Estimated radiation dose according to the craniocaudal angle in cerebral digital subtraction angiography: Patient and phantom study

被引:2
作者
Song, Yunsun [1 ]
Kim, Youngeun [2 ]
Han, Seongsik [2 ]
Kim, Tae Il [2 ]
Choi, Joon Ho [2 ]
Maeng, Jun Young [2 ]
Choi, Youngrak [3 ]
Lee, Deok Hee [1 ]
机构
[1] Univ Ulsan, Coll Med, Man Med Ctr, Dept Radiol,Res Inst Radiol, 88,Olymp Ro 43 Gil, Seoul 05505, South Korea
[2] Univ Ulsan, Asan Med Ctr, Coll Med, Dept Radiol Technol, Seoul, South Korea
[3] Siemens Healthineers Ltd, Seoul, South Korea
关键词
Radiation dose; Cerebral angiogram; Digital subtraction angiography; Tube angulation; IMAGE QUALITY; REDUCTION;
D O I
10.1016/j.neurad.2019.07.003
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background and purpose. - Routine use of cranial angulation with 15-20 degrees, craniocaudal angled (CC) view, for cerebral digital subtraction angiography (DSA) helps minimize bone subtraction artifacts with less overlapping of the vessels, however, it may increase the radiation dose. We designed the phantom and patient studies to determine the effect of the angulation to the radiation dose and the feasibility of true posteroanterior angled (PA) view, in cerebral DSA. Materials and methods. - In the phantom study, frontal DSA was simulated with variable angulations. In the patient study with thirty-one subjects, one internal carotid arteriogram was obtained with the CC view and the other, PA view in every patient. The dose-area product (DAP) and reference air-kerma (AK) were measured and compared between the angles. A qualitative analysis was performed to assess the diagnostic performance of the DSA over the angles. Results. - The phantom study confirmed that the greater craniocaudal angles caused higher radiation exposure. Especially, the radiation dose (AK) of the CC view was 5.4% higher than that of the PA view. In the patient study, the radiation dose of the PA view was significantly lower compared to the CC view (1.44 vs. 1.63 mGy, AK). In 4 patients, the dose particularly jumped when applying the CC view as the copper filter was automatically removed. The diagnostic ability of the DSA with the PA view tended to be higher without significance. Conclusions. - In a daily routine cerebral angiography, a simple modification of the angle may help to minimize the radiation dose. (C) 2019 Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:345 / 350
页数:6
相关论文
共 46 条
  • [21] Effect of kVp on image quality and accuracy in coronary CT angiography according to patient body size: a phantom study
    Lee, Sang Min
    Lee, Whal
    Chung, Jin Wook
    Park, Eun-Ah
    Park, Jae Hyung
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2013, 29 : 83 - 91
  • [22] Effect of X-ray Tube Parameters and Iodine Concentration on Image Quality and Radiation Dose in Cerebral Pediatric and Adult CT Angiography A Phantom Study
    Papadakis, Antonios E.
    Perisinakis, Kostas
    Raissaki, Maria
    Damilakis, John
    INVESTIGATIVE RADIOLOGY, 2013, 48 (04) : 192 - 199
  • [23] Automated Low-Kilovoltage Selection in Pediatric Computed Tomography Angiography Phantom Study Evaluating Effects on Radiation Dose and Image Quality
    Siegel, Marilyn Joy
    Ramirez-Giraldo, Juan Carlos
    Hildebolt, Charles
    Bradley, David
    Schmidt, Bernhard
    INVESTIGATIVE RADIOLOGY, 2013, 48 (08) : 584 - 589
  • [24] Optimizing Radiation Dose Levels in Prospectively Electrocardiogram-Triggered Coronary Computed Tomography Angiography Using Iterative Reconstruction Techniques: A Phantom and Patient Study
    Hou, Yang
    Zheng, Jiahe
    Wang, Yuke
    Yu, Mei
    Vembar, Mani
    Guo, Qiyong
    PLOS ONE, 2013, 8 (02):
  • [25] Significant Dose Reduction for Pediatric Digital Subtraction Angiography Without Impairing Image Quality: Preclinical Study in a Piglet Model
    Racadio, John
    Strauss, Keith
    Abruzzo, Todd
    Patel, Manish
    Kukreja, Kamlesh
    Johnson, Neil
    den Hartog, Mark
    Hoornaert, Bart
    Nachabe, Rami
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 203 (04) : 904 - 908
  • [26] Estimated Radiation Dose Reduction Using Adaptive Statistical Iterative Reconstruction in Coronary CT Angiography: The ERASIR Study
    Leipsic, Jonathon
    LaBounty, Troy M.
    Heilbron, Brett
    Min, James K.
    Mancini, G. B. John
    Lin, Fay Y.
    Taylor, Carolyn
    Dunning, Allison
    Earls, James P.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (03) : 655 - 660
  • [27] Assessment of image quality and radiation dose in some models of digital radiography systems - A Pehamed FLUORAD A plus D phantom study
    Kaheni, H.
    Chaparian, A.
    Nafisi-Moghadam, R.
    Hamzian, N.
    RADIOPROTECTION, 2018, 53 (02) : 139 - 144
  • [28] The potential use of ultra-low radiation dose images in digital mammography-a clinical proof-of-concept study in craniocaudal views
    Bluekens, A. M. J.
    Veldkamp, W. J. H.
    Schuur, K. H.
    Karssemeijer, N.
    Broeders, M. J. M.
    Den Heeten, G. J.
    BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1047)
  • [29] Reduction of the estimated radiation dose and associated patient risk with prospective ECG-gated 256-slice CT coronary angiography
    Efstathopoulos, E. P.
    Kelekis, N. L.
    Pantos, I.
    Brountzos, E.
    Argentos, S.
    Grebac, J.
    Ziaka, D.
    Katritsis, D. G.
    Seimenis, I.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (17) : 5209 - 5222
  • [30] Dual-source CT Angiography in Aortic Stent Grafting: An in vitro Aorta Phantom Study of Image Noise and Radiation Dose
    Sun, Zhonghua
    Ng, Curtise
    ACADEMIC RADIOLOGY, 2010, 17 (07) : 884 - 893