Estimated ECG Subtraction method for removing ECG artifacts in esophageal recordings of diaphragm EMG

被引:11
|
作者
Jonkman, Annemijn H. [1 ,2 ]
Juffermans, Ricardo [1 ,2 ]
Doorduin, Jonne [4 ,5 ]
Heunks, Leo M. A. [1 ,2 ]
Harlaar, Jaap [3 ,6 ]
机构
[1] Univ Amsterdam, Med Ctr, Dept Intens Care Med, Locat VUmc, Amsterdam, Netherlands
[2] Amsterdam Cardiovasc Sci Res Inst, Amsterdam, Netherlands
[3] Delft Univ Technol, Dept Biomech Engn, Delft, Netherlands
[4] Radboudumc, Donders Inst Brain Cognit & Behav, Nijmegen, Netherlands
[5] Radboudumc, Dept Neurol, Nijmegen, Netherlands
[6] Erasmus MC, Dept Orthoped, Rotterdam, Netherlands
关键词
Diaphragm EMG; ECG contamination; Template subtraction; Wavelet filter; MECHANICAL VENTILATION; SIGNALS; CONTAMINATION; CANCELLATION; ENHANCEMENT; VALIDATION;
D O I
10.1016/j.bspc.2021.102861
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The accuracy of diaphragm electromyogram (EMGdi) derived parameters, as used in critically ill intensive care unit (ICU) patients, can be compromised due to electrocardiographic (ECG) interference in the EMGdi signal. Removal of ECG contamination from the esophageal recordings of the EMGdi is challenging due to spectral overlapping of EMG and ECG signals and because of variability in ECG shape and amplitude. Therefore, we designed an Estimated ECG Subtraction (EES) method, based on three steps: (1) identification of the timing of the ECG artifact without an ECG reference channel, (2) estimation of the normalized ECG, considering the EMGdi as noise, and (3) subtraction of the denormalized ECG estimate from the EMGdi recordings. We evaluated the EES method against the use of a single wavelet-based adaptive filter. Using EMGdi signals of ten ICU patients and simulated contaminated EMG, we demonstrated that the EES method yields uncontaminated EMGdi, and showed that it is more effective than a wavelet-based adaptive filter only. Implementation of this technique may offer means to improve diaphragm activity monitoring and control in clinical practice.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Morphological ECG subtraction method for removing ECG artifacts from diaphragm EMG
    Guo, Liang
    Li, Zhi-Wei
    Zhang, Han
    Li, Shuang-Miao
    Zhang, Jian-Heng
    TECHNOLOGY AND HEALTH CARE, 2023, 31 : S333 - S345
  • [2] Removing Movement Artifacts from Equine ECG recordings acquired with Textile Electrodes
    Lanata, Antonio
    Guidi, Andrea
    Baragli, Paolo
    Paradiso, Rita
    Valenza, Gaetano
    Scilingo, Enzo Pasquale
    2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2015, : 1955 - 1958
  • [3] A novel method based on shifted rank-1 reconstruction for removing EMG artifacts in ECG signals
    Chen, Xieqi
    Zheng, Shubin
    Peng, Lele
    Zhong, Qianwen
    He, Liu
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2023, 85
  • [4] Analysis of diaphragm EMG signals: Comparison of gating vs subtraction for removal of ECG contamination
    Bartolo, A
    Roberts, C
    Dzwonczyk, RR
    Goldman, E
    JOURNAL OF APPLIED PHYSIOLOGY, 1996, 80 (06) : 1898 - 1902
  • [5] Method of removing ECG artifacts from EEG based on ICA
    Li, XY
    ICEMI'2003: PROCEEDINGS OF THE SIXTH INTERNATIONAL CONFERENCE ON ELECTRONIC MEASUREMENT & INSTRUMENTS, VOLS 1-3, 2003, : 1900 - 1903
  • [6] Utilizing SimEMG to Evaluate the Effectiveness of Adaptive Filters in Removing EMG Contamination from ECG Recordings
    Al-Safi, Amean
    ELEKTROTEHNISKI VESTNIK, 2024, 91 (05): : 238 - 246
  • [7] Utilizing SimEMG to Evaluate the Effectiveness of Adaptive Filters in Removing EMG Contamination from ECG Recordings
    Al-Safi, Amean
    Elektrotehniski Vestnik/Electrotechnical Review, 2024, 91 (05): : 238 - 246
  • [8] Removing ECG contamination from EMG recordings: A comparison of ICA-based and other filtering procedures
    Willigenburg, Nienke W.
    Daffertshofer, Andreas
    Kingma, Idsart
    van Dieen, Jaap H.
    JOURNAL OF ELECTROMYOGRAPHY AND KINESIOLOGY, 2012, 22 (03) : 485 - 493
  • [9] Enhancement of signal quality in esophageal recordings of diaphragm EMG
    Sinderby, CA
    Beck, JC
    Lindstrom, LH
    Grassino, AE
    JOURNAL OF APPLIED PHYSIOLOGY, 1997, 82 (04) : 1370 - 1377
  • [10] An automated ECG-artifact removal method for trunk muscle surface EMG recordings
    Mak, Joseph N. F.
    Hu, Yong
    Luk, Keith D. K.
    MEDICAL ENGINEERING & PHYSICS, 2010, 32 (08) : 840 - 848