Tracking Ophthalmic Drugs in the Eye Using Confocal Fluorescence Microscopy

被引:4
作者
Buttenschoen, K. K. [1 ]
Girkin, J. M. [1 ]
Daly, D. [2 ]
机构
[1] Univ Durham, Dept Phys, CfAI, South Rd, Durham DH1 3LE, England
[2] Reading Enterprise Ctr, Lein Appl Diag, Reading RG6 6BU, Berks, England
来源
ADVANCED BIOMEDICAL AND CLINICAL DIAGNOSTIC SYSTEMS X | 2012年 / 8214卷
关键词
Confocal microscopy; fluorescence microscopy; ocular pharmacokinetics; drug diffusion; aqueous humour; EXCITATION;
D O I
10.1117/12.906135
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We report on the development of a non-invasive instrument based on scanning confocal microscopy for tracking inherently fluorescent drugs and measuring spatial features in the anterior chamber of the eye. The new instrument incorporates all features of the initial instrument(1) with the addition of fluorescence detection from within the anterior chamber of the eye. We have measured the diffusion of Fluorescein with high time resolution within a cuvette, an artificial eye and ex vivo porcine eyes. Results are be presented that demonstrate the capability of the instrument to accurately measure the concentration and the location of the fluorescent drug over a given period of time along the optical axis of the eye with an axial resolution of under 200 mu m and temporal resolution of < 1s. We show that the instrument has high sensitivity and can measure concentrations of < 1 mu M/L of compounds having a quantum yield as low as 0.01 with high specificity for the compound of interest over competing background signals. The role of the instrument in assessing the efficiency of any inherently fluorescent ophthalmic drug as well as monitoring other medication that might produce fluorescent compounds in the eye will be discussed. We furthermore believe that the instrument might also be capable of monitoring certain bodily processes which have an impact on the compounds present in the eye.
引用
收藏
页数:8
相关论文
共 9 条
[1]   USE OF A XENON FLASH TUBE AS EXCITATION SOURCE IN A NEW SLIT-LAMP FLUOROPHOTOMETER [J].
BRUBAKER, RF ;
COAKES, RL .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1978, 86 (04) :474-484
[2]  
Buttenschön KK, 2010, CLIN OPTOM, V2, P67
[3]  
Lakowicz J. R., 1956, PRINCIPLES FLUORESCE, P56
[4]   FLUOROPHOTOMETRIC APPARATUS FOR THE OBJECTIVE DETERMINATION OF FLUORESCENCE IN THE ANTERIOR CHAMBER OF THE LIVING EYE [J].
LANGHAM, M ;
WYBAR, KC .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1954, 38 (01) :52-57
[5]  
MCLAREN JW, 1985, INVEST OPHTH VIS SCI, V26, P144
[6]  
Noecker Robert J, 2006, Ther Clin Risk Manag, V2, P193, DOI 10.2147/tcrm.2006.2.2.193
[7]   Permeability of cornea, sclera, and conjunctiva: A literature analysis for drug delivery to the eye [J].
Prausnitz, MR ;
Noonan, JS .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1998, 87 (12) :1479-1488
[8]  
Resnikoff S, 2004, B WORLD HEALTH ORGAN, V82, P844
[9]   IMPROVED OBJECTIVE SLIT-LAMP FLUOROPHOTOMETER USING TUNGSTEN-HALOGEN LAMP EXCITATION AND SYNCHRONOUS DETECTION [J].
SMITH, AT ;
JONES, DP ;
STURROCK, GD ;
WRIGHT, P .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1977, 61 (11) :722-725