Modeling and Estimation of Tip Contact Force for Steerable Ablation Catheters

被引:67
|
作者
Khoshnam, Mahta [1 ,2 ]
Skanes, Allan C. [3 ,4 ]
Patel, Rajni V. [1 ,2 ,5 ]
机构
[1] Univ Western Ontario, Dept Elect & Comp Engn, London, ON N6A 5A5, Canada
[2] Canadian Surg Technol & Adv Robot, London, ON N6A 5A5, Canada
[3] Univ Western Ontario, Schulich Sch Med & Dent, London, ON N6A 5A5, Canada
[4] Univ Western Ontario Hosp, London Hlth Sci Ctr, Electrophysiol Lab, London, ON N6A 5A5, Canada
[5] Univ Western Ontario, Dept Surg, London, ON N6A 5A5, Canada
基金
加拿大创新基金会; 加拿大自然科学与工程研究理事会;
关键词
Ablation catheter; contact force; kinematics; modeling;
D O I
10.1109/TBME.2015.2389615
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: The efficacy of catheter-based cardiac ablation procedures can be significantly improved if real-time information is available concerning contact forces between the catheter tip and cardiac tissue. However, the widely used ablation catheters are not equipped for force sensing. This paper proposes a technique for estimating the contact forces without direct force measurements by studying the changes in the shape of the deflectable distal section of a conventional 7-Fr catheter (henceforth called the "deflectable distal shaft," the "deflectable shaft," or the "shaft" of the catheter) in different loading situations. Method: First, the shaft curvature when the tip is moving in free space is studied and based on that, a kinematic model for the deflectable shaft in free space is proposed. In the next step, the shaft shape is analyzed in the case where the tip is in contact with the environment, and it is shown that the curvature of the deflectable shaft provides useful information about the loading status of the catheter and can be used to define an index for determining the range of contact forces exerted by the ablation tip. Results: Experiments with two different steerable ablation catheters show that the defined index can detect the range of applied contact forces correctly in more than 80% of the cases. Based on the proposed technique, a framework for obtaining contact force information by using the shaft curvature at a limited number of points along the deflectable shaft is constructed. Conclusion: The proposed kinematic model and the force estimation technique can be implemented together to describe the catheter's behavior before contact, detect tip/tissue contact, and determine the range of contact forces. Significance: This study proves that the flexibility of the catheter's distal shaft provides a means of estimating the force exerted on tissue by the ablation tip.
引用
收藏
页码:1404 / 1415
页数:12
相关论文
共 50 条
  • [21] Quantification of Mechanical Characteristics of Conventional Steerable Ablation Catheters for Treatment of Atrial Fibrillation Using a Heart Phantom
    Miller, Jacob
    Sewani, Alykhan
    Rezazada, Jeffrey
    Alawneh, Yara
    Kazemian, Pedram
    Terricabras, Maria
    Wright, Graham
    Tavallaei, M. Ali
    CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2023, 14 (03) : 419 - 427
  • [22] Impact of Steerable Sheaths on Contact Forces and Reconnection Sites in Ablation for Persistent Atrial Fibrillation
    Ullah, Waqas
    Hunter, Ross J.
    Mclean, Ailsa
    Dhinoja, Mehul
    Earley, Mark J.
    Sporton, Simon
    Schilling, Richard J.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2015, 26 (03) : 266 - 273
  • [23] Robotics-Assisted Control of Steerable Ablation Catheters Based on the Analysis of Tendon-Sheath Transmission Mechanisms
    Khoshnam, Mahta
    Patel, Rajni V.
    IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2017, 22 (03) : 1473 - 1484
  • [24] Risk of atrioesophageal fistula formation with contact force-sensing catheters
    Black-Maier, Eric
    Pokorney, Sean D.
    Barnett, Adam S.
    Zeitler, Emily P.
    Sun, Albert Y.
    Jackson, Kevin P.
    Bahnson, Tristram D.
    Daubert, James P.
    Piccini, Jonathan P.
    HEART RHYTHM, 2017, 14 (09) : 1328 - 1333
  • [25] A randomized trial of contact force in atrial flutter ablation
    Giehm-Reese, Mikkel
    Kronborg, Mads Brix
    Lukac, Peter
    Kristiansen, Steen Buus
    Jensen, Henrik Kjaerulf
    Gerdes, Christian
    Kristensen, Jens
    Nielsen, Jan Moller
    Nielsen, Jens Cosedis
    EUROPACE, 2020, 22 (06): : 947 - 955
  • [26] Contact force ablation of accessory pathways in pediatric patients
    Pook, Caitlin
    Kuhn, Evelyn
    Singh, Anoop
    Kovach, Joshua
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2021, 32 (02) : 370 - 375
  • [27] Contact force with magnetic-guided catheter ablation
    Bessiere, Francis
    Zikry, Christopher
    Rivard, Lena
    Dyrda, Katia
    Khairy, Paul
    EUROPACE, 2018, 20 : 1 - 4
  • [28] Force-Sensing Catheters During Pediatric Radiofrequency Ablation: The FEDERATION Study
    Dalal, Aarti S.
    Nguyen, Hoang H.
    Bowman, Tammy
    Van Hare, George F.
    Silva, Jennifer N. Avari
    JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2017, 6 (05):
  • [29] A Comparison of Remote Magnetic Irrigated Tip Ablation versus Manual Catheter Irrigated Tip Catheter Ablation With and Without Force Sensing Feedback
    Weiss, J. Peter
    May, Heidi T.
    Bair, Tami L.
    Crandall, Brian G.
    Cutler, Michael J.
    Day, John D.
    Osborn, Jeffrey S.
    Mallender, Charles
    Bunch, T. Jared
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2016, 27 : S5 - S10
  • [30] Contact Force-Sensing versus Standard Catheters in Non-Fluoroscopic Radiofrequency Catheter Ablation of Idiopathic Outflow Tract Ventricular Arrhythmias
    Karkowski, Grzegorz
    Kuniewicz, Marcin
    Zabek, Andrzej
    Kozluk, Edward
    Debski, Maciej
    Matusik, Pawel T.
    Lelakowski, Jacek
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (03)