条
  • [1] Analytical reconstruction for helical cone-beam SPECT with non-uniform attenuation correction
    Huang, Qiu
    Zeng, Gengsheng L.
    Gullberg, Grant T.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 1856 - 1859
  • [2] Analytical Cone Beam SPECT Reconstruction with Non-uniform Attenuation
    Wen, Junhai
    Chen, Yunbin
    Wang, Li
    2009 3RD INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOMEDICAL ENGINEERING, VOLS 1-11, 2009, : 2136 - 2139
  • [3] Towards analytical solution for 3D SPECT reconstruction with non-uniform attenuation and distance-dependent resolution variation: A Monte Carlo simulation study
    Lu, HB
    Wen, JH
    Li, X
    Li, TF
    Han, GP
    Liang, ZG
    MEDICAL IMAGING 2002: IMAGE PROCESSING, VOL 1-3, 2002, 4684 : 20 - 28
  • [4] Analytical cone-beam SPECT reconstruction algorithm with non-uniform attenuation for general non-circular orbit
    Wen, Junhai
    Zhang, Kangping
    Li, Cuifen
    Liang, Zhengrong
    COMPUTERS IN BIOLOGY AND MEDICINE, 2013, 43 (09) : 1221 - 1233
  • [5] A non-iterative solution for quantitative myocardial perfusion SPECT reconstruction with non-uniform attenuation and distance-dependent spatial resolution.
    Lu, H
    Wen, J
    Li, X
    Liang, JZ
    JOURNAL OF NUCLEAR MEDICINE, 2002, 43 (05) : 105P - 105P
  • [6] Rebinning Methods for Cone Beam SPECT Reconstruction with Non-uniform Attenuation
    Chen, Yunbin
    Wen, Junhai
    Wang, Li
    Hu, Peng
    Zhang, Hao
    PROCEEDINGS OF THE 2009 2ND INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS, VOLS 1-4, 2009, : 337 - 340
  • [7] Effect of de-noising and DDRV correction on cone-beam SPECT reconstruction with non-uniform attenuation
    Zhang, Hao
    Wen, Junhai
    Yin, Wen
    Li, Cuifen
    Zhang, Kangping
    Liang, Zhengrong
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [8] Analytical fan-beam and cone-beam reconstruction algorithms with uniform attenuation correction for SPECT
    Tang, QL
    Zeng, GSL
    Gullberg, GT
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (13): : 3153 - 3170
  • [9] Super resolution SPECT reconstruction with non-uniform attenuation
    Yan, Ziye
    Lu, Yao
    Wen, Junhai
    Li, Cuifen
    COMPUTERS IN BIOLOGY AND MEDICINE, 2012, 42 (06) : 651 - 656
  • [10] Analytical algorithm for 3D cone-beam SPECT reconstruction with uniform attenuation correction
    Bronnikov, A.V.
    IEEE Nuclear Science Symposium and Medical Imaging Conference, 1999, 3 : 1417 - 1423