Microfluidic device for single step measurement of protein C in plasma samples for sepsis prognosis

被引:5
作者
Damodara, Sreekant [1 ]
Arora, Jaskirat [2 ]
Dwivedi, Dhruva J. [2 ]
Liaw, Patricia C. [2 ]
Fox-Robichaud, Alison E. [2 ]
Selvaganapathy, P. Ravi [1 ]
机构
[1] McMaster Univ, Dept Mech Engn, Hamilton, ON, Canada
[2] McMaster Univ, Dept Med, Hamilton, ON, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
FACTOR-IX; CHROMATOGRAPHY;
D O I
10.1039/d1lc01084k
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Protein C is a vitamin K dependant protein in plasma that plays an essential role in regulating the coagulation cascade and inflammatory response. As a result of its importance in these roles, it has been suggested as a biomarker for prognosis of patients affected by sepsis. Sepsis is a dysregulated host response to an infection that is the leading cause of mortality in U.S. hospitals and results in the highest cost of hospitalization. It was found that protein C concentration in non-surviving sepsis patients is significantly lower (1.8 mu g mL(-1)) than in survivors and healthy patients who have a protein C concentration of 3.9-5.9 mu g mL(-1). Current methods for diagnosing sepsis rely on expensive immunoassays or functional assays that require multiple steps for isolation and activation of protein C. We demonstrate in this paper a low cost, single step assay for detection of protein C in blood plasma. This was done by combining isoelectric gates with barium-immobilized metal affinity trapping. The electric field was optimized for use with immobilized metal affinity using COMSOL simulation. The integrated device was tested with samples containing buffered protein C, protein C in the presence of high concentration bovine serum albumin and alpha 1-proteinase inhibitor, and in blood plasma with spiked protein C. The stability of the measured values was tested by monitoring the intensity of a mixture of protein C with BSA and A1PI every minute to determine that measurement after 40 minutes was optimal. The results showed that the device could be used to distinguish a reduction in protein C from 4.46 mu g mL(-1) to 1.96 mu g mL(-1) with greater than 98% confidence in plasma making it suitable for sepsis prognosis.
引用
收藏
页码:2566 / 2577
页数:12
相关论文
共 39 条
[1]  
Abcam, HUM PROT C ELISA KIT
[2]  
AssayGenie, HUM VIT K DEP PROT C
[3]  
Boyer C. B, 1985, A History of Mathematics
[4]  
Canfield W.M., 1981, METHOD ENZYMOL, V80, P320
[5]  
COMP PC, 1984, BLOOD, V63, P15
[6]   PURIFICATION OF HUMAN VITAMIN-K-DEPENDENT PROTEIN-S AND ITS LIMITED PROTEOLYSIS BY THROMBIN [J].
DAHLBACK, B .
BIOCHEMICAL JOURNAL, 1983, 209 (03) :837-846
[7]  
Dalton JC, 1997, ADV EXP MED BIOL, V411, P419
[8]   Single step separation and concentration of biomarker proteins using agarose based miniaturized isoelectric gates for point of care diagnostics [J].
Damodara, Sreekant ;
Dwivedi, Dhruva J. ;
Liaw, Patricia C. ;
Fox-Robichaud, Alison E. ;
Selvaganapathy, P. Ravi .
SENSORS AND ACTUATORS B-CHEMICAL, 2021, 330
[9]  
DIB JG, 2004, HOSP PHARM, V39, P558
[10]   CHARACTERIZATION OF PROTEIN S, A GAMMA-CARBOXYGLUTAMIC ACID CONTAINING PROTEIN FROM BOVINE AND HUMAN-PLASMA [J].
DISCIPIO, RG ;
DAVIE, EW .
BIOCHEMISTRY, 1979, 18 (05) :899-904