Over 10 years with an implantable cardioverter-defibrillator - a long term follow-up of 60 patients

被引:0
|
作者
Sterlinski, Maciej [1 ]
Przybylski, Andrzej [1 ]
Gepner, Katarzyna [1 ]
Syska, Pawel [1 ]
Maciag, Aleksander [1 ]
Lewandowski, Michal [1 ]
Kowalik, Ilona [1 ]
Pytkowski, Manusz [1 ]
Sitkowska-Rysiak, Ewa [1 ]
Lichomski, Jerzy [1 ]
Szwed, Hanna [1 ]
Sadowski, Zygmunt [1 ]
机构
[1] Inst Cardiol, PL-02637 Warsaw, Poland
关键词
implantable cardioverter defibrillator; long term observation; survival; appropriate and inappropriate interventions; complications; SUDDEN CARDIAC DEATH; MYOCARDIAL-INFARCTION; TEMPORAL TRENDS; UNITED-STATES; HIGH-RISK; THERAPY; PREVENTION; EXPERIENCE; RECIPIENTS; SURVIVAL;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Transvenous implantable cardioverter defibrillators (ICD) have been implanted in Poland since 1995 As the method spreads it is important to consider its long-term benefits and disadvantages Aim To assess survival efficacy and complication rate in ICD patients, who received the device more than ten years earlier Methods Retrospective analysis of 60 ICD patients implanted between 1995-1999 Results There were 42 (70%) males, mean age 50 6 +/- 16 4 years In 59 patients ICD was implanted for sudden cardiac death (SCD) secondary prevention Thirty eight patients (34 M, 63 3%) had coronary artery disease (CAD) The CAD was diagnosed in 89 5% of males and 10 5% of females (p < 0 0001) Mean follow-up time was 75 4 +/- 34 7 months During this time 22 patients died (37%, 19 M, 3 F) Three deaths were SCD Mean one-year mortality was 6 7% Deaths were more frequent among males 45 2% vs 16 7%, p < 0 005 In CAD mortality was higher than in non-CAD patients (50% vs 13 6% p < 0 005) Appropriate ICD discharges in the ventricular fibrillation (VF) zone occurred in 35 (58%) patients and in ventricular tachycardia (VT) zone - in 26 (43%) patients Mean intervention rate per year was 3 7 for VF and 0 6 for VT Complications occurred in 27 (45%) patients and 5 (8%) of them had no ICD intervention during follow-up In 5 patients more than one complication was diagnosed There were inappropriate discharges in 15 (25%) patients, 11 (18%) had electrical storm and ICD related infections were noted in 3 (5%) patients During the penoperative period, lead revisions were done in 4 patients, in 3 with discharges induced by T-wave oversensing and in one with lead dislocation Four cases of lead failure occurred during follow up, requiring new lead implantation In 4 patients, electrical storm (3 patients) and supraventricular tachycardia with ICD discharges (1 patient) were treated with radiofrequency ablation Only 10 (17%) patients did not demonstrate any ICD interventions or ICD related complications Conclusions 1 ICD interventions caused by malignant ventricular arrhythmias occurred in 75% patients with the device implanted more than 10 years earlier 2 Almost a half of the analysed population suffered from complications and side effects related to implanted ICD and they were present in 8% of subjects without ICD intervention Neither ICD interventions nor device related adverse events were recorded in 17% of patients
引用
收藏
页码:1023 / 1031
页数:9
相关论文
共 50 条
  • [1] Long-term follow-up of children and young adults treated with implantable cardioverter-defibrillator: the authors' own experience with optimal implantable cardioverter-defibrillator programming
