Mechanical cavopulmonary assist for the univentricular fontan circulation using a novel folding propeller blood pump

被引:26
作者
Throckmorton, Amy L.
Ballman, Kimberly K.
Myers, Cynthia D.
Litwak, Kenneth N.
Frankel, Steven H.
Rodefeld, Mark D.
机构
[1] Virginia Commonwealth Univ, Dept Engn Mech, Richmond, VA 23284 USA
[2] Indiana Univ Sch Med, James Whitcomb Riley Hosp Children, Dept Surg, Sect Cadiothoracic Surg, Indianapolis, IN USA
[3] Indiana Univ Sch Med, James Whitcomb Riley Hosp Children, Dept Pediat, Indianapolis, IN USA
[4] Univ Louisville, Dept Surg, Louisville, KY 40292 USA
[5] Purdue Univ, Sch Mech Engn, W Lafayette, IN 47907 USA
关键词
D O I
10.1097/MAT.0b013e318159d14b
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A blood pump specifically designed to operate in the unique anatomic and physiologic conditions of a cavopulmonary connection has never been developed. Mechanical augmentation of cavopulmonary blood flow in a univentricular circulation would reduce systemic venous pressure, increase preload to the single ventricle, and temporarily reproduce a scenario analogous to the normal two-ventricle circulation. We hypothesize that a folding propeller blood pump would function optimally in this cavopulmonary circulation. The hydraulic performance of a two-bladed propeller prototype was characterized in an experimental flow loop using a blood analog fluid for 0.5-3.5 Ipm at rotational speeds of 3,6004,000 rpm. We also created five distinctive blood pump designs and evaluated their hydraulic performance using computational fluid dynamics (CFD). The two-bladed prototype performed well over the design range of 0.5-3.5 Ipm, producing physiologic pressure rises of 5-18 mm Hg. Building upon this proof-of-concept testing, the CFD analysis of the five numerical models predicted a physiologic pressure range of 5-40 mm Hg over 0.5-4 Ipm for rotational speeds of 3,000-7,000 rpm. These preliminary propeller designs and the two-bladed prototype achieved the expected hydraulic performance. Optimization of these configurations will reduce fluid stress levels, remove regions of recirculation, and improve the hydraulic performance of the folding propeller. This propeller design produces the physiologic pressures and flows that are in the ideal range to mechanically support the cavopulmonary circulation and represents an exciting new therapeutic option for the support of a univentricular Fontan circulation.
引用
收藏
页码:734 / 741
页数:8
相关论文
共 34 条
[1]   Assessment of hemolysis related quantities in a microaxial blood pump by computational fluid dynamics [J].
Apel, J ;
Paul, R ;
Klaus, S ;
Siess, T ;
Reul, H .
ARTIFICIAL ORGANS, 2001, 25 (05) :341-347
[2]   3-DIMENSIONAL NUMERICAL PREDICTION OF STRESS LOADING OF BLOOD PARTICLES IN A CENTRIFUGAL PUMP [J].
BLUDSZUWEIT, C .
ARTIFICIAL ORGANS, 1995, 19 (07) :590-596
[3]   MODEL FOR A GENERAL MECHANICAL BLOOD DAMAGE PREDICTION [J].
BLUDSZUWEIT, C .
ARTIFICIAL ORGANS, 1995, 19 (07) :583-589
[4]  
BOYD J, 2004, 90 ANN CLIN C AM COL
[5]  
Breslin JB., 1994, HYDRODYNAMICS SHIP P
[6]   Computational fluid dynamics as a development tool for rotary blood pumps [J].
Burgreen, GW ;
Antaki, JF ;
Wu, ZJ ;
Holmes, AJ .
ARTIFICIAL ORGANS, 2001, 25 (05) :336-340
[7]  
Cebeci T., 2004, Analysis of Turbulent Flows with Computer Programs
[8]   Ventricular dysfunction of the functionally univentricular heart: management and outcomes [J].
Gewillig, Marc .
CARDIOLOGY IN THE YOUNG, 2005, 15 :31-34
[9]   ESTIMATION OF SHEAR STRESS-RELATED BLOOD DAMAGE IN HEART-VALVE PROSTHESES - INVITRO COMPARISON OF 25 AORTIC VALVES [J].
GIERSIEPEN, M ;
WURZINGER, LJ ;
OPITZ, R ;
REUL, H .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1990, 13 (05) :300-306
[10]   Development of design methods for a centrifugal blood pump with a fluid dynamic approach: Results in hemolysis tests [J].
Masuzawa, T ;
Tsukiya, T ;
Endo, S ;
Tatsumi, E ;
Taenaka, Y ;
Takano, H ;
Yamane, T ;
Nishida, M ;
Asztalos, B ;
Miyazoe, Y ;
Ito, K ;
Sawairi, T ;
Konishi, Y .
ARTIFICIAL ORGANS, 1999, 23 (08) :757-761