Comparison of Diffusion-Weighted Imaging in the Human Brain Using Readout-Segmented EPI and PROPELLER Turbo Spin Echo With Single-Shot EPI at 7 T MRI

被引:31
作者
Kida, Ikuhiro [1 ,2 ]
Ueguchi, Takashi [1 ,2 ]
Matsuoka, Yuichiro [1 ,2 ]
Zhou, Kun [3 ]
Stemmer, Alto [4 ]
Porter, David [5 ]
机构
[1] Natl Inst Informat & Commun Technol, Ctr Informat & Neural Networks, 1-4 Yamadaoka, Suita, Osaka 5650871, Japan
[2] Osaka Univ, Grad Sch Frontier Biosci, Osaka, Japan
[3] Siemens Shenzhen Magnet Resonance Ltd, Shenzhen, Peoples R China
[4] Siemens Healthcare GmbH, Erlangen, Germany
[5] Fraunhofer Inst Med Image Comp MEVIS, Bremen, Germany
关键词
diffusion-weighted imaging; 7; T; EPI; BLADE; RESOLVE; ADC; brain; PROPELLER; readout-segmented EPI; REDUCTION; ISCHEMIA;
D O I
10.1097/RLI.0000000000000248
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives The purpose of the present study was to compare periodically rotated overlapping parallel lines with enhanced reconstruction-type turbo spin echo diffusion-weighted imaging (pTSE-DWI) and readout-segmented echo planar imaging (rsEPI-DWI) with single-shot echo planar imaging (ssEPI-DWI) in a 7 T human MR system. We evaluated the signal-to-noise ratio (SNR), image distortion, and apparent diffusion coefficient values in the human brain. Materials and Methods Six healthy volunteers were included in this study. The study protocol was approved by our institutional review board. All measurements were performed at 7 T using pTSE-DWI, rsEPI-DWI, and ssEPI-DWI sequences. The spatial resolution was 1.2 x 1.2 mm(2) in-plane with a 3-mm slice thickness. Signal-to-noise ratio was measured using 2 scans. Results The ssEPI-DWI sequence showed significant image blurring, whereas pTSE-DWI and rsEPI-DWI sequences demonstrated high image quality with low geometrical distortion compared with reference T2-weighted, turbo spin echo images. Signal loss in ventral regions near the air-filled paranasal sinus/nasal cavity was found in ssEPI-DWI and rsEPI-DWI but not pTSE-DWI. The apparent diffusion coefficient values for ssEPI-DWI were 824 17 x 10(-6) and 749 +/- 25 x 10(-6) mm(2)/s in the gray matter and white matter, respectively; the values obtained for pTSE-DWI were 798 +/- 21 x 10(-6) and 865 +/- 40 x 10(-6) mm(2)/s; and the values obtained for rsEPI-DWI were 730 +/- 12 x 10(-6) and 722 +/- 25 x 10(-6) mm(2)/s. The pTSE-DWI images showed no additional distortion comparison to the T2-weighted images, but had a lower SNR than ssEPI-DWI and rsEPI-DWI. The rsEPI-DWI sequence provided high-quality images with minor distortion and a similar SNR to ssEPI-DWI. Conclusions Our results suggest that the benefits of the rsEPI-DWI and pTSE-DWI sequences, in terms of SNR, image quality, and image distortion, appear to outweigh those of ssEPI-DWI. Thus, pTSE-DWI and rsEPI-DWI at 7 T have great potential use for clinical diagnoses. However, it is noteworthy that both sequences are limited by the scan time required. In addition, pTSE-DWI has limitations on the number of slices due to specific absorption rate. Overall, rsEPI-DWI is a favorable imaging sequence, taking into account the SNR and image quality at 7 T.
引用
收藏
页码:435 / 439
页数:5
相关论文
共 18 条
  • [1] Adachi M, 1999, AM J NEURORADIOL, V20, P1500
  • [2] Elimination of eddy current artifacts in diffusion-weighted echo-planar images: The use of bipolar gradients
    Alexander, AL
    Tsuruda, JS
    Parker, DL
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (06) : 1016 - 1021
  • [3] Dietrich O, 2001, MAGNET RESON MED, V45, P448, DOI 10.1002/1522-2594(200103)45:3<448::AID-MRM1059>3.0.CO
  • [4] 2-W
  • [5] Diffusion-Weighted Imaging in Patients With Acute Brain Ischemia at 3 T Current Possibilities and Future Perspectives Comparing Conventional Echoplanar Diffusion-Weighted Imaging and Fast Spin Echo Diffusion-Weighted Imaging Sequences Using BLADE (PROPELLER)
    Fries, Peter
    Runge, Val M.
    Kirchin, Miles A.
    Stemmer, Alto
    Naul, L. Gill
    Williams, Kenneth D.
    Reith, Wolfgang
    Buecker, Arno
    Schneider, Guenther
    [J]. INVESTIGATIVE RADIOLOGY, 2009, 44 (06) : 351 - 359
  • [6] Diffusion Imaging in Humans at 7T Using Readout-Segmented EPI and GRAPPA
    Heidemann, Robin M.
    Porter, David A.
    Anwander, Alfred
    Feiweier, Thorsten
    Heberlein, Keith
    Knoesche, Thomas R.
    Turner, Robert
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2010, 64 (01) : 9 - 14
  • [7] Helenius J, 2002, AM J NEURORADIOL, V23, P194
  • [8] CSF-SUPPRESSED QUANTITATIVE SINGLE-SHOT DIFFUSION IMAGING
    KWONG, KK
    MCKINSTRY, RC
    CHIEN, D
    CRAWLEY, AP
    PEARLMAN, JD
    ROSEN, BR
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1991, 21 (01) : 157 - 163
  • [9] DIFFUSION-WEIGHTED MAGNETIC-RESONANCE-IMAGING OF ACUTE FOCAL CEREBRAL-ISCHEMIA - COMPARISON OF SIGNAL INTENSITY WITH CHANGES IN BRAIN WATER AND NA+,K+-ATPASE ACTIVITY
    MINTOROVITCH, J
    YANG, GY
    SHIMIZU, H
    KUCHARCZYK, J
    CHAN, PH
    WEINSTEIN, PR
    [J]. JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1994, 14 (02) : 332 - 336
  • [10] Evaluation of a Modified Stejs']jskal-Tanner Diffusion Encoding Scheme, Permitting a Marked Reduction in TE, in Diffusion-Weighted Imaging of Stroke Patients at 3 T
    Morelli, John N.
    Runge, Val M.
    Feiweier, Thorsten
    Kirsch, John E.
    Williams, Kenneth W.
    Attenberger, Ulrike I.
    [J]. INVESTIGATIVE RADIOLOGY, 2010, 45 (01) : 29 - 35