Dosimetric effect of translational and rotational errors for patients undergoing image-guided stereotactic body radiotherapy for spinal metastases

被引:100
作者
Wang, He [1 ]
Shiu, Almon [1 ]
Wang, Congjun [1 ]
O'Daniel, Jennifer [1 ]
Mahajan, Anita [2 ]
Woo, Shiao [2 ]
Liengsawangwong, Praimakorn [1 ]
Mohan, Radhe [1 ]
Chang, Eric L. [2 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Radiat Oncol, Houston, TX 77030 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2008年 / 71卷 / 04期
关键词
dosimetry; spinal metastases; image-guided; IG; stereotactic body radiotherapy; SBRT; setup errors;
D O I
10.1016/j.ijrobp.2008.02.074
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To investigate the dosimetric effects of translational and rotational patient positioning errors on the treatment of spinal and paraspinal metastases using computed tomography image-guided stereotactic body radiotherapy. The results of this study provide guidance for the treatment planning process and recognition of the dosimetric consequences of daily patient treatment setup errors. Methods and Materials: The data from 20 patients treated for metastatic spinal cancer using image-guided stereotactic body radiotherapy were investigated in this study. To simulate the dosimetric effects of residual setup uncertainties, 36 additional plans (total, 756 plans) were generated for each isocenter (total, 21 isocenters) on the planning computed tomography images, which included isocenter lateral, anteroposterior, superoinferior shifts, and patient roll, yaw, and pitch rotations. Tumor volume coverage and the maximal dose to the organs at risk were compared with those of the original plan. Six daily treatments were also investigated to determine the dosimetric effect with or without the translational and rotational corrections. Results: A 2-mm error in translational patient positioning error in any direction can result in >5 % tumor coverage loss and >25% maximal dose increase to the organs at risk. Rotational correction is very important for patients with multiple targets and for the setup of paraspinal patients when the isocenter is away from bony structures. Compared with the original plans, the daily treatment data indicated that translational adjustments could correct most of the setup errors to mean divergences of -1.4% for tumor volume coverage and -0.3% for the maximal dose to the organs at risk. Conclusion: For the best dosimetric results, spinal stereotactic treatments should have setup translational errors of <= 1 mm and rotational errors of <= 2 degrees. (C) 2008 Elsevier Inc.
引用
收藏
页码:1261 / 1271
页数:11
相关论文
共 20 条
[1]   A novel method to correct for pitch and yaw patient setup errors in helical tomotherapy [J].
Boswell, SA ;
Jeraj, R ;
Ruchala, KJ ;
Olivera, GH ;
Jaradat, HA ;
James, JA ;
Gutierrez, A ;
Pearson, D ;
Frank, G ;
Mackie, TR .
MEDICAL PHYSICS, 2005, 32 (06) :1630-1639
[2]   Phase I clinical evaluation of near-simultaneous computed tomographic image-guided stereotactic body radiotherapy for spinal metastases [J].
Chang, EL ;
Shiu, AS ;
Lii, MF ;
Rhines, LD ;
Mendel, E ;
Mahajan, A ;
Weinberg, JS ;
Mathews, LA ;
Brown, BW ;
Maor, MH ;
Cox, JD .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (05) :1288-1294
[3]   An analysis of the accuracy of the Cyberknife: A robotic nameless stereotactic radiosurgical system [J].
Chang, SD ;
Main, W ;
Martin, DP ;
Gibbs, IC ;
Heilbrun, MP .
NEUROSURGERY, 2003, 52 (01) :140-146
[4]   CyberKnife stereotactic radiosurgical treatment of spinal tumors for pain control and quality of life [J].
Degen, JW ;
Gagnon, GJ ;
Voyadzis, JM ;
McRae, DA ;
Lunsden, M ;
Dieterich, S ;
Molzahn, I ;
Henderson, FC .
JOURNAL OF NEUROSURGERY-SPINE, 2005, 2 (05) :540-549
[5]  
Gerszten PC, 2004, NEUROSURGERY, V55, P89, DOI 10.1227/01.NEU.0000126873.00845.A7
[6]   Image-guided radiotherapy using megavoltage cone-beam computed tomography for treatment of paraspinous tumors in the presence of orthopedic hardware [J].
Hansen, Eric K. ;
Larson, David A. ;
Aubin, Michele ;
Chen, Josephine ;
Descovich, Martina ;
Gillis, Amy M. ;
Morin, Olivier ;
Xia, Ping ;
Pouliot, Jean .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (02) :323-326
[7]   Evaluation of concave dose distributions created using an inverse planning system [J].
Hunt, MA ;
Hsiung, CY ;
Spirou, SV ;
Chui, CS ;
Amols, HI ;
Ling, CC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 54 (03) :953-962
[8]   Online planning and delivery technique for radiotherapy of spinal metastases using cone-beam CT:: Image quality and system performance [J].
Letourneau, Daniel ;
Wong, Rebecca ;
Moseley, Douglas ;
Sharpe, Michael B. ;
Ansell, Stephen ;
Gospodarowicz, Mary ;
Jaffray, David A. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (04) :1229-1237
[9]   Evaluation of image-guided helical tomotherapy for the retreatment of spinal metastases [J].
Mahan, SL ;
Ramsey, CR ;
Scaperoth, DD ;
Chase, DJ ;
Byrne, TE .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (05) :1576-1583
[10]  
MATTHIAS G, 2007, RADIOTHER ONCOL, V84, P56