Modeling and Estimation of Tip Contact Force for Steerable Ablation Catheters

被引:67
作者
Khoshnam, Mahta [1 ,2 ]
Skanes, Allan C. [3 ,4 ]
Patel, Rajni V. [1 ,2 ,5 ]
机构
[1] Univ Western Ontario, Dept Elect & Comp Engn, London, ON N6A 5A5, Canada
[2] Canadian Surg Technol & Adv Robot, London, ON N6A 5A5, Canada
[3] Univ Western Ontario, Schulich Sch Med & Dent, London, ON N6A 5A5, Canada
[4] Univ Western Ontario Hosp, London Hlth Sci Ctr, Electrophysiol Lab, London, ON N6A 5A5, Canada
[5] Univ Western Ontario, Dept Surg, London, ON N6A 5A5, Canada
基金
加拿大创新基金会; 加拿大自然科学与工程研究理事会;
关键词
Ablation catheter; contact force; kinematics; modeling;
D O I
10.1109/TBME.2015.2389615
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: The efficacy of catheter-based cardiac ablation procedures can be significantly improved if real-time information is available concerning contact forces between the catheter tip and cardiac tissue. However, the widely used ablation catheters are not equipped for force sensing. This paper proposes a technique for estimating the contact forces without direct force measurements by studying the changes in the shape of the deflectable distal section of a conventional 7-Fr catheter (henceforth called the "deflectable distal shaft," the "deflectable shaft," or the "shaft" of the catheter) in different loading situations. Method: First, the shaft curvature when the tip is moving in free space is studied and based on that, a kinematic model for the deflectable shaft in free space is proposed. In the next step, the shaft shape is analyzed in the case where the tip is in contact with the environment, and it is shown that the curvature of the deflectable shaft provides useful information about the loading status of the catheter and can be used to define an index for determining the range of contact forces exerted by the ablation tip. Results: Experiments with two different steerable ablation catheters show that the defined index can detect the range of applied contact forces correctly in more than 80% of the cases. Based on the proposed technique, a framework for obtaining contact force information by using the shaft curvature at a limited number of points along the deflectable shaft is constructed. Conclusion: The proposed kinematic model and the force estimation technique can be implemented together to describe the catheter's behavior before contact, detect tip/tissue contact, and determine the range of contact forces. Significance: This study proves that the flexibility of the catheter's distal shaft provides a means of estimating the force exerted on tissue by the ablation tip.
引用
收藏
页码:1404 / 1415
页数:12
相关论文
共 50 条
  • [41] Practical approach to subject-specific estimation of knee joint contact force
    Knarr, Brian A.
    Higginson, Jill S.
    JOURNAL OF BIOMECHANICS, 2015, 48 (11) : 2897 - 2902
  • [42] Assessing the Relationship of Applied Force and Ablation Duration on Lesion Size Using a Diamond Tip Catheter Ablation System
    Verma, Atul
    Schmidt, Megan M.
    Lalonde, Jean-Pierre
    Ramirez, David A.
    Getman, Michael K.
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2021, 14 (07) : 660 - 668
  • [43] Steerable versus non-steerable sheaths during pulmonary vein isolation: impact of left atrial enlargement on the catheter-tissue contact force
    Masuda, Masaharu
    Fujita, Masashi
    Iida, Osamu
    Okamoto, Shin
    Ishihara, Takayuki
    Nanto, Kiyonori
    Kanda, Takashi
    Shiraki, Tatsuya
    Sunaga, Akihiro
    Matsuda, Yasuhiro
    Uematsu, Masaaki
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2016, 47 (01) : 99 - 107
  • [44] Real-world safety of catheter ablation for atrial fibrillation with contact force or cryoballoon ablation
    Natale, Andrea
    Mohanty, Sanghamitra
    Goldstein, Laura
    Gomez, Tara
    Hunter, Tina D.
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2021, 60 (03) : 445 - 452
  • [45] Higher contact force, energy setting, and impedance rise during radiofrequency ablation predicts charring: New insights from contact force-guided in vivo ablation
    Makimoto, Hisaki
    Metzner, Andreas
    Tilz, Roland Richard
    Lin, Tina
    Heeger, Christian-H
    Rillig, Andreas
    Mathew, Shibu
    Lemes, Christine
    Wissner, Erik
    Kuck, Karl-Heinz
    Ouyang, Feifan
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2018, 29 (02) : 227 - 235
  • [46] Contact force-sensing catheters: performance in an ex vivo porcine heart model
    Eduardo Franco
    Daniel Rodríguez Muñoz
    Roberto Matía
    Antonio Hernández-Madrid
    Inmaculada Sánchez Pérez
    José Luis Zamorano
    Javier Moreno
    Journal of Interventional Cardiac Electrophysiology, 2018, 53 : 141 - 150
  • [47] Real-world safety of catheter ablation for atrial fibrillation with contact force or cryoballoon ablation
    Andrea Natale
    Sanghamitra Mohanty
    Laura Goldstein
    Tara Gomez
    Tina D. Hunter
    Journal of Interventional Cardiac Electrophysiology, 2021, 60 : 445 - 452
  • [48] Towards FBG-based End-Effector Force Estimation for a Steerable Continuum Robot
    Brumfiel, Timothy A.
    Sarma, Achraj
    Desai, Jaydev P.
    2022 INTERNATIONAL SYMPOSIUM ON MEDICAL ROBOTICS (ISMR), 2022,
  • [49] Estimation method of static height of contact wire by using pantograph contact force
    Usuda, Takayuki
    Ikeda, Mitsuru
    Quarterly Report of RTRI (Railway Technical Research Institute), 2015, 56 (03) : 175 - 180
  • [50] Lesion Size and Safety Comparison Between the Novel Flex Tip on the FlexAbility Ablation Catheter and the Solid Tips on the ThermoCool and ThermoCool SF Ablation Catheters
    Winterfield, Jeffrey R.
    Jensen, James
    Gilbert, Therese
    Marchlinski, Francis
    Natale, Andrea
    Packer, Douglas
    Reddy, Vivek
    Mahapatra, Srijoy
    Wilber, David J.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2016, 27 (01) : 102 - 109