Real-time optoacoustic monitoring of temperature in tissues

被引:176
作者
Larina, IV
Larin, KV
Esenaliev, RO [1 ]
机构
[1] Univ Texas, Ctr Biomed Engn, Med Branch, Galveston, TX 77555 USA
[2] Univ Texas, Med Branch, Dept Neurosci & Cell Biol, Galveston, TX 77555 USA
[3] Univ Texas, Med Branch, Dept Anesthesiol, Galveston, TX 77555 USA
关键词
D O I
10.1088/0022-3727/38/15/015
中图分类号
O59 [应用物理学];
学科分类号
摘要
To improve the safety and efficacy of thermal therapy, it is necessary to map tissue temperature in real time with submillimetre spatial resolution. Accurate temperature maps may provide the necessary control of the boundaries of the heated regions and minimize thermal damage to surrounding normal tissues. Current imaging modalities fail to monitor tissue temperature in real time with high resolution and accuracy. We investigated a non-invasive optoacoustic method for accurate, real-time monitoring of tissue temperature during thermotherapy. In this study, we induced temperature gradients in tissue and tissue-like samples and monitored the temperature distribution using the optoacoustic technique. The fundamental harmonic of a Q-switched Nd: YAG laser (lambda = 1064 nm) was used for optoacoustic wave generation and probing of tissue temperature. The tissue temperature was also monitored with a multi-sensor temperature probe inserted in the samples. Good agreement between optoacoustically measured and actual tissue temperatures was obtained. The accuracy of temperature monitoring was better than 1 degrees C, while the spatial resolution was about I mm. These data suggest that the optoacoustic technique has the potential to be used for non-invasive, real-time temperature monitoring during thermotherapy.
引用
收藏
页码:2633 / 2639
页数:7
相关论文
共 30 条
  • [11] Gusev V. E., 1993, LASER OPTOACOUSTICS
  • [12] Karabutov AA, 2003, LASER PHYS, V13, P711
  • [13] KIKOIN IK, 1976, TABLES PHYS PARAMETE
  • [14] Application of quasi-resonant ultrasonic detectors in the optoacoustic tomography of biological objects
    Kirillov, AG
    Lozhkarev, VV
    Mansfel'd, AD
    Reiman, AM
    Friedman, GI
    Yakovlev, IV
    [J]. QUANTUM ELECTRONICS, 1999, 29 (09) : 832 - 835
  • [15] Optoacoustic laser monitoring of cooling and freezing of tissues
    Larin, KV
    Larina, IV
    Motamedi, M
    Esenaliev, RO
    [J]. QUANTUM ELECTRONICS, 2002, 32 (11) : 953 - 958
  • [16] Effect of temperature on the optical properties of ex vivo human dermis and subdermis
    Laufer, J
    Simpson, R
    Kohl, M
    Essenpreis, M
    Cope, M
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (09) : 2479 - 2489
  • [17] Lehmann JF, 1990, Therapeutic Heat and Cold
  • [18] Noninvasive temperature estimation in tissue via ultrasound echo-shifts .1. Analytical model
    MaassMoreno, R
    Damianou, CA
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1996, 100 (04) : 2514 - 2521
  • [19] Noninvasive optoacoustic temperature determination at the fundus of the eye during laser irradiation
    Schüle, G
    Hüttmann, G
    Framme, C
    Roider, O
    Brinkmann, R
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2004, 9 (01) : 173 - 179
  • [20] NONINVASIVE ESTIMATION OF TISSUE TEMPERATURE RESPONSE TO HEATING FIELDS USING DIAGNOSTIC ULTRASOUND