    Lewandowski, Michal
    Sterlinski, Maciej
    Maciag, Aleksander
    Syska, Pawel
    Kowalik, Ilona
    Szwed, Hanna
    Chojnowska, Lidia
    Przybylski, Andrzej
    EUROPACE, 2010, 12 (09): : 1245 - 1250
  • [2] Implantable cardioverter-defibrillator in Brugada syndrome: Long-term follow-up
    El-Battrawy, Ibrahim
    Roterberg, Gretje
    Liebe, Volker
    Ansari, Uzair
    Lang, Siegfried
    Zhou, Xiaobo
    Borggrefe, Martin
    Akin, Ibrahim
    CLINICAL CARDIOLOGY, 2019, 42 (10) : 958 - 965
  • [3] Long-Term Follow-Up of Patients with Chronic Chagas Disease and Implantable Cardioverter-Defibrillator
    Moreira Pereira, Francisca Tatiana
    Rocha, Eduardo Arrais
    Martins Monteiro, Marcelo De Paula
    Rocha Neto, Almino Cavalcante
    Daher, Elisabeth De Francesco
    Martins Rodrigues Sobrinho, Carlos Roberto
    Pires Neto, Roberto Da Justa
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2014, 37 (06): : 751 - 756
  • [4] Relation of multicenter automatic defibrillator implantation trial implantable cardioverter-defibrillator score with long-term cardiovascular events in patients with implantable cardioverter-defibrillator
    Uslu, Abdulkadir
    Dogan, Cem
    Duman, Hakan
    Tanboga, Ibrahim Halil
    Askin, Lutfu
    Sevimli, Serdar
    NORTHERN CLINICS OF ISTANBUL, 2019, 6 (01) : 40 - 47
  • [5] Predictors of Appropriate Implantable Cardioverter-Defibrillator Therapy During Long-Term Follow-up of Patients With Coronary Artery Disease
    Blumer, Johannes
    Wolber, Thomas
    Hellermann, Jens
    Holzmeister, Johannes
    Binggeli, Christian
    Duru, Firat
    Brunckhorst, Corinna
    INTERNATIONAL HEART JOURNAL, 2009, 50 (03) : 313 - 321
  • [6] Clinical follow-up of patients with implantable cardioverter-defibrillator
    da Fonseca, Silvia Martelo Souza
    Belo, Luiz Gustavo
    Carvalho, Hecio
    Araujo, Nilson
    Munhoz, Claudio
    Siqueira, Leonardo
    Maciel, Washington
    Andrea, Eduardo
    Atie, Jacob
    ARQUIVOS BRASILEIROS DE CARDIOLOGIA, 2007, 88 (01) : 8 - 16
  • [7] The Implantable Cardioverter-Defibrillator Minimalist An Approach to Patient Follow-Up and Management of Implantable Defibrillators
    Sweeney, Michael O.
    CIRCULATION, 2012, 126 (03) : 369 - 377
  • [8] Cardiac Sarcoidosis: Electrophysiological Outcomes on Long-Term Follow-Up and the Role of the Implantable Cardioverter-Defibrillator
    Mohsen, Amr
    Jimenez, Alejandro
    Hood, Robert E.
    Dickfeld, Timm
    Saliaris, Anastasios
    Shorofsky, Stephen
    Saba, Magdi M.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2014, 25 (02) : 171 - 176
  • [9] The Wearable Cardioverter-Defibrillator: Experience in 153 Patients and a Long-Term Follow-Up
    Rosenkaimer, Stephanie L.
    El-Battrawy, Ibrahim
    Dreher, Tobias C.
    Gerhards, Stefan
    Roeger, Susanne
    Kuschyk, Juergen
    Borggrefe, Martin
    Akin, Ibrahim
    JOURNAL OF CLINICAL MEDICINE, 2020, 9 (03)
  • [10] Follow-up of the patient with automatic implantable cardioverter defibrillator
    Cruz Cardentey, Marleny
    Prieto Guerra, Maidelis
    Gutierrez Lopez, Alain
    Mengana Betancourt, Ana
    Ramos Cordero, Michel Y.
    Tannis Wilson, Eko
    Tamargo Barbeito, Teddy O.
    CORSALUD, 2020, 12 (03): : 283 - 291