Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads

被引:81
作者
Hauser, Robert G. [1 ]
Abdelhadi, Raed [1 ]
McGriff, Deepa [1 ]
Retel, Linda Kallinen [1 ]
机构
[1] Minneapolis Heart Inst Fdn, Minneapolis, MN 55407 USA
关键词
Implantable cardioverter-defibrillator; Lead; Insulation; Complications; Failure;
D O I
10.1016/j.hrthm.2012.03.048
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Riata and Riata ST leads (St Jude Medical, Sylmar, CA) were recalled in 2011. These leads are prone to externalized conductors. However, it is malfunctions that may result in serious adverse events, including death. OBJECTIVE To assess the deaths of patients with Riata and Riata ST leads that have been reported to the Food and Drug Administration to determine whether they were due to lead malfunction. A similar analysis was performed for Quattro Secure leads (Medtronic, Inc, Minneapolis, MN). METHODS In February 2012, we searched the Food and Drug Administration's Manufacturers and User Facility Device Experience database for deaths associated with Riata, Riata ST, and Quattro Secure leads. A lead-related death was a sudden or unexpected death accompanied by evidence of lead malfunction; an indeterminate death was a death that was nonsudden or evidence of lead malfunction was not provided; a not lead-related death was a death that was nonsudden and/or there was no evidence that the lead contributed to the patient's demise. RESULTS The Manufacturers and User Facility Device Experience database search found 133 deaths; of these, 22 were caused by Riata or Riata ST lead failure and 5 were caused by Quattro Secure lead failure. Riata and Riata ST lead failure deaths were typically caused by short circuits between high-voltage components. No death was due to externalized conductors. CONCLUSIONS Riata and Riata ST implantable cardioverter-defibrillator leads are prone to high-voltage failures that have resulted in death. These failures appeared to have been caused by insulation defects that resulted in short circuiting between high-voltage components. Externalized conductors were not a factor in these deaths.
引用
收藏
页码:1227 / 1235
页数:9
相关论文
共 13 条
  • [1] Carlson M, 2011, MED DEVICE ADVISORY
  • [2] Chester KM, 2010, IMPORTANT PRODUCT IN
  • [3] Implantable cardioverter-defibrillator lead disintegration at the level of the tricuspid valve
    Duray, Gabor Z.
    Israel, Carsten W.
    Schmitt, Joern
    Hohnloser, Stefan H.
    [J]. HEART RHYTHM, 2008, 5 (08) : 1224 - 1225
  • [4] Insulation Defects of Thin High-Voltage ICD Leads: An Underestimated Problem?
    Erkapic, Damir
    Duray, Gabor Z.
    Bauernfeind, Tamas
    De Rosa, Salvatore
    Hohnloser, Stefan H.
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2011, 22 (09) : 1018 - 1022
  • [5] Unexpected ICD pulse generator failure due to electronic circuit damage caused by electrical overstress
    Hauser, RG
    Hayes, DL
    Almquist, AK
    Epstein, AE
    Parsonnet, V
    Tyers, GFO
    Vlay, SC
    Schoenfeld, MH
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (07): : 1046 - 1054
  • [6] Early failure of a small-diameter high-voltage implantable cardioverter-defibrillator lead
    Hauser, Robert G.
    Kallinen, Linda M.
    Almquist, Adrian K.
    Gornick, Charles C.
    Katsiyiannis, William T.
    [J]. HEART RHYTHM, 2007, 4 (07) : 892 - 896
  • [7] Riata implantable cardioverter-defibrillator lead failure: Analysis of explanted leads with a unique insulation defect
    Hauser, Robert G.
    McGriff, Deepa
    Retel, Linda Kallinen
    [J]. HEART RHYTHM, 2012, 9 (05) : 742 - 749
  • [8] Longevity of Sprint Fidelis Implantable Cardioverter-Defibrillator Leads and Risk Factors for Failure Implications for Patient Management
    Hauser, Robert G.
    Maisel, William H.
    Friedman, Paul A.
    Kallinen, Linda M.
    Mugglin, Andrew S.
    Kumar, Kapil
    Hodge, David O.
    Morrison, Thomas B.
    Hayes, David L.
    [J]. CIRCULATION, 2011, 123 (04) : 358 - U247
  • [9] Deaths and cardiovascular injuries due to device-assisted implantable cardioverter-defibrillator and pacemaker lead extraction
    Hauser, Robert G.
    Katsiyiannis, William T.
    Gornick, Charles C.
    Almquist, Adrian K.
    Kallinen, Linda M.
    [J]. EUROPACE, 2010, 12 (03): : 395 - 401
  • [10] Keung E, VA NATL CARDIAC DEVI