The investigation of dose and image quality of chest computed tomography using different combinations of noise index and adaptive statistic iterative reconstruction level

被引:7
作者
Sookpeng, Supawitoo [1 ]
Martin, Colin J. [2 ]
Butdee, Chitsanupong [1 ]
机构
[1] Naresuan Univ, Fac Allied Hlth Sci, Dept Radiol Technol, Phitsanulok, Thailand
[2] Univ Glasgow, Dept Clin Phys, Glasgow, Lanark, Scotland
关键词
Adaptive statistic iterative reconstruction; automatic tube current modulation; computed tomography optimization; TUBE CURRENT MODULATION; CT; REDUCTION;
D O I
10.4103/ijri.IJRI_124_18
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Computed tomography (CT) automatic tube current modulation (ATCM) systems and iterative reconstruction (IR) play an important role in CT radiation dose optimization. How the two can best be used together is one of the challenges faced by radiology professionals. Aim: To determine optimum settings of ATCM noise index (NI) together with adaptive statistic iterative reconstruction (ASIR) for a general electric (GE) scanner that aims to achieve similar image quality to the standard protocol used in the hospital (Smart mA technique with NI of 11.57 and 30% ASIR reconstruction) with a lower dose. Methods: Different NI and ASIR levels were set for scans of a phantom. Objective image quality assessments in terms of noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), low-contrast detectability (LCD), and modulation transfer function (MTF) were carried out in an anthropomorphic chest and a Catphan 700 phantom. Subjective image quality assessment was also performed with five readers to confirm whether the image quality of the new protocols was adequate. Result and Conclusion: SNR and CNR increased with the strength of ASIR, and decreased with higher NI settings. The MTF improved slightly for higher dose levels and from filtered back projection (FBP) to higher strength of ASIR. LCD improved with ASIR compared to FBP and with higher strengths of ASIR. Qualitative scoring ranged between 3.0 and 4.6. A moderate degree of reliability was found between scoring. Use of NI 15.04 with 70% ASIR can reduce dose by 41% compared to the standard protocol of NI 11.57 with 30% ASIR without degradation of image quality.
引用
收藏
页码:53 / 60
页数:8
相关论文
共 28 条
[21]   Adaptive Statistical Iterative Reconstruction Technique for Radiation Dose Reduction in Chest CT: A Pilot Study [J].
Singh, Sarabjeet ;
Kalra, Mannudeep K. ;
Gilman, Matthew D. ;
Hsieh, Jiang ;
Pien, Homer H. ;
Digumarthy, Subba R. ;
Shepard, Jo-Anne O. .
RADIOLOGY, 2011, 259 (02) :565-573
[22]   INVESTIGATION OF THE INFLUENCE OF IMAGE RECONSTRUCTION FILTER AND SCAN PARAMETERS ON OPERATION OF AUTOMATIC TUBE CURRENT MODULATION SYSTEMS FOR DIFFERENT CT SCANNERS [J].
Sookpeng, Supawitoo ;
Martin, Colin J. ;
Gentle, David J. .
RADIATION PROTECTION DOSIMETRY, 2015, 163 (04) :521-530
[23]  
Task Group on Control of Radiation Dose in Computed Tomography, 2000, Ann ICRP, V30, P7
[24]  
The Phantom Laboratory, 2017, CATPH 700 US MAN CATPH 700 US MAN
[25]   A three-dimensional statistical approach to improved image quality for multislice helical CT [J].
Thibault, Jean-Baptiste ;
Sauer, Ken D. ;
Bouman, Charles A. ;
Hsieh, Jiang .
MEDICAL PHYSICS, 2007, 34 (11) :4526-4544
[26]   Radiation dose from CT scanning: can it be reduced? [J].
Trinavarat, Panruethai ;
Kritsaneepaiboon, Supika ;
Rongviriyapanich, Chantima ;
Visrutaratna, Pannee ;
Srinakarin, Jiraporn .
ASIAN BIOMEDICINE, 2011, 5 (01) :13-21
[27]   Image Quality Assessment of Standard- and Low-Dose Chest CT Using Filtered Back Projection, Adaptive Statistical Iterative Reconstruction, and Novel Model-Based Iterative Reconstruction Algorithms [J].
Vardhanabhuti, Varut ;
Loader, Robert J. ;
Mitchell, Grant R. ;
Riordan, Richard D. ;
Roobottom, Carl A. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2013, 200 (03) :545-552
[28]  
Xu Jingyan, 2009, J Am Coll Radiol, V6, P274, DOI 10.1016/j.jacr.2008.12.014