Optimization of an ultralow-dose high-resolution pediatric PET scanner design based on monolithic scintillators with dual-sided digital SiPM readout: a simulation study

被引:16
作者
Mikhaylova, Ekaterina [1 ,2 ]
Tabacchini, Valerio [3 ]
Borghi, Giacomo [3 ]
Mollet, Pieter [4 ]
D'Hoe, Ester [1 ,5 ]
Schaart, Dennis R. [3 ]
Vandenberghe, Stefaan [1 ]
机构
[1] Univ Ghent, iMinds Med IT, IBiTech, Dept Elect & Informat Syst,MEDISIP, Pintelaan 185,Block B, B-9000 Ghent, Belgium
[2] Univ Calif Davis, Dept Biomed Engn, One Shields Ave, Davis, CA 95616 USA
[3] Delft Univ Technol, Radiat Sci & Technol, Mekelweg 15, NL-2629 JB Delft, Netherlands
[4] Univ Ghent, iMinds Med IT, Molecubes, Ottergemsesteenweg Zuid 808 Bus 325, B-9000 Ghent, Belgium
[5] Vrije Univ Brussel, Labo ICMI, Dept Nucl Med, Laarbeeklaan 103, B-1090 Jette, Belgium
关键词
pediatric TOF-PET; monolithic LYSO; GATE simulation; TOF-PET; DETECTORS; PERFORMANCE; PET/MRI; MRI; CALIBRATION; CHALLENGES; SYSTEMS;
D O I
10.1088/1361-6560/aa8eb2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The goal of this simulation study is the performance evaluation and comparison of six potential designs for a time-of-flight PET scanner for pediatric patients of up to about 12 years of age. It is designed to have a high sensitivity and provide high-contrast and high-resolution images. The simulated pediatric PET is a full ring scanner, consisting of 32 x 32 mm(2) monolithic LYSO:Ce crystals coupled to digital silicon photomultiplier arrays. The six considered designs differ in axial lengths (27.2 cm, 54.4 cm and 102 cm) and crystal thicknesses (22 mm and 11 mm). The simulations are based on measured detector response data. We study two possible detector arrangements: 22 mm-thick crystals with dual-sided readout and 11 mm-thick crystals with back-sided readout. The six designs are simulated by means of the GEANT4 application for tomographic emission software, using the measured spatial, energy and time response of the monolithic scintillator detectors as input. The performance of the six designs is compared on the basis of four studies: (1) spatial resolution; (2) NEMA NU2-2012 sensitivity and scatter fraction (SF) tests; (3) non-prewhitening signal-to-noise ratio observer study; and (4) receiver operating characteristics analysis. Based on the results, two designs are identified as cost-effective solutions for fast and efficient imaging of children: one with 54.4 cm axial field-of-view (FOV) and 22 mm-thick crystals, and another one with 102 cm axial FOV and 11 cm-thick crystals. The first one has a higher center point sensitivity than the second one, but requires dual-sided readout. The second design has the advantage of allowing a whole-body scan in a single bed position acquisition. Both designs have the potential to provide an excellent spatial resolution (similar to 2 mm) and an ultra-high sensitivity (>100 cps kBq(-1)).
引用
收藏
页码:8402 / 8418
页数:17
相关论文
共 31 条
[1]  
Abbey C K, 1997, P SPIE
[2]   ROOT - A C++ framework for petabyte data storage, statistical analysis and visualization [J].
Antcheva, I. ;
Ballintijn, M. ;
Bellenot, B. ;
Biskup, M. ;
Brun, R. ;
Buncic, N. ;
Canal, Ph. ;
Casadei, D. ;
Couet, O. ;
Fine, V. ;
Franco, L. ;
Ganis, G. ;
Gheata, A. ;
Maline, D. Gonzalez ;
Goto, M. ;
Iwaszkiewicz, J. ;
Kreshuk, A. ;
Segura, D. Marcos ;
Maunder, R. ;
Moneta, L. ;
Naumann, A. ;
Offermann, E. ;
Onuchin, V. ;
Panacek, S. ;
Rademakers, F. ;
Russo, R. ;
Tadel, M. .
COMPUTER PHYSICS COMMUNICATIONS, 2009, 180 (12) :2499-2512
[3]   A 32 mm x 32 mm x 22 mm monolithic LYSO:Ce detector with dual-sided digital photon counter readout for ultrahigh-performance TOF-PET and TOF-PET/MRI [J].
Borghi, Giacomo ;
Peet, Bart Jan ;
Tabacchini, Valerio ;
Schaart, Dennis R. .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (13) :4929-4949
[4]   Towards monolithic scintillator based TOF-PET systems: practical methods for detector calibration and operation [J].
Borghi, Giacomo ;
Tabacchini, Valerio ;
Schaart, Dennis R. .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (13) :4904-4928
[5]   Experimental Validation of an Efficient Fan-Beam Calibration Procedure for k-Nearest Neighbor Position Estimation in Monolithic Scintillator Detectors [J].
Borghi, Giacomo ;
Tabacchini, Valerio ;
Seifert, Stefan ;
Schaart, Dennis R. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2015, 62 (01) :57-67
[6]   Future Image Acquisition Trends for PET/MRI [J].
Boss, Andreas ;
Weiger, Markus ;
Wiesinger, Florian .
SEMINARS IN NUCLEAR MEDICINE, 2015, 45 (03) :201-211
[7]   PET-MRI: Challenges and new directions [J].
Daftary, Aditya .
INDIAN JOURNAL OF NUCLEAR MEDICINE, 2010, 25 (01) :3-5
[8]   ON THE ANGULAR DISTRIBUTION OF 2-PHOTON ANNIHILATION RADIATION [J].
DEBENEDETTI, S ;
COWAN, CE ;
KONNEKER, WR ;
PRIMAKOFF, H .
PHYSICAL REVIEW, 1950, 77 (02) :205-212
[9]   NEMA NU 2-2012 performance studies for the SiPM-based ToF-PET component of the GE SIGNA PET/MR system [J].
Grant, Alexander M. ;
Deller, Timothy W. ;
Khalighi, Mohammad Mehdi ;
Maramraju, Sri Harsha ;
Delso, Gaspar ;
Levin, Craig S. .
MEDICAL PHYSICS, 2016, 43 (05) :2334-2343
[10]   PET/MR in children. Initial clinical experience in paediatric oncology using an integrated PET/MR scanner [J].
Hirsch, Franz Wolfgang ;
Sattler, Bernhard ;
Sorge, Ina ;
Kurch, Lars ;
Viehweger, Adrian ;
Ritter, Lutz ;
Werner, Peter ;
Jochimsen, Thies ;
Barthel, Henryk ;
Bierbach, Uta ;
Till, Holger ;
Sabri, Osama ;
Kluge, Regine .
PEDIATRIC RADIOLOGY, 2013, 43 (07) :860-875