Modeling and Estimation of Tip Contact Force for Steerable Ablation Catheters

被引:67
作者
Khoshnam, Mahta [1 ,2 ]
Skanes, Allan C. [3 ,4 ]
Patel, Rajni V. [1 ,2 ,5 ]
机构
[1] Univ Western Ontario, Dept Elect & Comp Engn, London, ON N6A 5A5, Canada
[2] Canadian Surg Technol & Adv Robot, London, ON N6A 5A5, Canada
[3] Univ Western Ontario, Schulich Sch Med & Dent, London, ON N6A 5A5, Canada
[4] Univ Western Ontario Hosp, London Hlth Sci Ctr, Electrophysiol Lab, London, ON N6A 5A5, Canada
[5] Univ Western Ontario, Dept Surg, London, ON N6A 5A5, Canada
基金
加拿大创新基金会; 加拿大自然科学与工程研究理事会;
关键词
Ablation catheter; contact force; kinematics; modeling;
D O I
10.1109/TBME.2015.2389615
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: The efficacy of catheter-based cardiac ablation procedures can be significantly improved if real-time information is available concerning contact forces between the catheter tip and cardiac tissue. However, the widely used ablation catheters are not equipped for force sensing. This paper proposes a technique for estimating the contact forces without direct force measurements by studying the changes in the shape of the deflectable distal section of a conventional 7-Fr catheter (henceforth called the "deflectable distal shaft," the "deflectable shaft," or the "shaft" of the catheter) in different loading situations. Method: First, the shaft curvature when the tip is moving in free space is studied and based on that, a kinematic model for the deflectable shaft in free space is proposed. In the next step, the shaft shape is analyzed in the case where the tip is in contact with the environment, and it is shown that the curvature of the deflectable shaft provides useful information about the loading status of the catheter and can be used to define an index for determining the range of contact forces exerted by the ablation tip. Results: Experiments with two different steerable ablation catheters show that the defined index can detect the range of applied contact forces correctly in more than 80% of the cases. Based on the proposed technique, a framework for obtaining contact force information by using the shaft curvature at a limited number of points along the deflectable shaft is constructed. Conclusion: The proposed kinematic model and the force estimation technique can be implemented together to describe the catheter's behavior before contact, detect tip/tissue contact, and determine the range of contact forces. Significance: This study proves that the flexibility of the catheter's distal shaft provides a means of estimating the force exerted on tissue by the ablation tip.
引用
收藏
页码:1404 / 1415
页数:12
相关论文
共 50 条
  • [31] Catheter ablation of right ventricular outflow tract tachycardia using contact force guidance
    S. D. A. Valk
    N. M. S. de Groot
    L. Jordaens
    Netherlands Heart Journal, 2014, 22 : 460 - 462
  • [32] Electromagnetic Contact-Force Sensing Electrophysiological Catheters: How Accurate Is the Technology?
    Bourier, Felix
    Hessling, Gabriele
    Ammar-Busch, Sonia
    Kottmaier, Marc
    Buiatti, Alessandra
    Grebmer, Christian
    Telishevska, Marta
    Semmler, Verena
    Lennerz, Carsten
    Schneider, Christine
    Kolb, Christof
    Deisenhofer, Isabel
    Reents, Tilko
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2016, 27 (03) : 347 - 350
  • [33] Modeling of contact area, contact force, and contact stiffness of mechanical systems with friction
    Abdo, Jamil
    Shamseldin, Elhanafi
    PROCEEDINGS OF THE ASME TRIBOLOGY DIVISION - 2005, 2005, : 295 - 302
  • [34] Fiberoptic Contact-Force Sensing Electrophysiological Catheters: How Precise Is the Technology?
    Bourier, Felix
    Gianni, Carola
    Dare, Matthew
    Deisenhofer, Isabel
    Hessling, Gabriele
    Reents, Tilko
    Mohanty, Sanghamitra
    Trivedi, Chintan
    Natale, Andrea
    Al-Ahmad, Amin
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2017, 28 (01) : 109 - 114
  • [35] Combined local impedance and contact force for radiofrequency ablation assessment
    Garrott, Kara
    Laughner, Jacob
    Gutbrod, Sarah
    Sugrue, Alan
    Shuros, Allan
    Sulkin, Matt
    Yasin, Omar
    Bush, Jamie
    Pottinger, Nathan
    Meyers, Jason
    Kapa, Suraj
    HEART RHYTHM, 2020, 17 (08) : 1371 - 1380
  • [36] Use of contact force technology for cardiac arrhythmia ablation in children
    Cruz-Baquero, Linibeth
    Pena Pachon, Paola
    Molano-Gonzalez, Nicolas
    Arenas, Alvaro
    HEART RHYTHM O2, 2022, 3 (06): : 833 - 838
  • [37] Persistent Atrial Fibrillation Ablation With or Without Contact Force Sensing
    Hussein, Ayman A.
    Barakat, Amr F.
    Saliba, Walid I.
    Tarakji, Khaldoun G.
    Bassiouny, Mohamed
    Baranowski, Bryan
    Tchou, Patrick
    Bhargava, Mandeep
    Dresing, Thomas
    Callahan, Thomas
    Cantillon, Daniel
    Kanj, Mohamed
    Lindsay, Bruce D.
    Wazni, Oussama M.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2017, 28 (05) : 483 - 488
  • [38] Factors Influencing Contact Force in Robotic Magnetic Navigation Ablation
    Orczykowski, Michal
    Bak, Maciej
    Kaczmarek, Krzysztof
    Urbanek, Piotr
    Robert, Bodalski
    Dubowski, Krzysztof
    Warminski, Grzegorz
    Derejko, Pawel
    Ptaszynski, Pawel
    Sterlinski, Maciej
    Bilinska, Maria
    Szumowski, Lukasz
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2025, : 855 - 862
  • [39] Estimation of wear and strain of contact wire using contact force of pantograph
    Current Collection Laboratory, Railway Dynamics Division
    Q Rep RTRI (Jpn), 2007, 3 (170-175): : 170 - 175
  • [40] Steerable versus non-steerable sheaths during pulmonary vein isolation: impact of left atrial enlargement on the catheter–tissue contact force
    Masaharu Masuda
    Masashi Fujita
    Osamu Iida
    Shin Okamoto
    Takayuki Ishihara
    Kiyonori Nanto
    Takashi Kanda
    Tatsuya Shiraki
    Akihiro Sunaga
    Yasuhiro Matsuda
    Masaaki Uematsu
    Journal of Interventional Cardiac Electrophysiology, 2016, 47 : 99 - 107