Determination of Cr(III) and Cr(VI) at Sub-ppb Levels in Water with Solid-Phase Extraction/Metal Furnace Atomic Absorption Spectrometry

被引:35
作者
Inui, Tetsuo [1 ]
Fujita, Kazuhiro [1 ]
Kitano, Masaru [1 ]
Nakamura, Toshihiro [1 ]
机构
[1] Meiji Univ, Dept Appl Chem, Tama Ku, Kawasaki, Kanagawa 2148571, Japan
关键词
DUAL MINI-COLUMNS; PLASMA-MASS SPECTROMETRY; COMPLEX-FORMATION; ICP-AES; SPECIATION; CHROMIUM(III); SYSTEM; PRECONCENTRATION; ACID; RESIN;
D O I
10.2116/analsci.26.1093
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple method using solid-phase extraction combined with metal furnace atomic absorption spectrometry was developed for the determination of Cr(III) and Cr(VI) at sub-ppb levels in water. A 500-ml water sample was adjusted to pH 3 with nitric acid and then passed through an iminodiacetate extraction disk placed on a cation-exchange extraction disk at a flow rate of 20 - 40 ml min(-1) for concentrating Cr(III) The filtrate was adjusted to pH 10 with aqueous ammonia and then passed through an anion-exchange extraction disk at a flow rate of 2 ml min(-1) for concentrating Cr(VI) The Cr(III) and Cr(VI) collected were eluted with 40 ml of 3 mol nitric acid for Cr(III) and 40 ml of 1 g diphenylcarbazide solution for Cr(VI). Each eluate was diluted to 50 ml with deionized water and injected into a U-type tungsten board on the metal furnace The calibration curves of Cr(III) and Cr(VI) showed good linearity in the range of 0.1 - 0.5 ng The detection limits corresponding to three times the standard deviation (n = 5) of blank values were 8 1 pg for both Cr(III) and Cr(VI) The analytical value of total Cr (Cr(III) + Cr(VI)) in certified reference material of river water (JSAC 0302-3) was in good agreement with the reference value The recovery test for 0 50 kg (1.00 mu g l(-1)) of Cr(III) and Cr(VI) added to 500 ml of the water samples showed sufficient values (98.1 - 106%). except for river water sampled downstream due to relatively higher COD(Mn) value. The relative standard deviations (n = 5) were less than 5% for both Cr(III) and Cr(VI).
引用
收藏
页码:1093 / 1098
页数:6
相关论文
共 35 条
[11]   Development of certified reference materials accomplished by the Japan Society for Analytical Chemistry [J].
Hobo, Toshiyuki ;
Iida, Yoshio ;
Ishibashi, Yo-Ichi ;
Okamoto, Kensaku ;
Kawase, Akira ;
Nakamura, Toshihiro ;
Nakamura, Hiroshi ;
Hirai, Shoji ;
Matsuda, Rieko ;
Yamasad, Shin-ichi ;
Yomota, Chikako ;
Ono, Akihiro ;
Kakita, Kazutoshi ;
Sakata, Mamoru ;
Takimoto, Ken-ichi .
BUNSEKI KAGAKU, 2008, 57 (06) :363-392
[12]   One-pot speciation method for Cr(III) and Cr(VI) with APCD/DIBK extraction system based on difference in rate of complex formation [J].
Honma, Y .
BULLETIN OF THE CHEMICAL SOCIETY OF JAPAN, 2003, 76 (07) :1397-1402
[13]  
IDA I, 2008, BUNSEKI, P206
[14]   Speciation analysis of chromium in environmental water samples from Niigata Prefecture by using ion-exchange ICP-AES [J].
Iwasaki, Y ;
Kano, N ;
Omori, K ;
Imaizumi, H ;
Ishizuka, T .
BUNSEKI KAGAKU, 2004, 53 (09) :1015-1019
[15]  
Japanese Industrial Standards Committee, 2008, K0102 JIS
[16]   Atomic absorption spectrometry using tungsten and molybdenum tubes as metal atomizer [J].
Kaneco, Satoshi ;
Katsumata, Hideyuki ;
Suzuki, Tohru ;
Ohta, Kiyohisa .
BUNSEKI KAGAKU, 2007, 56 (07) :535-546
[17]   Preparation and certification of the new reference materials, river waters (JS']JSAC 0301-1, JS']JSAC 0301-2, JS']JSAC 0302) for determination of inorganic components [J].
Kawase, Akira ;
Ohashi, Kazuo ;
Akabane, Isoko ;
Kuroiwa, Takayoshi ;
Murayama, Mariko ;
Yoshimura, Hajime ;
Yoshinaga, Jun ;
Uemoto, Michihisa ;
Ono, Akihiro ;
Kakita, Kazutoshi ;
Sakata, Mamoru .
BUNSEKI KAGAKU, 2006, 55 (11) :875-882
[18]   Decomposition characteristics of various organic compounds in chemical oxygen demand measurement [J].
Kikuchi, Yoichi ;
Kakizaki, Satomi ;
Koizumi, Die .
BUNSEKI KAGAKU, 2007, 56 (10) :857-862
[19]   Chromium occurrence in the environment and methods of its speciation [J].
Kotas, J ;
Stasicka, Z .
ENVIRONMENTAL POLLUTION, 2000, 107 (03) :263-283
[20]   Simple, rapid and sensitive determination of Cr,(VI) by spectropho tome tric flow injection system utilizing activated alumina column for in-line separation, preconcentration and reaction [J].
Kubota, Toshiaki ;
Yamane, Takeshi .
BUNSEKI KAGAKU, 2007, 56 (11) :927-932