A Simple Source Model for 6 MV Flattening Filter Free Photon Beams Monte Carlo Dose Calculations

被引:1
作者
Ezzati, Ahad Ollah [1 ]
Studenski, Matthew T. [2 ]
Jamshidi, Negin [1 ]
机构
[1] Univ Tabriz, Dept Phys, Tabriz 5166616471, Iran
[2] Univ Miami, Dept Radiat Oncol, Miami, FL 33331 USA
关键词
source model; FFF photon beams; Monte Carlo; IMRT; MULTIPLE SOURCE MODEL; VIRTUAL SOURCE MODEL; LINEAR-ACCELERATOR; SIMULATION; RADIOTHERAPY; VALIDATION;
D O I
10.1134/S1547477121070037
中图分类号
O412 [相对论、场论]; O572.2 [粒子物理学];
学科分类号
摘要
The purpose of this work was to develop a simple source model for Monte Carlo dose calculations of flattening filter free (FFF) 6 MV photon beams. MCNPX2.4 Monte Carlo code was used to simulate 6 MV FFF photon beam from a Varian 2100C/D Clinac. The FFF photon beam features including energy and angular distributions were calculated in a phase space plane located 25.1 cm downstream from the accelerator target. A simple point source model (SM) was developed based on correlated histograms. Differences between measured and calculated profiles were less than 2% of the maximum dose or 2 mm distance to agreement at 100 cm source to surface distance (SSD). Using the obtained SM, percent depth dose and beam profiles were calculated for different field sizes from 3 x 3 to 40 x 40 cm(2). Differences between the SM and Clinac full simulation data were less than 2% or 2 mm distance to agreement at 100 cm SSD. This model is independent of components located above the accelerator movable jaws. Also, it has source biasing capabilities that increase the simulation speed up to 2000 times for certain field sizes. Our simulations showed that the dose rates of the FFF beams are at least two times higher for the same current of incident electrons on the target.
引用
收藏
页码:791 / 798
页数:8
相关论文
共 50 条
  • [31] Monte carlo model and output factors of elekta Infinity™ 6 and 10 MV photon beam
    Yani, Sitti
    Budiansah, Indra
    Rhani, Mohamad Fahdillah
    Haryanto, Freddy
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2020, 25 (04) : 470 - 478
  • [32] Dose measurements compared with Monte Carlo simulations of narrow 6 MV multileaf collimator shaped photon beams
    De Vlamynck, K
    Palmans, H
    Verhaegen, F
    De Wagter, C
    De Neve, W
    Thierens, H
    MEDICAL PHYSICS, 1999, 26 (09) : 1874 - 1882
  • [33] A direction-selective flattening filter for clinical photon beams. Monte Carlo evaluation of a new concept
    Chofor, Ndimofor
    Harder, Dietrich
    Willborn, Kay
    Ruehmann, Antje
    Poppe, Bjoern
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (14) : 4355 - 4376
  • [34] Investigation of neutron contamination of flattening filter and flattening filter-free 10-MV photon beams in Elekta InfinityTM accelerator
    Yani S.
    Budiansah I.
    Lestari F.P.
    Tursinah R.
    Rhani M.F.
    Haryanto F.
    Iranian Journal of Medical Physics, 2020, 17 (02) : 126 - 132
  • [35] Validation of BEAMnrc Monte Carlo model for a 12 MV photon beam
    Mohammed, Maged
    El Bardouni, T.
    Chakir, E.
    Saeed, M.
    Jamal, Al Zain
    Mohamed, Landour
    JOURNAL OF KING SAUD UNIVERSITY SCIENCE, 2018, 30 (04) : 537 - 543
  • [36] Evaluation of dosimetric parameters of small fields of 6 MV flattening filter free photon beam measured using various detectors against Monte Carlo simulation
    Mamballikalam, Gopinath
    Senthilkumar, S.
    Jayadevan, P. M.
    Jaon Bos, R. C.
    Ahamed Basith, P. M.
    Inippully, Rohit
    Shine, N. S.
    Clinto, C. O.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2021, 20 (03) : 332 - 339
  • [37] Verification measurements and clinical evaluation of the iPlan RT Monte Carlo dose algorithm for 6 MV photon energy
    Petoukhova, A. L.
    van Wingerden, K.
    Wiggenraad, R. G. J.
    van de Vaart, P. J. M.
    van Egmond, J.
    Franken, E. M.
    van Santvoort, J. P. C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (16) : 4601 - 4614
  • [38] Effect of an integral quality monitor on 4-, 6-, 10-MV, and 6-MV flattening filter-free photon beams
    Nguyen, Trang Hong Thi
    Yokoyama, Haruna
    Kojima, Hironori
    Isomura, Naoki
    Takemura, Akihiro
    Ueda, Shinichi
    Noto, Kimiya
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (01): : 76 - 91
  • [39] Monte carlo study on 6 MV photon beams of a CyberKnife® stereotactic radiosurgery system
    Dong Xiaoqing
    Luo Wenyun
    Yue Kun
    Wang Chuanshan
    Zhou Shiqing
    Pan Fuxing
    Wang Enmin
    Wang Chaozhuang
    NUCLEAR SCIENCE AND TECHNIQUES, 2010, 21 (01) : 16 - 19
  • [40] Monte carlo study on 6 MV photon beams of a CyberKnife stereotactic radiosurgery system
    DONG Xiaoqing1 LUO Wenyun1 YUE Kun1 WANG Chuanshan1 ZHOU Shiqing1 PAN Fuxing1 WANG Enmin2 WANG Chaozhuang2 1 Shanghai Applied Radiation Institute
    Nuclear Science and Techniques, 2010, 21 (01) : 16 - 19