An Anthropomorphic Breast Model for Breast Imaging Simulation and Optimization

被引:47
作者
Chen, Baiyu [1 ,2 ]
Shorey, Jamie [3 ]
Saunders, Robert S., Jr. [2 ]
Richard, Samuel [2 ]
Thompson, John [2 ]
Nolte, Loren W. [3 ]
Samei, Ehsan [1 ,2 ,4 ,5 ]
机构
[1] Duke Univ, Med Phys Grad Program, Durham, NC 27705 USA
[2] Duke Univ, Carl E Ravin Adv Imaging Labs, Dept Radiol, Durham, NC 27705 USA
[3] Duke Univ, Dept Elect & Comp Engn, Durham, NC 27705 USA
[4] Duke Univ, Dept Biomed Engn, Durham, NC 27705 USA
[5] Duke Univ, Dept Phys, Durham, NC 27705 USA
基金
美国国家科学基金会;
关键词
Anthropomorphic breast phantom; breast imaging; system optimization; Monte Carlo simulation; virtual phantoms; DIGITAL MAMMOGRAPHY; 3D SIMULATION; PHANTOM; TOMOSYNTHESIS; CONSPICUOUSNESS; ACQUISITION; COMPRESSION; MASS;
D O I
10.1016/j.acra.2010.11.009
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: Optimization studies for x-ray based breast imaging systems using computer simulation can greatly benefit from a phantom capable of modeling varying anatomical variability across different patients. This study aimed to develop a three-dimensional phantom model with realistic and randomizable anatomical features. Materials and Methods: A voxelized breast model was developed consisting of an outer layer of skin and subcutaneous fat, a mixture of glandular and adipose, stochastically generated ductal trees, masses, and microcalcifications. Randomized realization of the breast morphology provided a range of patient models. Compression models were included to represent the breast under various compression levels along different orientations. A Monte Carlo (MC) simulation code was adapted to simulate x-ray based imaging systems for the breast phantom. Simulated projections of the phantom at different angles were generated and reconstructed with iterative methods, simulating mammography, breast tomosynthesis, and computed tomography (CT) systems. Phantom dose maps were further generated for dosimetric evaluation. Results: Region of interest comparisons of simulated and real mammograms showed strong similarities in terms of appearance and features. Noise-power spectra of simulated mammographic images demonstrated that the phantom provided target properties for anatomical backgrounds. Reconstructed tomosynthesis and CT images and dose maps provided corresponding data from a single breast enabling optimization studies. Dosimetry result provided insight into the dose distribution difference between modalities and compression levels. Conclusion: The anthropomorphic breast phantom, combined with the MC simulation platform, generated a realistic model for a breast imaging system. The developed platform is expected to provide a versatile and powerful framework for optimizing volumetric breast imaging systems.
引用
收藏
页码:536 / 546
页数:11
相关论文
共 25 条
[1]   Assessing the risk of breast cancer. [J].
Armstrong, K ;
Eisen, A ;
Weber, B .
NEW ENGLAND JOURNAL OF MEDICINE, 2000, 342 (08) :564-571
[2]   Mammogram synthesis using a 3D simulation. 1. Breast tissue model and image acquisition simulation [J].
Bakic, PR ;
Albert, M ;
Brzakovic, D ;
Maidment, ADA .
MEDICAL PHYSICS, 2002, 29 (09) :2131-2139
[3]   Mammogram synthesis using a 3D simulation. II. Evaluation of synthetic mammogram texture [J].
Bakic, PR ;
Albert, M ;
Brzakovic, D ;
Maidment, ADA .
MEDICAL PHYSICS, 2002, 29 (09) :2140-2151
[4]   Mammogram synthesis using a three-dimensional simulation. III. Modeling and evaluation of the breast ductal network [J].
Bakic, PR ;
Albert, M ;
Brzakovic, D ;
Maidment, ADA .
MEDICAL PHYSICS, 2003, 30 (07) :1914-1925
[5]   DISTRIBUTION OF THE GEOMETRIC PARAMETERS OF HUMAN AORTIC BIFURCATIONS [J].
BARGERON, CB ;
HUTCHINS, GM ;
MOORE, GW ;
DETERS, OJ ;
MARK, FF ;
FRIEDMAN, MH .
ARTERIOSCLEROSIS, 1986, 6 (01) :109-113
[6]   A three-dimensional breast software phantom for mammography simulation [J].
Bliznakova, K ;
Bliznakov, Z ;
Bravou, V ;
Kolitsi, Z ;
Pallikarakis, N .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (22) :3699-3719
[7]   Producing lesions for hybrid mammograms: Extracted tumours and simulated microcalcifications [J].
Burgess, AE ;
Chakraborty, S .
MEDICAL IMAGING 1999: IMAGE PERCEPTION AND PERFORMANCE, 1999, 3663 :316-322
[8]   Human observer detection experiments with mammograms and power-law noise [J].
Burgess, AE ;
Jacobson, FL ;
Judy, PF .
MEDICAL PHYSICS, 2001, 28 (04) :419-437
[9]   DEVELOPMENT OF AN ANTHROPOMORPHIC BREAST PHANTOM [J].
CALDWELL, CB ;
YAFFE, MJ .
MEDICAL PHYSICS, 1990, 17 (02) :273-280
[10]   Optimized image acquisition for breast tomosynthesis in projection and reconstruction space [J].
Chawla, Amarpreet S. ;
Lo, Joseph Y. ;
Baker, Jay A. ;
Samei, Ehsan .
MEDICAL PHYSICS, 2009, 36 (11) :4859-4869