Estimated radiation dose according to the craniocaudal angle in cerebral digital subtraction angiography: Patient and phantom study

被引:2
作者
Song, Yunsun [1 ]
Kim, Youngeun [2 ]
Han, Seongsik [2 ]
Kim, Tae Il [2 ]
Choi, Joon Ho [2 ]
Maeng, Jun Young [2 ]
Choi, Youngrak [3 ]
Lee, Deok Hee [1 ]
机构
[1] Univ Ulsan, Coll Med, Man Med Ctr, Dept Radiol,Res Inst Radiol, 88,Olymp Ro 43 Gil, Seoul 05505, South Korea
[2] Univ Ulsan, Asan Med Ctr, Coll Med, Dept Radiol Technol, Seoul, South Korea
[3] Siemens Healthineers Ltd, Seoul, South Korea
关键词
Radiation dose; Cerebral angiogram; Digital subtraction angiography; Tube angulation; IMAGE QUALITY; REDUCTION;
D O I
10.1016/j.neurad.2019.07.003
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background and purpose. - Routine use of cranial angulation with 15-20 degrees, craniocaudal angled (CC) view, for cerebral digital subtraction angiography (DSA) helps minimize bone subtraction artifacts with less overlapping of the vessels, however, it may increase the radiation dose. We designed the phantom and patient studies to determine the effect of the angulation to the radiation dose and the feasibility of true posteroanterior angled (PA) view, in cerebral DSA. Materials and methods. - In the phantom study, frontal DSA was simulated with variable angulations. In the patient study with thirty-one subjects, one internal carotid arteriogram was obtained with the CC view and the other, PA view in every patient. The dose-area product (DAP) and reference air-kerma (AK) were measured and compared between the angles. A qualitative analysis was performed to assess the diagnostic performance of the DSA over the angles. Results. - The phantom study confirmed that the greater craniocaudal angles caused higher radiation exposure. Especially, the radiation dose (AK) of the CC view was 5.4% higher than that of the PA view. In the patient study, the radiation dose of the PA view was significantly lower compared to the CC view (1.44 vs. 1.63 mGy, AK). In 4 patients, the dose particularly jumped when applying the CC view as the copper filter was automatically removed. The diagnostic ability of the DSA with the PA view tended to be higher without significance. Conclusions. - In a daily routine cerebral angiography, a simple modification of the angle may help to minimize the radiation dose. (C) 2019 Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:345 / 350
页数:6
相关论文
共 46 条
  • [31] Comprehensive organ-specific radiation dose mapping in coronary angiography using the Alderson Rando phantom: An experimental approach to optimize patient safety
    Kayaokay, Duygu Tuncman
    Kesmezacar, Fahrettin Fatih
    Yeyin, Nami
    Karacam, Songul Cavdar
    Gunay, Osman
    Karabag, Turgut
    Demir, Mustafa
    Issa, Bashar
    Tekin, H. O.
    Almisned, Ghada
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2025, 233
  • [32] Abdominal CT during pregnancy: a phantom study on the effect of patient centring on conceptus radiation dose and image quality
    G. Solomou
    A. E. Papadakis
    J. Damilakis
    [J]. European Radiology, 2015, 25 : 911 - 921
  • [33] Effect of automatic tube voltage selection on image quality and radiation dose in abdominal CT angiography of various body sizes: A phantom study
    Schindera, S. T.
    Winklehner, A.
    Alkadhi, H.
    Goetti, R.
    Fischer, M.
    Gnannt, R.
    Szucs-Farkas, Z.
    [J]. CLINICAL RADIOLOGY, 2013, 68 (02) : E79 - E86
  • [34] Radiation dose and physical image quality in 128-section dual-source computed tomographic coronary angiography: a phantom study
    Matsubara, Kosuke
    Koshida, Haruka
    Sakuta, Keita
    Takata, Tadanori
    Horii, Junsei
    Iida, Hiroji
    Koshida, Kichiro
    Ichikawa, Katsuhiro
    Matsui, Osamu
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (05): : 252 - 261
  • [35] Halved contrast medium dose coronary dual-layer CT-angiography - phantom study of tube current and patient characteristics
    Kristiansen, C. H.
    Tetteroo, P. M.
    Dobrolinska, M. M.
    Lauritzen, P. M.
    Velthuis, B. K.
    Greuter, M. J. W.
    Sucha, D.
    de Jong, P. A.
    van der Werf, N. R.
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2024, 40 (04) : 931 - 940
  • [36] Cerebral Arteriovenous Malformations at 3.0 T Intraindividual Comparative Study of 4D-MRA in Combination With Selective Arterial Spin Labeling and Digital Subtraction Angiography
    Kukuk, Guido M.
    Hadizadeh, Dariusch R.
    Bostroem, Azize
    Gieseke, Jurgen
    Bergener, Julia
    Nelles, Michael
    Muertz, Petra
    Urbach, Horst
    Schild, Hans H.
    Willinek, Winfried A.
    [J]. INVESTIGATIVE RADIOLOGY, 2010, 45 (03) : 126 - 132
  • [37] Pseudo-continuous and territorial arterial spin labeling MRI for assessment of cerebral perfusion in moyamoya disease after revascularization: A comparative study with digital subtraction angiography
    Li, Juan
    Meng, Qinghu
    Huang, Ling
    Pylypenko, Dmytro
    Zhong, Hai
    [J]. HELIYON, 2024, 10 (17)
  • [38] Radiation dose, image quality, stenosis measurement, and CT densitometry using ECG-triggered coronary 64-MDCT angiography: A phantom study
    Horiguchi, Jun
    Kiguchi, Masao
    Fujioka, Chikako
    Shen, Yun
    Arie, Ryuichi
    Sunasaka, Kenichi
    Ito, Katsuhide
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (02) : 315 - 320
  • [39] Reduction of Radiation Dose to Eye Lens in Cerebral 3D Rotational Angiography Using Head Off-Centering by Table Height Adjustment: A Prospective Study
    Ryu, Jae-Chan
    Yoon, Jong-Tae
    Kim, Byung Jun
    Kim, Mi Hyeon
    Moon, Eun Ji
    Suh, Pae Sun
    Roh, Yun Hwa
    Moon, Hye Hyeon
    Kwon, Boseong
    Lee, Deok Hee
    Song, Yunsun
    [J]. KOREAN JOURNAL OF RADIOLOGY, 2023, 24 (07) : 681 - 689
  • [40] Image quality and radiation dose of low tube voltage 3rd generation dual-source coronary CT angiography in obese patients: a phantom study
    Felix G. Meinel
    Christian Canstein
    U. Joseph Schoepf
    Martin Sedlmaier
    Bernhard Schmidt
    Brett S. Harris
    Thomas G. Flohr
    Carlo N. De Cecco
    [J]. European Radiology, 2014, 24 : 1643 - 1650