Validation of the blood pressure measurement device Erkameter 125 PRO according to the European Society of Hypertension International Protocol revision 2010

被引:0
|
作者
Beime, Beate [1 ]
Deutsch, Cornelia [1 ]
Krueger, Ralf [1 ]
Zimmermann, Erik [2 ]
Bramlage, Peter [1 ]
机构
[1] Inst Pharmacol & Prevent Med, Bahnhofstr 20, D-49661 Cloppenburg, Germany
[2] ERKA Kallmeyer Med Tech, Bad Tolz, Germany
关键词
blood pressure measuring device; European Society of Hypertension; validation; SELF-MEASUREMENT; MONITOR; PREVALENCE; AWARENESS;
D O I
10.1097/MBP.0000000000000216
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective The aim of the present study was to validate the blood pressure (BP) measurement device Erkameter 125 PRO according to the International Protocol revision 2010 of the European Society of Hypertension (ESH-IP2). Patients and methods In 33 patients aged 32-79 years, BP measurements were performed alternately using the Erkameter 125 PRO and the reference mercury sphygmomanometer according to ESH-IP revision 2010. For the analysis, a total of 99 comparisons were included. Results All absolute differences between the test device and the reference were within 10 mmHg for systolic blood pressure (SBP), and all except one for diastolic blood pressure (DBP). A total of 93 out of 99 comparisons for SBP showed an absolute difference within 5 mmHg and 92 out of 99 for DBP. The mean +/- SD difference between the Erkameter 125 PRO and the standard reference was -0.5 +/- 3.5 mmHg for SBP and 0.5 +/- 3.5 mmHg for DBP. As to part 2 of ESH-IP 2010, all patients had a minimum of two out of three measurements within 5 mmHg difference for SBP and 31 out of 33 patients for DBP. Conclusion The Erkameter 125 PRO fulfilled the requirements of parts 1 and 2 of the ESH-IP revision 2010 and can be recommended for office BP measurements in adults. Copyright (C) 2016 Wolters Kluwer Health, Inc. All rights reserved.
引用
收藏
页码:356 / 360
页数:5
相关论文
共 50 条
  • [1] Validation of the blood pressure measurement device Beurer BM 28 according to the European Society of Hypertension International Protocol revision 2010
    Deutsch, Cornelia
    Bramlage, Carsten
    Botta, Beate
    Krueger, Ralf
    Forstner, Klaus
    Bramlage, Peter
    Beime, Beate
    BLOOD PRESSURE MONITORING, 2021, 26 (04) : 292 - 298
  • [2] Validation of the Thermor BIOS BD215 device for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010
    Benetti, Elisabetta
    Fania, Claudio
    Hernandez, Veronica Marquez
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2014, 19 (03) : 176 - 179
  • [3] Validation of the A&D UM-201 device for office blood pressure measurement according to the European Society of Hypertension International Protocol Revision 2010
    Fania, Claudio
    Albertini, Federica
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2017, 22 (04) : 234 - 237
  • [4] Validation of the A&D UM-211 device for office blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010
    Fania, Claudio
    Albertini, Federica
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2017, 22 (05) : 302 - 305
  • [5] Validation of the A&D BP UA-651 device for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010
    Benetti, Elisabetta
    Fania, Claudio
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2014, 19 (01) : 50 - 53
  • [6] Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
    Beime, Beate
    Deutsch, Cornelia
    Krueger, Ralf
    Wolf, Andreas
    Mueller, Peter
    Hammel, Gertrud
    Bramlage, Peter
    EUROPEAN JOURNAL OF PEDIATRICS, 2017, 176 (05) : 573 - 580
  • [7] Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
    Beate Beime
    Cornelia Deutsch
    Ralf Krüger
    Andreas Wolf
    Peter Müller
    Gertrud Hammel
    Peter Bramlage
    European Journal of Pediatrics, 2017, 176 : 573 - 580
  • [8] Validation of the custo screen 400 ambulatory blood pressure-monitoring device according to the European Society of Hypertension International Protocol revision 2010
    Bramlage, Peter
    Deutsch, Cornelia
    Krueger, Ralf
    Wolf, Andreas
    Mueller, Peter
    Zwingers, Thomas
    Beime, Beate
    Mengden, Thomas
    VASCULAR HEALTH AND RISK MANAGEMENT, 2014, 10 : 303 - 309
  • [9] Validation of the Hingmed WBP-02A device for ambulatory blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Fania, Claudio
    Vezzu, Leonardo
    Lazzaretto, Ilaria
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2019, 24 (03) : 151 - 154
  • [10] Validation of the A&D BP UB-542 wrist device for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010
    Saladini, Francesca
    Benetti, Elisabetta
    Fania, Claudio
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2013, 18 (04) : 219 - 222