Enzymatic electrochemical biosensor for urea with a polyaniline grafted conducting hydrogel composite modified electrode

被引:61
作者
Das, Joyati [1 ]
Sarkar, Priyabrata [1 ]
机构
[1] Univ Calcutta, Dept Polymer Sci & Technol, Biosensor Lab, 92 APC Rd, Kolkata 700009, W Bengal, India
关键词
COVALENTLY IMMOBILIZED UREASE; SENSOR; NANOCOMPOSITE; FILM; FABRICATION; FIBER; SERUM;
D O I
10.1039/c6ra12159d
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Conducting polymer hydrogels (CPH) are a unique class of materials, which have both the advantageous features of hydrogels and organic conductors. Herein, we report a new urea biosensor in which the working electrode is modified by a CPH, specifically a composite made of conducting electro-polymerized aniline in the non-conducting hydrogel of polyacrylamide and polyvinyl alcohol for immobilization of the enzyme urease. Scanning electron microscopy (SEM) and atomic force microscopy (AFM) studies depict the porous structure of the hydrogel, which favours the high density immobilization of enzymes. Fourier transform infrared spectroscopy (FTIR) studies confirmed the grafting of polyaniline within the matrix. Electrochemical impedance spectroscopy (EIS), differential pulse voltammetry (DPV) and cyclic voltammetry (CV) predict high enzyme loading in the hydrogel. Bode plot analysis with the [FeCN6](3-/4-) redox couple shows the partial capacitive behaviour of the bioelectrode in the frequency range of 10(2) to 10(3) Hz. Factors, such as scan rate and pH of the electrolyte, were optimized to give the best results. DPV could detect 1.5-1000 mu M urea with the lowest detection limit (LOD) of 60 nM and sensitivity of 878 mu A mM(-1) cm(-2). The relatively low Michaelis-Menten constant (K-m) value of 113 mM indicates the high affinity of the enzyme for urea. EIS depicts an increase in charge transfer resistance (RCT) with an increase in urea concentration (LOD = 14 mu M). The proposed sensor exhibits a very good performance with respect to detection limit, sensitivity, K-m value, preparation method, cost, sample preparation, stability, selectivity and reproducibility. This sensor was successfully applied for the detection of urea in milk, puffed rice, soil and human blood. Statistical analysis of the results shows good agreement with commercial spectrophotometric analysis.
引用
收藏
页码:92520 / 92533
页数:14
相关论文
共 43 条
[1]   Highly stable urea sensor based on ZnO nanorods directly grown on Ag/glass electrodes [J].
Ahmad, Rafiq ;
Tripathy, Nirmalya ;
Hahn, Yoon-Bong .
SENSORS AND ACTUATORS B-CHEMICAL, 2014, 194 :290-295
[2]   Urea sensor based on tin oxide thin films prepared by modified plasma enhanced CVD [J].
Ansari, S. G. ;
Ansari, Z. A. ;
Seo, Hyung-Kee ;
Kim, Gil-Sung ;
Kim, Young-Soon ;
Khang, Gilson ;
Shin, Hyung-Shik .
SENSORS AND ACTUATORS B-CHEMICAL, 2008, 132 (01) :265-271
[3]   Conducting polymer coated carbon surfaces and biosensor applications [J].
Ates, Murat ;
Sarac, A. Sezai .
PROGRESS IN ORGANIC COATINGS, 2009, 66 (04) :337-358
[4]   A novel thin film amperometric urea biosensor based on urease-immobilized on poly(N-glycidylpyrrole-co-pyrrole) [J].
Bozgeyik, Ibrahim ;
Senel, Mehmet ;
Cevik, Emre ;
Abasiyanik, M. Fatih .
CURRENT APPLIED PHYSICS, 2011, 11 (04) :1083-1088
[5]  
Ciana L.D., 1996, CLIN CHEM, V42, P1079
[6]   An aqueous ammonia sensor based on an inkjet-printed polyaniline nanoparticle-modified electrode [J].
Crowley, Karl ;
O'Malley, Eimer ;
Morrin, Aoife ;
Smyth, Malcolm R. ;
Killard, Anthony J. .
ANALYST, 2008, 133 (03) :391-399
[7]   Amperometric urea biosensors based on sulfonated graphene/polyaniline nanocomposite [J].
Das, Gautam ;
Yoon, Hyon Hee .
INTERNATIONAL JOURNAL OF NANOMEDICINE, 2015, 10 :55-66
[8]   Preparation of urease/nano-structured polyaniline-Nafion®/Au/Al2O3 electrode for inhibitive detection of mercury ion [J].
Do, Jing-Shan ;
Lin, Kai-Hsin ;
Ohara, Reiko .
JOURNAL OF THE TAIWAN INSTITUTE OF CHEMICAL ENGINEERS, 2011, 42 (04) :662-668
[9]  
Ensa A. A., 2012, ANAL CHIM ACTA, V726, P93
[10]   Immobilization of urease on nanostructured polymer membrane and preparation of urea amperometric biosensor [J].
Gabrovska, Katya ;
Ivanov, Javor ;
Vasileva, Ioana ;
Dimova, Nedyalka ;
Godjevargova, Tzonka .
INTERNATIONAL JOURNAL OF BIOLOGICAL MACROMOLECULES, 2011, 48 (04) :620-626